Bard® X-Force® Balloon Dilation Catheter 998404Recalls & reported problems
C. R. Bard, Inc. · Bard® X-Force® Balloon Dilation Catheter · Class 2
All models of Bard X-Force Balloon Dilation Catheter (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10801741052641), received between May 2, 2017 and Jun 17, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Inflation Problem3
- Crack2
- Material Fragmentation2
- Pressure Problem2
- Burst Container or Vessel1
- Device Handling Problem1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Foreign Body In Patient1
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJun 17, 2026Material Fragmentation
IT WAS REPORTED THAT A PATIENT WHO HAD A STENT PLACED AT OUR FACILITY AND WAS MANAGED AT ANOTHER FACILITY FOR APPROXIMATELY ONE YEAR PRESENTED TO OUR HOSPITAL WITH URINARY RETENTION. WHILE IT WAS POSSIBLE THAT THE STENT REMAINS INSIDE THE BODY, DETAILS CANNOT BE DETERMINED UNTIL THE URINARY RETENTION IS RESOLVED VIA A NEPHROSTOMY APPROACH. THEREFORE, THEY ARE REQUESTING INFORMATION ON WHETHER THERE HAVE BEEN ANY CASES IN THE PAST WHERE A STENT BROKE AND REMAINED INSIDE THE BODY AFTER ONE YEAR OF PLACEMENT. POSSIBILITY OF URINARY TRACT OBSTRUCTION DURING URETERAL STENT PLACEMENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body In Patient
MAUDE report 25537328 ↗MalfunctionFeb 11, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT BALLOON PART OF THE BALLOON DILATION CATHETER WAS DAMAGED. THE BALLOON DID NOT INFLATE WHEN USED. PER FOLLOW UP INFORMATION RECEIVED VIA IBC ON 25JAN2025, STATED THAT THE DAMAGE TO THE BALLOON COULD NOT BE CONFIRMED BY VISUAL INSPECTION. THE BALLOON HAD NOT BURST. IT SEEMS TO BE LEAKING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21354577 ↗MalfunctionAug 8, 2023Fracture; Fracture
IT WAS REPORTED THAT THE CONNECTION TO THE EAGLE INFLATION DEVICE WAS CRACKED, AND THE BALLOON COULD NOT BE EXPANDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17480091 ↗MalfunctionFeb 8, 2023Burst Container or Vessel; Material Fragmentation; Burst Container or Vessel; Material Fragmentation
IT WAS REPORTED THAT THE SHAFT BELOW THE BALLOON GOT BROKEN DURING USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16317930 ↗MalfunctionDec 29, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE CONTRAST AGENT LEAKED FROM THE BASE OF THE CATHETER BALLOON AND THE BALLOON WAS UNABLE TO BE DILATED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11089819 ↗MalfunctionSep 25, 2020Break; Leak/Splash; Device Handling Problem; Device Handling Problem
IT WAS REPORTED THAT THE CATHETER LEAKED THE WATER BECAUSE THE INFLATION DEVICE WAS BROKEN BEFORE THE USE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10586886 ↗MalfunctionApr 11, 2019Crack; Inflation Problem; Pressure Problem; Inflation Problem
IT WAS REPORTED THAT THE PRESSURE DID NOT RISE AFTER THE CATHETER INSERTION. FOLLOWING REMOVAL, A CRACK WAS FOUND ON THE SHAFT. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8504736 ↗MalfunctionApr 11, 2019Break; Crack; Pressure Problem; Break
IT WAS REPORTED THAT THE PRESSURE DID NOT RISE AFTER THE CATHETER WAS INSERTED. FOLLOWING REMOVAL, A CRACK WAS FOUND ON THE SHAFT. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8503256 ↗MalfunctionNov 24, 2017Inflation Problem
IT WAS REPORTED THAT THE INFLATION DEVICE DID NOT MOVE DURING USAGE. THE INFLATION DEVICE WAS REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7057740 ↗MalfunctionMay 19, 2017Inflation Problem; Material Rupture
IT WAS REPORTED THAT THE SHAFT OF THE CATHETER HAD RUPTURED AS THE USER ATTEMPTED PRESSURE WITH THE INFLATION DEVICE. THE USER INSERTED THE CATHETER INTO THE STRUCTURE AREA OF THE URETERAL, AND ATTEMPTED PRESSURE WITH THE INFLATION DEVICE, BUT THE BALLOON OF THE CATHETER COULD NOT BE INFLATED AND THE METER DID NOT INDICATE APPLIED PRESSURE. THE USER FOUND THAT THE SHAFT OF THE CATHETER HAD RUPTURED, THEN THE CATHETER WAS REMOVED FROM THE PATIENT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6578670 ↗MalfunctionMay 2, 2017Break
IT WAS REPORTED THAT THE JUNCTION PART OF THE SHAFT, WHICH WAS ON THE FRONT SIDE OF THE BIFURCATION AREA, WAS BROKEN AS THE USER ATTEMPTED TO DILATE THE URETER WITH THE URETER CATHETER BY APPLYING 27 DIASTOLIC PRESSURE. ANOTHER OF THE SAME TYPE OF DEVICE WAS USED, AND THE TREATMENT WAS COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6538814 ↗
Other DILATOR, CATHETER, URETERAL devices
- Porges Coloplast RBD014Coloplast A/S · DILATOR, CATHETER, URETERAL1 recall0 reports
- Bard® X-Force® Balloon Dilation Catheter 998706C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL38 reports
- Bard® X-Force® Balloon Dilation Catheter 998606C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL27 reports
- Bard® X-Force® Balloon Dilation Catheter 998006C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL25 reports
- Bard® X-Force® Balloon Dilation Catheter 998806C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL23 reports
- Bard® X-Force® Balloon Dilation Catheter 998504C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL16 reports
- Bard® X-Force® Balloon Dilation Catheter 998004C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL15 reports
- Bard® X-Force® Balloon Dilation Catheter 998506C. R. Bard, Inc. · DILATOR, CATHETER, URETERAL15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page