Skip to content

UroForce® Balloon Dilation Catheter 887404Recalls & reported problems

C. R. Bard, Inc. · UroForce® Balloon Dilation Catheter · Class 2

All models of UroForce Balloon Dilation Catheter (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801741052948), received between May 24, 2018 and May 24, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Known Impact Or Consequence To Patient1
  • Rupture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 24, 2018Burst Container or Vessel

    IT WAS REPORTED THAT THE UROFORCE CATHETER BALLOON WAS INSERTED AND INFLATED TO 4 ATM USING THE EAGLE INFLATION DEVICE. IT WAS REPORTED THAT THE PHYSICIAN THEN ASKED THE NURSE TO INFLATE THE CATHETER BALLOON TO 8 ATM, AT WHICH TIME THE NURSE BEGAN TO USE THE EAGLE INFLATION DEVICE TO INCREASE THE PRESSURE IN THE BALLOON. IT WAS REPORTED THAT, WHEN THE PRESSURE GOT TO 5 ATM, THE BALLOON OF THE CATHETER BURST. IT WAS REPORTED THAT THERE WERE NO MISSING PIECES; HOWEVER, THE PATIENT'S URETER "RUPTURED," AND A STENT WAS PLACED TO ALLOW THE AREA TO HEAL.

    Patient problems: No Known Impact Or Consequence To Patient; Rupture

    MAUDE report 7541989 ↗

Other DILATOR, CATHETER, URETERAL devices

All DILATOR, CATHETER, URETERAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page