ARROW IPN046639Recalls & reported problems
Teleflex Incorporated · Epidural Anesthesia Kit · Class 2
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10801902013832), received between Oct 31, 2025 and Oct 31, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break12
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434999 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434996 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892,9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434953 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434931 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434913 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434878 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434789 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434762 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, .9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884, 9680794-2025-00885 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434720 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD SOME KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9680794-2025-00884 AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434688 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD 11 KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, AND 9680794-2025-00880.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434587 ↗MalfunctionOct 31, 2025Break; Break
IT WAS REPORTED THAT: "WE HAVE HAD 11 KITS ALL HAD BROKEN SODIUM CHLORIDE BOTTLES IN THE KITS. THE LIQUID WAS DRIED UP BUT THERE WERE SHARDS OF GLASS IN THE KIT. ALL WITH THE SAME LOT NUMBER". ASSOCIATED COMPLAINTS 9680794-2025-00881, 9680794-2025-00886, 9680794-2025-00888, 9680794-2025-00889, 9680794-2025-00890, 9680794-2025-00891, 9680794-2025-00892, 9680794-2025-00893, 9680794-2025-00894, 9 680794-2025-00884.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23434550 ↗
FDA clearance and classification
The FDA regulates product code CAZ (anesthesia conduction kit) under 21 CFR 868.5140 as a Class II device. 21 CFR 868.5140 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ARROW IPN046639
Has the ARROW IPN046639 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10801902013832), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ARROW IPN046639?
The FDA's MAUDE database lists 12 adverse event reports for this device, received on Oct 31, 2025. In those reports, the most common specific device problem is “Break” (12). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN046639?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN046639 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the ARROW IPN046639, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN046639). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Anesthesia conduction kit devices
- BD Luer-Lok, BD Glaspak 405671Becton, Dickinson And Company · Anesthesia conduction kit1 recall68 reports
- ARROW IPN922678Teleflex Incorporated · Anesthesia conduction kit2 recalls0 reports
- BD Luer-Lok, BD Glaspak 405672Becton, Dickinson And Company · Anesthesia conduction kit1 recall26 reports
- BD Luer-Lok, BD Glaspak 405673Becton, Dickinson And Company · Anesthesia conduction kit1 recall10 reports
- BD Luer-Lok, BD Glaspak 405637Becton, Dickinson And Company · Anesthesia conduction kit1 recall7 reports
- BD Luer-Lok, BD Glaspak 405707Becton, Dickinson And Company · Anesthesia conduction kit1 recall4 reports
- BD Luer-Lok, BD Glaspak 405609Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
- BD Luer-Lok, BD Glaspak 405735Becton, Dickinson And Company · Anesthesia conduction kit1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page