Skip to content

ARROW IPN033506Recalls & reported problems

Teleflex Incorporated · ARROW® PICC with Chlorag+ard® Technology · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10801902095715), received between Apr 8, 2024 and Dec 19, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

11 other recalls cover devices cleared under the same FDA submission (K112896); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2024Material Separation; Material Separation

    IT WAS REPORTED 'THE PATIENT HAS A UNILUMEN PICC LINE IN THE RIGHT UPPER LIMB FOR THE ADMINISTRATION OF TREATMENT. DURING THE DISINFECTION OF THE CATHETER PORT, THE PORT BROKE, AND THE BLUE SAFETY CLAMP DETACHED, THE REST OF THE CATHETER REMAINED INTACT, ALLOWING THE VENOCLYSIS EQUIPMENT TO BE CONNECTED, AND THERAPY WAS SUBSEQUENTLY STARTED'. THE PATIENT WAS FINE AND HAD NO HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20976129 ↗
  • MalfunctionSep 30, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: " A REPORT IS RECEIVED FROM ONCOLOGY INDICATING THAT THE CLAMP SPLIT AND DETACHED, LEAVING THE PATIENT WITH PICC INSERTED, BUT WITHOUT A SAFETY KEY."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20332820 ↗
  • MalfunctionApr 8, 2024Material Deformation; Material Deformation

    IT WAS REPORTED "DURING THE INSERTION OF THE PICC, AT THE MOMENT OF ADVANCING THE DILATOR AND INTRODUCER AND EVEN THOUGH NO FORCE WAS APPLIED, THE PEELABLE INTRODUCER OPENS AT THE TIP AND THE USE OF ANOTHER CATHETER IS NECESSARY." NO MEDICAL INTERVENTION REQUIRED. NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19059555 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112896. One submission often covers many models, so check the product and lot details in each record.

  • Z-2378-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR
  • Z-1820-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code:PR 45541 HPHNM
  • Z-1819-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code: PR 45541 HPHNL
  • Z-1811-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance PICC Product Code: CDC45541HPK1A
  • Z-1822-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code:PR 45552 HPHNM
  • Z-1327-2018 · Dec 22, 2017 · Terminated
    Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard Technology, Part Numbers: (1) ASK-45541-RWJ4; (2) ASK-45552-RWJ4
  • Z-0996-2018 · Nov 27, 2017 · Terminated
    Pressure Injectable Multi- Lumen Central Venous Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology Product Usage: The Arrow CVC is indicated to provide short-ter
  • Z-2698-2017 · Jun 12, 2017 · Terminated
    Arrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology
  • Z-2331-2017 · May 9, 2017 · Terminated
    1) Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology The Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology is indicated for
  • Z-1705-2017 · Mar 3, 2017 · Terminated
    The CG+ Arrow PIC Catheters permit venous access to the central circulation through a peripheral vein.
  • Z-0207-2016 · Oct 16, 2015 · Terminated
    ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates