Skip to content

ARROW IPN038527Recalls & reported problems

Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Plus(R) Three-Lumen CVC Kit · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10801902142167), received between Apr 29, 2026 and May 26, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

54 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 21, 2026Open, Classified

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

    Z-2384-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Material Separation2
  • Physical Resistance/Sticking2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2026Material Separation

    IT WAS REPORTED THAT "THE NURSE PRACTITIONER DURING INITIAL INSERTION OF THE CENTRAL VEIN CATHETER (CVC) OBSERVED THAT THE GUIDEWIRE BROKE (EXTERNAL TO THE PATIENT)." THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS "FINE , " AND NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25284408 ↗
  • MalfunctionMay 26, 2026Material Separation

    IT WAS REPORTED THAT THE PROVIDER ENCOUNTERED RESISTANCE WHILE ATTEMPTING TO WITHDRAW THE GUIDEWIRE THROUGH THE CATHETER. DURING THIS PROCESS, THE GUIDEWIRE FRACTURED WHILE BOTH THE GUIDEWIRE AND CATHETER WERE STILL POSITIONED WITHIN THE PATIENT. AN X-RAY WAS PERFORMED, WHICH CONFIRMED THAT NO FRAGMENTS REMAINED WITHIN THE PATIENT. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS "FINE." NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED BEYOND THE DIAGNOSTIC IMAGING PERFORMED TO CONFIRM THE ABSENCE OF RETAINED FRAGMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25281079 ↗
  • MalfunctionApr 29, 2026Physical Resistance/Sticking

    IT WAS REPORTED "THE CATH WOULD NOT THREAD OVER THE GUIDEWIRE." THIS WAS FOUND PRIOR TO USE. THE CUSTOMER REPORTED SIMILAR INCIDENTS HAS OCCURED HOWEVER FURTHER DETAILS WERE NOT PROVIDED FOR THOSE INCIDENTS. ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2026-00380 AND .

    Patient problems: Insufficient Information

    MAUDE report 25033255 ↗
  • MalfunctionApr 29, 2026Physical Resistance/Sticking

    IT WAS REPORTED "THE CATH WOULD NOT THREAD OVER THE GUIDEWIRE." THIS WAS FOUND PRIOR TO USE. THE CUSTOMER REPORTED SIMILAR INCIDENTS HAS OCCURED HOWEVER FUTHER DETAILS WERE NOT PROVIDED FOR THOSE INCIDENTS. ASSOCIATED MDR NUMBERS INCLUDE: (B)(4) AND (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25032979 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.

  • Z-2376-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
  • Z-1914-2026 · Mar 11, 2026 · Open, Classified
    ARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
  • Z-2123-2026 · Jan 7, 2026 · Open, Classified
    Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
  • Z-0003-2024 · Aug 10, 2023 · Open, Classified
    Arrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0002-2024 · Aug 10, 2023 · Open, Classified
    Arrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
  • Z-0352-2023 · Oct 27, 2022 · Terminated
    Pressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
  • Z-0351-2023 · Oct 27, 2022 · Terminated
    Arrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
  • Z-1604-2022 · May 25, 2022 · Open, Classified
    TELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
  • Z-1993-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable ARROWg+ard Blue PLUS(R) Two-Lumen Central Venous Catheterization Kit; Product Code: ASK-42802-PWHC1
  • Z-1990-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Code: ASK-45802-PHMW
  • Z-1989-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Quad-Lumen CVC Kit; Product Codes: CDC-42854-P1A and CDC-45854-P1A
  • Z-1995-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip, ARROWg+ard Blue PLUS Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product
  • Z-1988-2018 · Apr 11, 2018 · Terminated
    Pressure Injectable ARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Codes: ASK-42703-PAM1, ASK-42703-PSHR1, ASK-45703-PNY, ASK-45703-PWHC1
  • Z-1994-2018 · Apr 11, 2018 · Terminated
    ARROWg+ard Blue PLUS Pressure Injectable Multi-Lumen CVC Kit; Product Codes: CDC-45703-P1A; CDC-45703-PB1A; CDC-42703-P1A; and CDC-15703-P1A
  • Z-1987-2018 · Apr 11, 2018 · Terminated
    Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Codes: ASK-42703-MHW and ASK-42703-OLL1

…and 39 more.

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.