ARROW recallZ-2384-2026 · Arrow International, LLC · initiated Apr 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
Recall details
- Recall number
- Z-2384-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jun 13, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 1,061,069
- Product code
- LRS
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Product description
Description/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDT1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20CM/ASK-15703-LMDTG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-15703-PVH; 1L CVC KIT: 16 GA X 8 IN (20CM)/ASK-24301-JHH1; CVC KIT: 16 GA X 16 CM/CDC-24306-1A; CVC KIT: 3-L 7 FR X 30 CM ANTIMICROBIAL/CDC-24703-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-1A; CVC KIT: 2L 8 FR X 6IN (16 CM)/ASK-42802-ECL3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-ACH; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-BJC1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-DU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-EMC; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6IN/ASK-42703-LMDT1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-LMDTG1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-MCY1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-NS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAGP1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM)/ASK-42703-PAM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAMC2; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PAU1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16/ASK-42703-PB1; PI CVC KIT: 3L 7 FR X 16CM AGB+/ASK-42703-PBSWH2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16 IN/ASK-42703-PCAM1; MULTILUMEN PI CVC KIT: 3L 7 FR X 16 CM/ASK-42703-PCSJ2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PECL2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PGM4; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHHN1; MULTI LUMEN CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PLC; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMEM; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMHC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-POHU2; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU2; MULTILUMEN CVC KIT 7 FR X 6 IN (16 CM)/ASK-42703-PTJH; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUCH2; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-PUCI1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUHC2; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PUMH; MULTI-LUMEN PI CVC KIT: 3-L 7 FR/ASK-42703-PUPM1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-SHUF2; MULTI LUMEN PI CVC KIT: 3L 7 FR/ASK-42703-SL1; CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-42703-UMC; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-WHH; CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-CHM; CVC KIT: 2L 8 FR X 16 CM/ASK-42802-LMDT1; PI CATHETER 2L: 8FR X 16CM AGB+/ASK-42802-MCY1; PI CVC KIT: 2L 8FR X 16CM AGB+/ASK-42802-PBSWH2; MULTILUMEN CVC KIT: 2L 8 FR X 6/ASK-42802-POSU2; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PUNCC2; PI CVC KIT: 2-L 8 FR/ASK-42802-PUPM1; PI CVC KIT: 2L 8 FR/ASK-42802-TGH1; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-BJC1; CVC KIT: 4L 8.5 FR X (16CM) AGB+/ASK-42854-PBMH; PI CVC KIT: 4L 8.5FR X 16CM AGB+/ASK-42854-PMM; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-PUCH2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-ACH; 3L CVC KIT: 7 FR X 20 CM AGB+/ASK-45703-ARM; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-BJC1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-BLH; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-DU1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-HHC4; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-LMDT1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-LMDTG1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-MCY1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-NO; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-NS1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PADV1; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PAGP1; MULTI LUMEN CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PAMC1; MULTI LUMEN CVC KIT: 3L 7 FR X 8 IN/ASK-45703-PAU1; PI CVC KIT: 3L 7FR X 20CM AGB+/ASK-45703-PBSWH2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PCAM1; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCSJ2; PI CVC KIT: 3L 7 FR X 8 IN (20 CM) AGB+/ASK-45703-PDM1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PECL2; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8 IN/A
Code information (models, lots, serial numbers)
Show the full code information (26k characters)
REF(Material)/UDI-DI/Lot(Batch): ASK-04001-JHH/10801902193275/33F24K0748, 33F26A1669, 33F25G0558; ASK-04001-LMDTG/10801902193275, 10801902218442/33F26A0405, 33F24E0333; ASK-12703-PSPH1/10801902214338, 10801902193275, 10801902214338/33F24F0787, 33F24L0304, 33F25A0242, 33F25C0268, 33F25D0470, 33F25E0537, 33F25E0854, 33F25J0365, 33F25M0247, 33F26A1555, 33F26B0855, 33F26C0742, 33F24K0672; ASK-12703-WMC2/10801902193275/33F24K0493, 33F24L0305, 33F25B0156, 33F25E0593, 33F25H0483, 33F25J0367, 33F25M0218, 33F24K0545; ASK-15703-LMDT1/10801902193275/33F24J0164, 33F24L0307, 33F25B0192, 33F25C0230; ASK-15703-LMDTG1/10801902193275/33F25D0537, 33F25E0670, 33F25G0295, 33F25J0146, 33F25K0093, 33F25M0138, 33F26B0131, 33F25E0312; ASK-15703-PVH/10801902133226, 10801902193275, 10801902210217/33F24D0809, 33F24F0669, 33F24K0462, 33F24L0226, 33F25A0167, 33F25B0193, 33F25B0547, 33F25D0265, 33F25D0653, 33F25G0169, 33F25G0404, 33F25J0369, 33F26A1459, 33F24M0270; ASK-24301-JHH1/10801902193275, 10801902214819, 10801902210217/33F25A0170, 33F25B0552, 33F25E0547, 33F25G0222, 33F25L0056, 33F26B0421, 33F24E0546; CDC-24306-1A/30801902116919, 10801902193275, 30801902116919/33F24D0786, 33F24E0620, 33F24F0926, 33F24K0446, 33F24L0990, 33F25A0310, 33F25D0495, 33F25E0880, 33F25J0323, 33F26B0377, 33F26C0693, 33F24L1111; CDC-24703-X1A/10801902193275/33F25D0497, 33F25E0748, 33F25G0318, 33F25H0570, 33F25J0325, 33F26A1563, 33F25B0619; CDC-25123-1A/10801902193275, 10801902191622/33F24H0012, 33F24L0481, 33F25C0445, 33F25D0500, 33F25F0382, 33F25G0258, 33F26A1560, 33F26B0394, 33F25B0675; ASK-42802-ECL3/10801902193275/33F25E0567, 33F25F0128, 33F25H0464, 33F25J0286, 33F26A1399; ASK-42703-ACH/10801902193275, 30801902119514/33F24L0327, 33F24L1064, 33F25C0292, 33F25E0675, 33F25H0420, 33F25J0267, 33F25L0122, 33F25M0091, 33F26A1505, 33F24D0032; ASK-42703-BJC1/10801902193275, 10801902202519/33F24K0500, 33F24L0328, 33F25B0556, 33F25G0173, 33F25J0268, 33F26A1513, 33F26B0767, 33F25C0744; ASK-42703-DU1/10801902193275, 1080190216099/33F25B0141, 33F25C0232, 33F25E0786, 33F25J0105, 33F25M0014, 33F26C0342, 33F24F0060; ASK-42703-EMC/10801902160475, 10801902118421/33F24F0881, 33F25A0214, 33F25C0293, 33F25D0596, 33F25H0632, 33F25M0115, 33F26C0343, 33F24F0132; ASK-42703-LMDT1/10801902205657, 10801902193275/33F24F0688, 33F24J0157, 33F24L0228, 33F24L0329, 33F24L1096, 33F25A0252, 33F25B0557, 33F25C0352, 33F25E0557, 33F25E0722, 33F25G0400, 33F25K0139, 33F26A1370, 33F26B0210, 33F25C0294; ASK-42703-LMDTG1/10801902193275, 10801902205695, 10801902220896/33F25E0558, 33F25E0860, 33F25G0271, 33F25G0399, 33F25H0419, 33F26B0387, 33F24G0070; ASK-42703-MCY1/10801902193275, 0801902220896/33F24H0198, 33F24K0425, 33F25A0198, 33F25B0651, 33F25C0353, 33F25D0405, 33F25E0676, 33F25H0452, 33F25J0220, 33F25M0183, 33F26A1479, 33F26B0863, 33F25H0409; ASK-42703-NS1/10801902193275, 10801902216479/33F25A0300, 33F25B0652, 33F25C0375, 33F25E0723, 33F25H0459, 33F25J0221, 33F26A1579, 33F26B0825, 33F25C0354; ASK-42703-PAGP1/10801902193275, 10801902158717, 10801902136685/33F25D0406, 33F25E0560, 33F25E0790, 33F25H0560, 33F25L0114, 33F26A1353, 33F26A1506, 33F26B0751, 33F24D0181; ASK-42703-PAM1/10801902136685, 10801902193275/33F24E0310, 33F24L0895, 33F25A0256, 33F25B0496, 33F25C0477; ASK-42703-PAMC2/10801902193275, 10801902217131/33F25F0188, 33F25J0270, 33F26C0161, 33F24K0400; ASK-42703-PAU1/10801902193275/33F24L0414, 33F25B0143, 33F25C0474, 33F25G0397, 33F25M0313, 33F26A1349, 33F25B0232; ASK-42703-PB1/10801902193275, 10801902220377/33F25E0458, 33F25K0427, 33F24F0065; ASK-42703-PBSWH2/10801902220377, 10801902193275, 10801902220377, 10801902203189/33F24F0651, 33F24K0368, 33F24L0230, 33F24L0415, 33F24L1079, 33F25B0167, 33F25C0440, 33F25D0539, 33F25E0014, 33F25E0596, 33F25G0205, 33F25H0511, 33F25J0222, 33F25K0112, 33F25M0394, 33F26B0412, 33F24D0720; ASK-42703-PCAM1/10801902203189, 10801902193275, 10801902141030/33F24F0675, 33F24L1058, 33F25B0560, 33F25C0355, 33F25D0351, 33F25E0725, 33F25G0366, 33F25J0386, 33F25M0388, 33F26B0791, 33F24F0874; ASK-42703-PCSJ2/10801902193275, 10801902211672/33F25H0468, 33F25J0271, 33F26B0389, 33F26C0179, 33S24L0906, 33F25B0445; ASK-42703-PECL2/10801902193275/33F25H0466, 33F25J0272, 33F26A0011; ASK-42703-PGM4/10801902193275, 10801902221824/33F25B0653, 33F25D0541, 33F25J0106, 33F25J0387, 33F25K0140, 33F26B0749, 33F26C0027, 33F25E1195; ASK-42703-PHHN1/10801902193275, 10801902144734/33F25H0486, 33F25K0064, 33F25M0673, 33F26A0249, 33F26B0906, 33F25E0009; ASK-42703-PLC/10801902193275/33F25D0407, 33F25E0791, 33F25F0099, 33F25H0606, 33F26A1393, 33F25E1196; ASK-42703-PMEM/10801902193275, 10801902142167/33F25G0554, 33F25K0054, 33F25L0256, 33F26C0050, 33F24L0418; ASK-42703-PMHC1/10801902193275/10801902135244/33F25E0561, 33F25H0430, 33F25H0638, 33F25J0388, 33F25M0232, 33F26B0135, 33F26B0796, 33F25F0363; ASK-42703-PNHS1/10801902193275, 10801902225495/33F25D0542, 33F25E0599, 33F25H0498, 33F25H0650, 33F26D0066; ASK-42703-POHU2/10801902193275, 10801902212020/33F25E0562, 33F25G0284, 33F25J0389, 33F26A1421, 33F24F0662; ASK-42703-POSU2/10801902193275/33F24K0401, 33F24L0419, 33F24L0949, 33F25B0446, 33F25D0758; ASK-42703-PTJH/10801902193275, 10801902219012/33F24L0190, 33F25A0015, 33F25B0366, 33F25C0658, 33F25E0316, 33F24E0660; ASK-42703-PUCH2/10801902219012, 10801902193275, 10801902219012, 10801902135022/ ASK-42703-PUCI1/33F24F1025, 33F24L0336, 33F24L1009, 33F25B0476, 33F25C0356, 33F25E0021, 33F25F0189, 33F25H0455, 33F25J0275, 33F25K0142, 33F25M0120, 33F26A1518, 33F26B0182, 33F26B0781, 33F24F0850; ASK-42703-PUHC2/10801902193275, 10801902159608/33F25D0353, 33F25E0602, 33F25G0288, 33F25H0635, 33F25J0277, 33F25K0143, 33F26B0386, 33F26C0149, 33F24D0765; ASK-42703-PUMH/10801902129465, 10801902193275, 10801902159592/33F24F0697, 33F24K0709, 33F24L0997, 33F25C0301, 33F25E0677, 33F25J0278, 33F26A0256, 33F26B0789, 33F26C0227, 33F24D0698; ASK-42703-PUPM1/10801902159592, 10801902193275, 10801902225150/33F24F0761, 33F24K0499, 33F24L1074, 33F25B0519, 33F25C0357, 33F25D0354, 33F25D0598, 33F25E0678, 33F25G0346, 33F25J0279, 33F25M0126, 33F26B0363; ASK-42703-SHUF2/10801902193275, 10801902208368/33F25D0355, 33F25E0792, 33F25H0473, 33F25J0280, 33F25M0204, 33F26A1501, 33F26B0832, 33F24L0339; ASK-42703-SL1/10801902193275, 10801902205206/33F25B0483, 33F25C0443, 33F25D0482, 33F25E0793, 33F25G0306, 33F25J0281, 33F26A1437, 33F24D0631; ASK-42703-UMC/10801902193275, 10801902205206/33F24K0363, 33F24L0340, 33F25B0568,33F25E0801, 33F25G0328, 33F25J0282, 33F25M0194, 33F26A1444, 33F26B0769, 33F26C0221, 33F25B0158; ASK-42703-WHH/10801902193275, 10801902159189, 30801902119576/33F25E0565, 33F25G0350, 33F25J0283, 33F26B0777, 33F24E0552; ASK-42802-CHM/10801902193275/33F24H0178, 33F24K0480, 33F24L0859, 33F25D0236, 33F25F0190, 33F25H0518, 33F25J0284, 33F25M0139, 33F26B0175, 33F25E0005; ASK-42802-LMDT1/10801902205664, 10801902193275/33F24F0737, 33F24K0381, 33F24L0250, 33F24L0423, 33F25B0570, 33F25C0377, 33F25F0129, 33F25G0281, 33F25J0158, 33F25K0146, 33F26C0156, 33F25F0533; ASK-42802-MCY1/10801902193275, 10801902220384/33F24K0328, 33F24K0754, 33F24F0964; ASK-42802-PBSWH2/10801902193275, 10801902220384/33F24H0200, 33F24K0408, 33F24L0424, 33F24L1006, 33F25B0477, 33F25D0357, 33F25G0203, 33F25G0204, 33F25H0659, 33F25J0390, 33F25M0195, 33F25M0335, 33F26B0424, 33F24F0759; ASK-42802-POSU2/10801902193275/33F24K0473, 33F25D0759; ASK-42802-PUNCC2/10801902164251, 10801902193275, 10801902159585/33F24F0948, 33F24L0425, 33F25H0443, 33F25J0394, 33F24D0660; ASK-42802-PUPM1/10801902159585, 10801902193275, 10801902225143/33F24F0668, 33F24K0422, 33F24L0341, 33F24L1042, 33F25C0305, 33F25E0679, 33F25F0194, 33F25J0159, 33F25M0123, 33F26B0364; ASK-42802-TGH1/10801902213829, 10801902193275, 10801902213829/33F24L1054, 33F25B0521, 33F25D0311, 33F25E0604, 33F25G0225, 33F25J0288, 33F25L0204, 33F26A1543, 33F26B0844, 33F26C0224, 33F25C0306; ASK-42854-BJC1/10801902202540, 10801902193275, 10801902202540/33F24F0644, 33F24K0348, 33F24L0232, 33F25A0260, 33F25B0571, 33F25D0358, 33F25D0601, 33F25E0863, 33F25G0367, 33F26A1524, 33F26B0764, 33F26C0140, 33F24H0126; ASK-42854-PBMH/10801902193275, 40801902131469/33F26A0241, 33F24E0639; ASK-42854-PMM/40801902131469, 10801902193275/33F24F1014, 33F24L1013, 33F25C0307, 33F25D0546, 33F25E0802, 33F25H0480, 33F25J0398, 33F25A0019; ASK-42854-PUCH2/10801902193275, 10801902219050/33F25K0065, 33F26B0902, 33F24L0347; ASK-45703-ACH/10801902193275, 10801902203257, 30801902119644/33F25B0235, 33F25E0007, 33F25F0237, 33F25K0046, 33F25M0675, 33F26B0221, 33F24D0033; ASK-45703-ARM/10801902193275, 1080190215812/33F24L0903, 33F25B0449, 33F25D0363, 33F25E0682, 33F25H0479, 33F25J0228, 33F25K0150, 33F26A1423, 33F26B0849, 33F25F0004; ASK-45703-BJC1/10801902193275, 10801902202533, 10801902213782/33F24K0426, 33F24L0826, 33F25B0450, 33F25B0577, 33F25D0547, 33F25F0104, 33F25F0199, 33F25M0096, 33F26A1348, 33F26B0378, 33F26B0750, 33F24F0776; ASK-45703-BLH/10801902193275, 10801902213782, 10801902135381/33F24L1067, 33F25J0110, 33F25K0071, 33F25K0151, 33F25M0331, 33F26A1457, 33F26B0776, 33F24F0165; ASK-45703-DU1/10801902193275, 10801902160970, 10801902160468/33F25B0451, 33F25D0548, 33F25E0683, 33F25H0558, 33F26A1451, 33F26B0208, 33F26B0707, 33F26B0845, 33F24D0579; ASK-45703-HHC4/10801902193275, 10801902211467, 10801902205671/33F25D0364 ,33F25D0661, 33F25E0867, 33F25G0246, 33F25H0598, 33F25J0229, 33F25M0152, 33F26A1537, 33F26B0818, 33S25E0867, 33F24F0873; ASK-45703-LMDT1/10801902193275, 10801902205671/33F24K0440, 33F24L0432, 33F24L0919, 33F25B0160, 33F25D0365, 33F25D0603, 33F25E0608, 33F25G0326, 33F25H0475, 33F25J0230, 33F25M0153, 33F26A1552, 33F26B0880, 33F25B0453; ASK-45703-LMDTG1/10801902193275, 10801902205749, 10801902220902/33F25B0579, 33F25E0018, 33F25E0609, 33F25F0200, 33F25G0398, 33F25J0401, 33F25M0190, 33F25M0332, 33F26B0482, 33F26C0226, 33F24G0072; ASK-45703-MCY1/10801902193275/33F24J0132, 33F24K0445, 33F24L0433, 33F24L1028, 33F25B0657, 33F25E0022, 33F25E0610, 33F25F0201, 33F25H0657, 33F25M0144, 33F26A1573, 33F25L0143; ASK-45703-NO/10801902205213, 10801902193275, 10801902216417/33F24F0844, 33F24L0352, 33F25B0581, 33F25E0611, 33F24D0769; ASK-45703-NS1/10801902216417, 10801902193275, 10801902216417/33F24F0835, 33F24L0434, 33F24L1047, 33F25E0612, 33F25F0130, 33F25G0214, 33F25M0176, 33F26A1544, 33F26B0864, 33F25F0006; ASK-45703-PADV1/10801902193275/33F24J0146, 33F24K0458, 33F24L0353, 33F25C0311, 33F25E0423, 33F25E0684, 33F25H0561, 33F25M0127, 33F26A0775, 33F25A0265; ASK-45703-PAGP1/10801902193275, 10801902158724/33F25E0613, 33F25F0202, 33F25G0409, 33F25J0297, 33F26A1339, 33F26C0125, 33F25B0658; ASK-45703-PAMC1/10801902193275, 10801902217148, 10801902208344/33F25D0549, 33F25E0728, 33F25J0298, 33F25K0152, 33F26B0199, 33F24D0738; ASK-45703-PAU1/10801902208344, 10801902193275, 10801902220353/33F24F0783, 33F24L0354, 33F25D0483, 33F25E0729, 33F25H0534, 33F25J0404, 33F24F0070; ASK-45703-PBSWH2/10801902220353, 10801902193275, 10801902220353, 10801902203158/33F24F0937, 33F24L0437, 33F24L1112, 33F25A0306, 33F25B0659, 33F25E0614, 33F25E0803, 33F25F0080, 33F25H0368, 33F25H0513, 33F25K0153, 33F25M0191, 33F26C0157, 33F24F0660; ASK-45703-PCAM1/10801902193275/33F24L0355, 33F25C0312, 33F25E0016, 33F25E0804, 33F25J0162, 33F26A1447, 33F24K0479; ASK-45703-PCSJ2/10801902193275, 10801902211689, 10801902202861/33F25G0245, 33F25J0165, 33F25K0155, 33F26A1546, 33F26B0853, 33F24D0896; ASK-45703-PDM1/10801902202861, 10801902193275/33F24F0747, 33F25A0268, 33F25B0584, 33F25D0368, 33F25D0604, 33F25E0617, 33F25G0330, 33F25H0612, 33F25J0166, 33F25M0196, 33F26A1460, 33F25B0585; ASK-45703-PECL2/10801902193275/33F25D0369, 33F25D0605, 33F25E0619, 33F25G0242, 33F25J0168, 33F26A0013; ASK-45703-PGM4/10801902193275, 10801902221817/33F25B0660, 33F25D0412, 33F25J0169, 33F25K0078, 33F26C0176, 33F24M0174; ASK-45703-PGVH/10801902193275, 10801902200508/33F25E0310, 33F25H0781, 33F24F0101; ASK-45703-PHF4/)10801902200508, 10801902193275/33F24F0887, 33F24L0253, 33F24L0439, 33F25A0269, 33F25B0586, 33F25C0314, 33F25D0606, 33F25E0575, 33F25F0118, 33F25M0060, 33F25M0170, 33F26A1480, 33F26B0877, 33F25B0587; ASK-45703-PHUMC1/10801902193275, 10801902201017/33F25K0047, 33F25L0137, 33F26C0116, 33F25B0588; ASK-45703-PJHH1/10801902214833, 10801902193275/33F24D0929, 33F24F0765, 33F24L0254, 33F25B0162, 33F25B0589, 33F25E0013, 33F25E0023, 33F25E0622, 33F25G0223, 33F25J0171, 33F26A1461, 33F26B0798, 33F26C0195, 33F25A0271; ASK-45703-PMEM/10801902142174, 10801902193275/33F26C0051, 33F25B0591; ASK-45703-PMHC1/10801902193275, 10801902135237/33F25D0485, 33F25F0146, 33F25H0562, 33F25K0405, 33F26A1462, 33F26B0871, 33F25A0272; ASK-45703-PNHS1/10801902193275, 10801902193275/33F25C0317, 33F25E0732, 33F25G0293, 33F25H0649, 33F26D0067; ASK-45703-PNO2/10801902193275, 10801902205220, 10801902212082/33F25E0685, 33F25H0536, 33F25K0073, 33F25K0157, 33F26A1373, 33F26B0154, 33F26C0006, 33F24F0731; ASK-45703-POSU2/10801902193275/33F24K0404, 33F24L0868, 33F25B0457, 33F25B0594, 33F25D0760; ASK-45703-PPHPB1/10801902143850, 10801902193275/33F24F0721, 33F24K0379, 33F24L0444, 33F24L0849, 33F25B0595, 33F25D0487, 33F25E0805, 33F25H0496, 33F25J0299, 33F25K0158, 33F25E0444; ASK-45703-PRGK1/10801902193275, 10801902136654/33F25E0577, 33F25E0806, 33F25G0267, 33F25G0353, 33F25J0300, 33F26A1439, 33F26B0865, 33F24L0869; ASK-45703-PSAU1/10801902193275, 10801902159462/33F25L0262, 33F26C0049, 33F24H0106; ASK-45703-PSCH1/10801902193275/33F25B0241, 33F25E0010, 33F25H0777, 33F24M0274; ASK-45703-PSFR1/10801902193275/33F25H0451, 33F25H0654, 33F25M0145, 33F26A1540, 33F25C0483; ASK-45703-PSHS2/10801902193275/33F25E0808, 33F25H0517, 33F24L0873; ASK-45703-PSJR1/10801902193275, 10801902221053, 10801902125207/33F25D0371, 33F25E0625, 33F25G0198, 33F25M0162, 33F26C0128, 33F24F0626; ASK-45703-PSSM1/10801902193275, 10801902119749/33F24K0571, 33F25A0022, 33F25A0432, 33F25B0231, 33F25B0436, 33F25C0656, 33F25G0556, 33F26A0245, 33F24D0917; ASK-45703-PUCH2/10801902219029, 10801902193275, 10801902135046/33F24F0960, 33F24L0447, 33F24L1029, 33F25C0362, 33F25E0734, 33F25H0661, 33F25M0171, 33F26A1527, 33F26B0763, 33F24D0665; ASK-45703-PUCI1/10801902193275, 10801902135046/33F24J0172, 33F24L0363, 33F25B0460, 33F25C0321, 33F25E0626, 33F25F0205, 33F25J0304, 33F25M0187, 33F26C0013, 33F24L0971; ASK-45703-PUHC2/10801902193275, 10801902159615, 10801902118674/33F25A0277, 33F25E0627, 33F25G0302, 33F25J0232, 33F25M0119, 33F26B0128, 33F26B0879, 33F24D0730; ASK-45703-PUMC/10801902193275, 10801902113228/33F25C0322, 33F25E0688, 33F25H0648, 33F25M0122, 33F26B0130, 33F25F0376; ASK-45703-PUMH/10801902193275, 10801902129489, 10801902156355/33F25A0010, 33F25D0029, 33F25G0009, 33F25K0055, 33F25L0257, 33F26C0118, 33F24D0694; ASK-45703-PUN1/10801902156355, 10801902193275, 10801902164268/33F24F0769, 33F24L0948, 33F25B0462, 33F25C0323, 33F25E0689, 33F25M0131, 33F26C0130, 33F24D0696; ASK-45703-PUNMH/10801902193275, 10801902144161, 10801902159554/33F24L0264, 33F24L0366, 33F25B0599, 33F25D0374, 33F25E0691, 33F25G0305, 33F25J0406, 33F25K0161, 33F25M0222, 33F26B0716, 33F26B0870, 33F24D0848; ASK-45703-PUPM1/10801902159554, 10801902193275, 10801902159554, 10801902225181/33F24F0689, 33F24L0265, 33F24L1128, 33F25A0278, 33F25B0600, 33F25C0324, 33F25E0629, 33F25E0869, 33F25G0408, 33F25H0633, 33S25E0869,, 33F26B0365; ASK-45703-PUWH/10801902193275, 10801902112948/33F24M0046, 33F25A0023, 33F25A0429, 33F25E1174, 33F25G0266, 33F25H0779, 33F25K0049, 33F25L0131, 33F26A1132, 33F24D0310; ASK-45703-PVCU1/10801902112948, 10801902193275/33F24D0869, 33F24L0451, 33F24L0924, 33F25B0602, 33F25D0552, 33F25F0131, 33F25G0304, 33F25J0305, 33F26B0797, 33F25B0463; ASK-45703-PWH1/10801902213904/33F24E0561; ASK-45703-PYNH1/10801902193275/33F24G0519, 33F24K0268, 33F25C0184, 33F25C0909, 33F25E0244, 33F25L0336, 33F25M0843, 33F25G0261; ASK-45703-RR/10801902206944, 10801902193275/33F24F0961, 33F24K0447, 33F24L0452, 33F24L0879, 33F25B0603, 33F25D0608, 33F25E0871, 33F25K0087, 33F25M0395, 33F26A1425, 33F26A1426, 33F26C0695, 33F25B0604; ASK-45703-SHUF2/10801902193275, 10801902208351, 10801902202366/33F25E0631, 33F25F0206, 33F25G0396, 33F25J0174, 33F26B0778, 33F24F0771; ASK-45703-SL1/10801902193275, 10801902202366, 10801902203127/33F24L0453, 33F25B0464, 33F25D0413, 33F25E0809, 33F25H0538, 33F25J0409, 33F26A1410, 33F26B0173, 33F24D0583; ASK-45703-TC/10801902203127, 10801902193275/33F24F0877, 33F24L0454, 33F25B0146, 33F25H0539, 33F26B0186, 33F24L0163; ASK-45854-BJC1/10801902193275, 10801902202557/33F24L0164, 33F24L1019, 33F25B0605, 33F25E0810, 33F26C0137, 33F25B0465; ASK-45854-LMDTG1/10801902193275, 10801902205763/33F25D0376, 33F25D0662, 33F25E0735, 33F25H0535, 33F25J0411, 33F26A1409, 33F26C0126, 33F24H0216; ASK-45854-MCY1/10801902193275/33F24K0494, 33F25L0145; ASK-45854-PBMH/10801902193275/33F25M0676, 33F24L0456; ASK-45854-PBSWH2/10801902193275, 10801902220360/33F25A0307, 33F25C0364, 33F25E0736, 33F25H0512, 33F25J0412, 33F26B0788, 33F25B0606; ASK-45854-PECL2/10801902193275, 10801902224269/33F25K0056, 33F25M0678, 33F26B0227; ASK-45854-PNHS1/10801902193275, 10801902225471/33F25C0326, 33F25D0489, 33F25J0235, 33F26A1351, 33F26D0065; ASK-45854-PPHPB1/10801902193275/33F24H0170, 33F24L0370, 33F25B0607, 33F25C0475, 33F25E0692, 33F25H0581, 33F25J0306, 33F25D0031; ASK-45854-PSHS2/10801902193275/33F25F0371, 33F25K0062, 33F25M0679; ASK-45854-PUN1/10801902193275, 10801902159578/33F25H0484, 33F24D0717; ASK-45854-PUPM1/10801902159578, 10801902193275, 10801902225167/33F24E0665, 33F24F0898, 33F24L1007,33F25B0163, 33F25B0608, 33F25D0414, 33F25E0811, 33F25H0559, 33F25M0011, 33F26B0366; ASK-45854-PUR1/10801902193275/33F24M0178, 33F25B0429, 33F25E1200, 33F25K0057, 33F26A0250, 33F26A1394; ASK-46702-PJHH1/10801902193275, 10801902214888, 10801902116700/33F24L0835, 33F25D0490, 33F25J0307, 33F26C0351, 33F24F1020; CDC-42854-XPCN1A/10801902162134, 10801902193275/33F24F0702, 33F24K0345, 33F24L0130,33F24L0387,33F24L1147, 33F25B0631, 33F25C0334, 33F25D0565, 33F25E0703, 33F25E0897, 33F25G0344, 33F25J0254, 33F26A1476, 33F26B0897, 33F26C0067, 33F25B0534; CDC-42854-XPN1A/10801902193275, 10801902162141/33F25C0335, 33F25D0566, 33F25D0673, 33F26B0810, 33F26B0944, 33F24J0098; CDC-45854-XPCN1A/10801902162158, 10801902193275/33F24E0224, 33F24F0217, 33F24F0245, 33F24G0110, 33F24H0071, 33F24H0430, 33F24J0227, 33F24M0059, 33F24M0117, 33F25A0370, 33F25B0320, 33F25E1128, 33F25F0014, 33F25G0068, 33F25H0291, 33F25J0620, 33F25J0977, 33F25L0089, 33F26A1621, 33F26A1631, 33F26B0625, 33F25A0319; CDC-45854-XPN1A/10801902193275, 10801902162165/33F25B0633, 33F25C0336, 33F25D0572, 33F25E0758, 33F25G0259, 33F25G0342, 33F25J0331, 33F26A0419, 33F26A1569, 33F26C0215, 33F25B0634; AK-42703-CDC/10801902193275/33F25C0802, 33F25H0042, 33F25J0849, 33F25L0415, 33F25B0538; AK-45703-ACDC/10801902167825, 10801902193275, 10801902167825/33F24C0123, 33F24E0379, 33F24G0382, 33F24J0536, 33F24K0144, 33F24L0547, 33F24L1161, 33F25A0490, 33F25C0824, 33F25E0093, 33F25E0949, 33F25G0756, 33F25H0043, 33F25J0489, 33F25L0418, 33F25M0677, 33F26A1054, 33F26A1256, 33F26B0561, 33F26B0976, 33F25B0513; CDC-42703-P1A/10801902170863, 10801902193275/33F24D0984,33F24E0263, 33F24E0840, 33F24F0157, 33F24F0174, 33F24F0234, 33F24H0403, 33F24H0437, 33F24H0459, 33F24J0255, 33F24K0045, 33F24M0042, 33F24M0078, 33F24M0136, 33F25A0386, 33F25B0048, 33F25B0120, 33F25C0220, 33F25C0580, 33F25D0737, 33F25E1104, 33F25G0438, 33F25G0661, 33F25G0764, 33F25H0211, 33F25H0337, 33F25H0819, 33F25J0965, 33F25L0013, 33F25M0513, 33F26B0641, 33F26B1041, 33F26C0322, 33S25G0438, 33F24M0049; CDC-42703-XP1A/10801902193275, 20801902117087, 10801902162080/33F25K0249, 33F25K0386, 33F25L0091, 33F25L0227, 33F25L0271, 33F25L0301, 33F25M0514, 33F25M0526, 33F25M0546, 33F25M0573, 36F25B0028, 36F25B0066, 36F25D0016, 36F25D0017, 33F24C1215; CDC-42703-XPCN1A/10801902162080, 10801902193275/33F24C1227, 33F24E0237, 33F24E0837, 33F24F0001, 33F24F0161, 33F24F0242,33F24F0243,33F24G0209,33F24G0210, 33F24G0211, 33F24G0212, 33F24H0336, 33F24H0362, 33F24H0404, 33F24J0197, 33F24J0222, 33F24J0259, 33F24K0006, 33F24K0654, 33F24M0023, 33F24M0024, 33F24M0025, 33F24M0058, 33F24M0093, 33F24M0138, 33F24M0151, 33F24M0198, 33F25B0277, 33F25B0376, 33F25B0377, 33F25D0063, 33F25E0797, 33F25E0798, 33F25E1188, 33F25F0067, 33F25F0068, 33F25H0882, 33F25J0680, 33F25J0692, 33F25J0696, 33F25J0701,33F25J0704, 33F25J0709, 33F25J0713, 33F25J0714, 33F25J0716, 33F25J0720, 33F25J0727, 33F25J0731, 33F25K0320, 33F25K0387, 33F25K0388, 33F25K0389, 33F25L0165, 33F25L0183, 33F25L0197, 33F25L0280, 33F25L0294, 33F25M0015, 33F25M0018, 33F25M0024, 33F25M0027, 33F25M0036, 33F25M0574, 33F26A1664, 33F26B0626, 33F26B0627, 33F26C0430, 33F26C0431, 33F26C0432, 36F25B0067, 36F25C0003, 36F25C0004, 36F25D0018, 36F25D0019,36F25D0020, 36F25D0029,36F25E0002, 36F25E0003, 36F25E0004, 36F25E0013, 36F25E0027, 36F25G0001, 36F25G0004, 36F25G0005, 36F25G0008, 36F25G0014, 36F25H0020, 36F25H0021, 36F25H0022, 36F25J0004, 36F25J0006, 33F25F0069; CDC-42802-XPCN1A/10801902162103, 10801902193275/33F24D0791, 33F24F0096, 33F24F0768, 33F24K0367, 33F24L0126, 33F24L0383, 33F24L0893, 33F25A0317, 33F25B0533, 33F25D0563, 33F25D0616, 33F25D0637, 33F25E0702, 33F25E0828, 33F25E0895, 33F25F0220, 33F25H0628, 33F25J0438, 33F26A1567, 33F26C0551, 33F25C0246; CDC-42802-XPN1A/10801902193275, 10801902162097, 10801902170887/33F25D0617, 33F25D0671, 33F25E0896, 33F26A1568, 33F26B0811, 33F26B0945, 33F24D0858; CDC-42854-P1A/10801902170887, 10801902193275, 10801902170887/33F24F0749, 33F24L0128, 33F24L0385, 33F25C0371, 33F25D0672, 33F25E0753, 33F25G0300, 33F25H0548, 33F25J0182, 33F25K0189, 33F25M0081, 33F26A1474, 33F26B0709, 33F26B0806, 33F25H0877; CDC-45703-P1A/10801902171525/33F24D0988, 33F24E0239, 33F24E0252, 33F24E0841, 33F24F0164, 33F24F0225, 33F24G0208, 33F24G0253, 33F24H0079, 33F24H0308, 33F24H0337, 33F24J0225, 33F24J0262, 33F24K0049, 33F24K0690, 33F24L0015, 33F24M0033, 33F24M0034, 33F24M0080, 33F24M0094, 33F24M0140, 33F24M0152, 33F25A0369, 33F25A0387, 33F25B0230, 33F25B0300, 33F25B0345, 33F25C0534, 33F25D0725, 33F25E0419, 33F25E1127, 33F25E1157, 33F25F0026, 33F25F0053, 33F25G0036, 33F25G0055, 33F25G0517, 33F25H0364, 33F25H0784, 33F25J0591, 33F25K0244, 33F25L0053, 33F25L0235, 33F24H0376; CDC-45703-XPB1A/10801902193275, 10801902168372/33F25D0377, 33F25H0553, 33F25J0329, 33F26B0885, 33F24D0499; CDC-45703-XPCN1A/10801902161243, 10801902193275, 10801902161243/33F24E0212, 33F24E0216, 33F24E0217, 33F24E0230, 33F24E0231, 33F24E0243, 33F24E0244, 33F24E0255, 33F24E0256, 33F24E0827, 33F24F0237, 33F24F0240, 33F24F0241, 33F24G0214,33F24G0215, 33F24G0216, 33F24G0256, 33F24G0257, 33F24G0258, 33F24H0083, 33F24H0311, 33F24H0312, 33F24H0313, 33F24H0314, 33F24H0340, 33F24H0366, 33F24H0407, 33F24H0440, 33F24H0462, 33F24K0635, 33F24K0636, 33F24K0637, 33F24K0646, 33F24K0647, 33F24K0653, 33F24M0026, 33F24M0027, 33F24M0028, 33F24M0029, 33F24M0030, 33F24M0031, 33F24M0184, 33F24M0195, 33F24M0196, 33F24M0201, 33F24M0202, 33F25C0589, 33F25C0691, 33F25C0716, 33F25C0717, 33F25E0787, 33F25E0789, 33F25E0799, 33F25E0800, 33F25E1181, 33F25E1182, 33F25E1183, 33F25E1185, 33F25E1186, 33F25E1187, 33F25F0061, 33F25F0063, 33F25F0065, 33F25F0329, 33F25F0330, 33F25F0331, 33F25F0343, 33F25F0344, 33F25F0345, 33F25J0086, 33F25J0087, 33F25J0088, 33F25J0290, 33F25J0293, 33F25J0663, 33F25J0674, 33F25J0675, 33F25J0676, 33F25J0677, 33F25J0683, 33F25J0685, 33F25J0686, 33F25J0694, 33F25J0698, 33F25J0699, 33F25J0703, 33F25J0706, 33F25J0707, 33F25J0719, 33F25K0306, 33F25K0307, 33F25K0391, 33F25K0392, 33F25K0393, 33F25L0001, 33F25L0051, 33F25L0052, 33F25L0065, 33F25L0080, 33F25L0087, 33F25L0088, 33F25L0215, 33F25L0247, 33F25L0282, 33F25L0300, 33F25L0321, 33F25L0326, 33F25M0004, 33F25M0022, 33F25M0023, 33F25M0030, 33F25M0038,33F25M0473, 33F25M0516, 33F25M0535, 33F26A0211, 33F26A1640, 33F26A1641, 33F26A1642, 33F26A1643, 33F26B0612, 33F26B0613, 33F26B0614, 33F26B0615, 33F26B0616, 33F26B0617, 33F26B0618, 33F26B0619, 33F26C0321, 33F26C0543, 33F26C0553, 33F26C0564, 33F26C0568, 33F26C0844, 33F26C0861, 33F26C0877, 33F26C0878, 33F26C0879, 33F26C0880, 33R24M0184, 33X25H0001, 33X25K0005, 36F25C0001, 36F25C0002, 36F25C0009, 36F25C0012, 36F25D0007, 36F25D0012, 36F25D0024, 36F25D0025, 36F25D0026, 36F25D0036, 36F25D0037, 36F25E0005, 36F25E0006, 36F25E0010, 36F25E0011, 36F25E0024, 36F25E0025, 36F25G0006, 36F25G0007, 36F25G0009, 36F25G0010, 36F25G0016, 36F25G0017, 36F25G0018, 36F25H0023, 36F25H0024, 33F25E0795; CDC-45703-XPN1A/10801902193275, 10801902162042/33F25F0333, 33F25F0338, 33F25J0294, 33F25J0673, 33F25J0684, 33F25J0780, 33F25L0249, 33F25M0684,33F25M0775, 33F26A0311, 33F26B0608, 33F26B0609, 33F26B0610, 33F26B0611, 33F26B0685, 33F26B1036, 33F26B1055, 33F26B1057, 33F26B1058, 36F25D0027, 36F25D0033, 36F25D0039,36F25D0040, 36F25E0007, 36F25E0012, 36F25E0014, 36F25E0015, 36F25E0022, 36F25E0023, 36F25G0011, 36F25G0015, 36F25G0026, 36F25G0027, 36F25G0028, 36F25H0003, 36F25H0004, 36F25H0005, 36F25H0012, 36F25H0018, 36F25H0019, 36F25H0025, 36F25H0027, 36F25J0001,36F25J0002, 33F25B0505; CDC-45802-XP1A/10801902193275, 10801902162110/33F25D0568, 33F25E0755, 33F25G0275, 33F25H0434, 33F25J0119, 33F25L0107, 33F25M0347, 33F26A1473, 33F24D0771; CDC-45802-XPCN1A/10801902193275, 10801902162110/33F24K0407, 33F24L0495, 33F25B0490, 33F25D0569, 33F25E0756, 33F25G0380, 33F25J0120, 33F26B0890, 33F25A0290; CDC-46702-XP1A/10801902193275, 10801902168389/33F25E0447, 33F25E0706, 33F25J0332, 33F26A1471, 33F26B0218, 33F25B0506; CDC-47702-XP1A/10801902193275, 10801902168396, 20801902118787/33F25D0573, 33F25F0133, 33F25H0441, 33F25J0185, 33F25K0192, 33F25M0409, 33F26B0710, 33F26B0804, 33F26C0779, 33F24A0137; ASK-45703-PTJH/20801902116400/33F25A0016, 33F25C0585;
Devices named in this recall
125 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.
- ARROW IPN925037Teleflex Incorporated · CATHETER, CONDUCTION, ANESTHETIC7 recalls0 reports
- ARROW IPN038522Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days3 recalls1 report
- ARROW IPN915628Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls5 reports
- ARROW IPN915630Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls2 reports
- ARROW IPN904896Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 report
- ARROW IPN055818Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN055820Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN056571Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN917372Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN923102Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN927593Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN037185Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN056182Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN915625Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN056570Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN922680Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN922682Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN922908Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN923103Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN927740Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00178Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00181Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00182Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPNB00180Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN036856Teleflex Incorporated · Syringe, piston2 recalls0 reports
- ARROW IPN915614Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall48 reports
- ARROW IPN037072Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall13 reports
- ARROW IPN915619Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 death9 reports
- ARROW IPN915623Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall8 reports
- ARROW IPN036440Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall6 reports
- ARROW IPN038527Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall4 reports
- ARROW IPN037417Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall3 reports
- ARROW IPN927426Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall3 reports
- ARROW IPN922683Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall2 reports
- ARROW IPN037947Teleflex Incorporated · THERMOMETER, CLINICAL COLOR CHANGE1 recall1 report
- ARROW IPN038109Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN038962Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 report
- ARROW IPN922636Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 report
- ARROW IPN923109Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 report
- ARROW IPN924041Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 report
- ARROW IPN930566Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall1 report
- ARROW IPN037099Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall1 report
- ARROW IPN036334Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036287Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN036525Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036922Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN036860Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037608Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037609Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037948Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN038019Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN037836Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN038112Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall0 reports
- ARROW IPN038528Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN029651Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN029716Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055552Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055726Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055727Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055788Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055807Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055816Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055819Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055801Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN055802Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN056183Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN056267Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN915624Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN915626Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN915629Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN915631Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916999Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916837Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916838Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916839Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917374Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN917148Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN916805Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN921752Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN921951Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922681Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922084Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922109Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922909Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922918Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922983Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922984Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN922986Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN923104Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN923108Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN923112Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN923895Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN924039Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN924040Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN925293Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN926164Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN926160Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN926613Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN926622Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN927374Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 98793).
- Z-2363-2026 · Open, ClassifiedPARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU
- Z-2364-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;…
- Z-2365-2026 · Open, ClassifiedDescription/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU
- Z-2366-2026 · Open, ClassifiedDescription/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM
- Z-2367-2026 · Open, ClassifiedDescription/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) V…
- Z-2368-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018…
- Z-2369-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 …
- Z-2370-2026 · Open, ClassifiedDescription/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT…
- Z-2371-2026 · Open, ClassifiedDescription/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION K…
- Z-2372-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A
- Z-2373-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYS…
- Z-2374-2026 · Open, ClassifiedDescription/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-1…
- Z-2376-2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-…
- Z-2377-2026 · Open, ClassifiedDescription/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE…
- Z-2378-2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR …
- Z-2379-2026 · Open, ClassifiedDescription/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JA…
- Z-2380-2026 · Open, ClassifiedDescription/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000
- Z-2381-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
- Z-2382-2026 · Open, ClassifiedDescription/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP,…
- Z-2383-2026 · Open, ClassifiedDescription/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU
- Z-2385-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH;…
- Z-2387-2026 · Open, ClassifiedDescription/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
- Z-2388-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 C…
- Z-2389-2026 · Open, ClassifiedDescription/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINA…
- Z-2390-2026 · Open, ClassifiedDescription/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B
- Z-2391-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHIL…
- Z-2393-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; …
- Z-2394-2026 · Open, ClassifiedDescription/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MI…
- Z-2396-2026 · Open, ClassifiedDescription/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
- Z-2397-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.