ARROW IPN039773Recalls & reported problems
Teleflex Incorporated · Arrow(R) VPS(R) Stylet Bedside Procedure Kit · Class 2
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10801902145779), received between Jul 13, 2023 and Apr 28, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (4 reports).
4 other recalls cover devices cleared under the same FDA submission (K103255); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings4
- Connection Problem1
- Material Frayed1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2026Material Frayed
IT WAS REPORTED, \"STYLET TIP COMPROMISED, WIRE COMING LOOSE.\"\\GOOD FAITH EFFORTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED DEVICE ISSUE AND TO ASSESS WHETHER ANY PATIENT HARM, INJURY, OR ADVERSE HEALTH CONSEQUENCES OCCURRED. A TOTAL OF THREE DOCUMENTED ATTEMPTS WERE MADE TO CONTACT THE ACCOUNT; HOWEVER, NO RESPONSE OR ADDITIONAL INFORMATION WAS RECEIVED.
Patient problems: Insufficient Information
MAUDE report 25024000 ↗MalfunctionJul 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED "STYLET WAS PLACED IN A BD DOUBLE LUMEN 4FR PICC AND PICC WAS PLACED. AFTER BBE WAS ACHIEVED THEY WERE PULLING THE STYLET OUT OF THE PICC AND THEY NOTICED THE TIP OF THE VPS STYLET WAS SPLIT. THE PATIENT IS FINE AND NOTHING FELL OFF THE STYLET." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396505 ↗MalfunctionNov 28, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT: UNABLE TO GET BLUE BULLS' EYE DESPITE HAVE GOOD WAVE FORMS. THE PATIENT LOST A SMALL AMOUNT OF BLOOD, AND THE PROCEDURE WAS PROLONGED. THE PATIENT WAS REPORTED AS FINE POST THE PROCEDURE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 18216038 ↗MalfunctionNov 28, 2023Incorrect, Inadequate or Imprecise Result or Readings; Connection Problem; Incorrect, Inadequate or Imprecise Result or Readings; Connection Problem
IT WAS REPORTED THAT: LOW AMPLITUDE AND INCONSISTENT DOPPLER WAVEFORM AND UNABLE TO GET BULLSEYE. CHEST X-RAY WAS NEEDED TO CONFIRM PICC POSITION. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18215405 ↗MalfunctionSep 27, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
REPORT FROM AHS CANADA RECEIVED: PATIENT HAD A PICC FROM ANOTHER MANUFACTURER BRAND INSERTED ON JUNE 23RD UTILIZING TELEFLEX VPS G4 TECHNOLOGY. AS PER PICC RECORD "TIP CONFIRMED IN THE LOWER 1/3 SVC/CAJ AS PER G4 TECHNOLOGY." PICC SLUGGISH TO FLUSH WITH NO BLOOD RETURN PRESENT FROM BOTH PICC LUMENS. WITHDRAWAL OCCLUSION PRESENT TO BOTH PICC LUMENS. CHEST X-RAY WAS DONE AND TIP OF PICC WAS MALPOSITIONED AND LOCATED IN THE MID SVC. EXTERNAL LENGTH OF THE PICC HAD NOT CHANGED AND REMAINED THE SAME AS INSERTION. THE TELEFLEX VPS G4 TECHNOLOGY UTILIZED DURING THE INSERTION HAD THE RESULT THAT THE PICC TIP WAS IN THE LOWER 1/3 SVC (BLUE BULLSEYE OBTAINED), BUT THE CHEST XRAY SHOWED THE PICC TIP WAS NOT IN THE LOWER 1/3 SVC BUT IN THE MID SVC. IT WAS REPORTED ALTHOUGH THE MID SVC IS NOT AN OPTIMAL PICC TIP POSITION, IT WAS STILL SAFE TO TREAT THE WITHDRAWAL OCCLUSION WITH CATH-FLO, AS A WITHDRAWAL OCCLUSION CANNOT BE LEFT UNTREATED. A PICC EXCHANGE OR NEW PICC WAS NOT INSERTED AT THIS TIME AS THE PATIENT HAD LIMITED DURATION LEFT FOR HIS ANTIBIOTICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17826361 ↗MalfunctionJul 13, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED PATIENT'S PICC WAS 8CM DEEP IN THE VENTRICLE. THE BULLSEYE SHOWED DESPITE DEEP PLACEMENT. UPON CT SCAN, A CHEST X-RAY WAS ORDERED AND FOUND THE TIP 8CM IN THE VENTRICLE. THE PICC TEAM WAS ASKED TO RETRACT THE PICC 12CMS. NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17321409 ↗
FDA clearance and classification
Cleared through 510(k) K103255 on Feb 18, 2011 (decision: Substantially Equivalent). Submitted by Vasonova, Inc. as “Vascular Positioning System Tm (Vps Tm System) Stylet”. The FDA review took 107 days. FDA record ↗
The FDA regulates product code OBJ (catheter, ultrasound, intravascular) under 21 CFR 870.1200 as a Class II device. 21 CFR 870.1200 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K103255. One submission often covers many models, so check the product and lot details in each record.
- Z-0179-2018 · Jun 1, 2017 · TerminatedArrow(R) VPS(R) Access Kit for use with 4 Fr. Peripherally Inserted Central Venous Catheter The intended use of the VPS Stylet and Console (VPS System) is to quickly and accurately
- Z-2170-2017 · Apr 28, 2017 · TerminatedArrow(R) VPS(R) Access Kit for use with 4 and 5 Fr. Peripherally Inserted Central Venous Catheters
- Z-2489-2015 · Jul 30, 2015 · TerminatedArrow International, Inc. PICC Kits and VPS Access Kits with Touhy Borst Component Catheter, Intravascular, Therapeutic, Ultrasound, Long Term Greater than 30 Days
- Z-2462-2015 · Jul 30, 2015 · TerminatedPICC Kits with Peel-Away Sheath Component; catheter, intravascular, therapeutic, ultrasound Product Usage: catheter, intravascular, therapeutic, ultrasound
Common questions about the ARROW IPN039773
Has the ARROW IPN039773 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10801902145779), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ARROW IPN039773?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jul 13, 2023 and Apr 28, 2026. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (4), “Connection Problem” (1) and “Material Frayed” (1); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN039773?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN039773 FDA approved?
It was cleared through 510(k) K103255 on Feb 18, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the ARROW IPN039773, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN039773). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, ultrasound, intravascular devices
- OptiCross H749518100Boston Scientific Corporation · Catheter, ultrasound, intravascular2 deaths460 reports
- OptiCross™ HD H74939352030Boston Scientific Corporation · Catheter, ultrasound, intravascular1 death444 reports
- ViewFlex™ D087031Irvine Biomedical, Inc. · Catheter, ultrasound, intravascular3 deaths341 reports
- OptiCross 18 H7493932800180Boston Scientific Corporation · Catheter, ultrasound, intravascular186 reports
- OptiCross™ 6 HD H74939354080Boston Scientific Corporation · Catheter, ultrasound, intravascular4 deaths177 reports
- Visions PV .014P RX Digital IVUS Catheter 014RPhilips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall1 death85 reports
- Visions PV .018 Digital IVUS Catheter 86700Philips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall67 reports
- OptiCross™ HD H74939352040Boston Scientific Corporation · Catheter, ultrasound, intravascular3 deaths116 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page