ARROW IPN915317Recalls & reported problems
Teleflex Incorporated · RediGuard(R) IAB · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10801902161892), received between Oct 2, 2022 and Jan 4, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Apr 29, 2024Open, Classified
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
Z-1878-2024 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Connection Problem1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 4, 2023Connection Problem; Connection Problem
THE EVENT OCCURRED ON (B)(6) 2022. IT WAS REPORTED THAT PRIOR TO INSERTION DURING CALIBRATION, "IT DID NOT FIT AND THE CALIBRATION DIDN'T HAPPEN". UPON FURTHER COMMUNICATION ON (B)(6) 2022, THE CUSTOMER STATED THAT THE PATIENT HAD EXPIRED. MULTIPLE INQUIRIES SURROUNDING THE EVENT AND DEATH WERE MADE. ON (B)(6) 2023, THE CUSTOMER STATED "THE DEVICE DIDN'T CAUSE OR CONTRIBUTE TO THE PATIENT'S DEATH. WE WILL NOT ANSWER OTHER QUESTIONS DUE TO THE PROTECTION OF PATIENT AND DOCTOR DATA". IF TELEFLEX RECEIVES ADDITIONAL INFORMATION ON THIS EVENT, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16095818 ↗MalfunctionJan 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
REPORTED AS "IAB RUPTURE/ABRASION CATHETER REMOVED". AS REPORTED BY THE CLINICAL SUPPORT SPECIALIST: "EMAILED RECEIVED 27DEC2022 TO REPORT INCIDENT ON (B)(6) 2022 11:20AM. RT CALLED TO CONFIRM BLOOD SPECS IN HELIUM LINE. TUBING CLAMPED AND CALL PLACED TO [PHYSICIAN]. [PHYSICIAN] REMOVED CATHETER AT (B)(6) 2022 AT 1:56 PM. HEMATOMA AT GROIN SITE. NO DIFFICULTY NOTED ON REMOVAL, NO COMPLICATIONS. PATIENT CURRENTLY CRITICAL DUE TO NATURE OF ILLNESS AND COMPLICATED COURSE (NOT DUE TO IAB REMOVAL). NO IAB CATHETER REPLACED. NO BLOOD NOTED IN PUMP OR CONDENSATION TUBING, PUMP SENT TO BIOMED FOR INSPECTION, CATHETER RETAINED FOR RETURN."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16086865 ↗MalfunctionOct 26, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PATIENT WAS RECEIVED VIA TRANSFER FROM AN OUTSIDE FACILITY, (B)(6), THERE WERE RUSTY COLORED FLECKS OF BLOOD NOTED IN THE GAS LINE AND THE FLECKS ARE ON THE INSIDE OF THE TUBING BUT THERE WERE NO ALARMS. NO ALARM STRIPS WERE PROVIDED. THE IAB WAS REMOVED AND REPLACED IN THE SAME SITE. THERE WERE NO COMPLICATIONS WITH REMOVAL OR REINSERTION. PATIENT IS CURRENTLY STABLE ON THE PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15673590 ↗MalfunctionOct 2, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE BALLOON WAS DEFECTIVE; IT WOULD NOT PUMP/WORK. IT WAS USED AS INDICATED AND THE REPORTED ISSUE OCCURRED RIGHT AT THE FIRST INFLATION OF THE CATHETER. THERE WERE NO ISSUES AT THE INSERTION SITE. THE CATHETER WAS REPLACED WITH A NEW ONE THAT WORKED AS INTENDED. NO PUMP STRIPS WERE GENERATED FOR REVIEW. THERE WAS NO MEDICAL INTERVENTION OR NEGATIVE PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15526208 ↗
Other System, balloon, intra-aortic and control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.