Skip to content

ARROW recallZ-1878-2024 · ARROW INTERNATIONAL Inc. · initiated Apr 29, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ARROW INTERNATIONAL Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Teleflex/ARROW INTERNATIONAL issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 04/29/2024 via FedEx 2-day mail. 1. Prior to use, ensure that a backup IAB catheter is available. 2. Prior to insertion, inspect all in-scope IAB catheters for signs of an over-twisted balloon wrap or bent balloon shaft. Do not insert catheters suspected of manifesting an over-twisted wrap. 3. Insert all in-scope IAB catheters under fluoroscopic guidance. Use fluoroscopy to assess complete balloon inflation. Span the full length of the balloon, include several cycles of inflation/deflation. Keep fluoroscopy available for up to 3 minutes after one-to-one cyclic counterpulsation begins, or until full balloon inflation is confirmed. 4. Be vigilant in responding to pump alarms consistent with the Instructions For Use and Users Guide (IFU). . when an Arrow AC3 or Autocat-2 pump is used, a high-pressure or high baseline alarm may sound. These red-condition alarms require immediate attention per the IFU. The alarms provide a troubleshooting algorithm that includes incomplete IAB inflation. . when a Getinge CardioSave pump is used, an IAB catheter restriction alarm may sound. . A pump alarm indicating helium leakage (helium loss or gas loss) may signal the IAB catheter is not performing as expected. . Real-time telephone support for Arrow IABPs and IAB catheters is available. 5. Be vigilant in responding to bedside indicators that an IAB catheter is not performing as expected, including Blood in helium pathway and Lower than expected diastolic pressure augmentation. Immediate actions If an IAB catheter is suspected of having this issue: remove and replace the catheter using steps outlined in the IFU. A replacement balloon catheter may be inserted contralaterally or ipsilaterally as determined by the responsible physician after considering the patient s individual femoral access options and relevant clinical features.

Recall details

Recall number
Z-1878-2024
Status
Open, Classified
Initiated by the company
Apr 29, 2024
Posted by the FDA
Jun 3, 2024
Recalling firm
ARROW INTERNATIONAL Inc.
Quantity
7939 units
Product code
DSP
FDA submissions
K200634
Distribution
Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.

Product description

Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

a) REF IAB-05830-LWS, UDI: (01)10801902161892(17)241231(10)18F23A0038, (01)10801902161892(17)250131(10)18F23B0023, (01)10801902161892(17)250430(10)18F23D0024, (01)10801902161892(17)250731(10)18F23H0060, (01)10801902161892(17)250930(10)18F23J0075, (01)10801902161892(17)250930(10)18F23K0013, (01)10801902161892(17)251031(10)18F23L0019, (01)10801902161892(17)251031(10)18F23L0041, (01)10801902161892(17)260131(10)18F24B0008, (01)10801902161892(17)260228(10)18F24C0058, (01)10801902161892(17)260228(10)18F24C0101, (01)10801902161892(17)260331(10)18F24D0002, (01)10801902172843(17)250228(10)18F23C0026, (01)10801902172843(17)250531(10)18F23F0049, (01)10801902172843(17)250831(10)18F23J0001, (01)10801902172843(17)251031(10)18F23L0054, (01)10801902172843(17)251130(10)18F23M0056, (01)10801902172843(17)251231(10)18F24A0015, (01)10801902172843(17)260131(10)18F24A0040, (01)10801902172843(17)260131(10)18F24B0013, (01)10801902172843(17)260228(10)18F24C0061, (01)10801902172843(17)260228(10)18F24C0109, (01)10801902172843(17)260228(10)18F24C0110, (01)10801902172843(17)260331(10)18F24C0172, (01)10801902172843(17)260331(10)18F24C0173, (01)10801902172843(17)46112(10)18F24C0088, (01)10801902172843(17)46112(10)18F24C0089, (01)10801902172843(17)46112(10)18F24C0174, (01)10801902172843(17)46112(10)18F24C0175, (01)10801902172843(17)46112(10)18F24C0176, (01)10801902172843(17)46112(10)18F24C0185; b) REF IAB-05840-LWS, UDI: (01)10801902167733(17)240531(10)18F22F0072, (01)10801902167733(17)240630(10)18F22G0021, (01)10801902167733(17)240630(10)18F22G0031, (01)10801902167733(17)240630(10)18F22G0046, (01)10801902167733(17)240731(10)18F22H0009, (01)10801902167733(17)240731(10)18F22H0023, (01)10801902167733(17)240731(10)18F22H0058, (01)10801902167733(17)240930(10)18F22K0015, (01)10801902167733(17)241130(10)18F22M0017, (01)10801902167733(17)241130(10)18F22M0018, (01)10801902167733(17)241231(10)18F23A0004, (01)10801902167733(17)241231(10)18F23A0005, (01)10801902167733(17)241231(10)18F23A0018, (01)10801902167733(17)241231(10)18F23A0026, (01)10801902167733(17)250131(10)18F23B0008, (01)10801902167733(17)250131(10)18F23B0016, (01)10801902167733(17)250531(10)18F23E0054, (01)10801902167733(17)250630(10)18F23G0017, (01)10801902167733(17)250731(10)18F23H0029, (01)10801902167733(17)250731(10)18F23H0064, (01)10801902167733(17)251031(10)18F23L0009, (01)10801902167733(17)251031(10)18F23L0028, (01)10801902167733(17)251231(10)18F23M0068, (01)10801902167733(17)251231(10)18F24A0030, (01)10801902167733(17)251231(10)18F24A0039, (01)10801902167733(17)260131(10)18F24B0015, (01)10801902167733(17)260131(10)18F24B0031, (01)10801902167733(17)260228(10)18F24C0008, (01)10801902167733(17)260228(10)18F24C0069, (01)10801902167733(17)260228(10)18F24C0071, (01)10801902167733(17)260228(10)18F24C0072, (01)10801902167733(17)260228(10)18F24C0075, (01)10801902167733(17)260228(10)18F24C0076, (01)10801902167733(17)260228(10)18F24C0077, (01)10801902167733(17)260228(10)18F24C0078, (01)10801902167733(17)260228(10)18F24C0079, (01)10801902167733(17)260228(10)18F24C0080, (01)10801902167733(17)260228(10)18F24C0081, (01)10801902167733(17)260331(10)18F24C0083, (01)10801902167733(17)260331(10)18F24C0186, (01)10801902167733(17)260331(10)18F24C0187, (01)10801902167733(17)260331(10)18F24C0188, (01)10801902167733(17)260331(10)18F24C0193, (01)10801902167733(17)260331(10)18F24D0020, (01)10801902167733(17)260331(10)18F24D0021, (01)10801902167733(17)260331(10)18F24D0022, (01)10801902167733(17)260331(10)18F24D0031, (01)10801902167733(17)260331(10)18F24D0032, (01)10801902167733(17)260331(10)18F24D0037, (01)10801902167733(17)260331(10)18F24D0038, (01)10801902167733(17)260331(10)18F24D0039, (01)10801902167733(17)260331(10)18F24D0042, (01)10801902167733(17)260331(10)18F24D0044, (01)10801902167733(17)260331(10)18F24D0045, (01)10801902167733(17)260331(10)18F24D0049, (01)10801902167733(17)260331(10)18F24D0050, (01)10801902167733(17)260331(10)18F24D0051, (01)10801902167733(17)260331(10)18F24D0054, (01)10801902167733(17)260331(10)18F24D0055, (01)10801902167733(17)260331(10)18F24D0072, (01)10801902167733(17)260331(10)18F24D0087, (01)10801902167733(17)260331(10)18F24D0088, (01)10801902167733(17)260331(10)18F24D0089, (01)10801902167733(17)260331(10)18F24D0090, (01)10801902167733(17)260331(10)18F24D0091, (01)10801902167733(17)260331(10)18F24D0092, (01)10801902172867(17)240430(10)18F22E0008, (01)10801902172867(17)240430(10)18F22E0029, (01)10801902172867(17)240531(10)18F22F0006, (01)10801902172867(17)240531(10)18F22F0018, (01)10801902172867(17)240531(10)18F22F0039, (01)10801902172867(17)240630(10)18F22G0002, (01)10801902172867(17)240630(10)18F22G0003, (01)10801902172867(17)240630(10)18F22G0019, (01)10801902172867(17)240630(10)18F22G0059, (01)10801902172867(17)240731(10)18F22H0035, (01)10801902172867(17)240831(10)18F22J0025, (01)10801902172867(17)240930(10)18F22K0016, (01)10801902172867(17)250430(10)18F23E0020, (01)10801902172867(17)250531(10)18F23F0002, (01)10801902172867(17)250531(10)18F23F0026, (01)10801902172867(17)250630(10)18F23G0008, (01)10801902172867(17)250731(10)18F23G0057, (01)10801902172867(17)250731(10)18F23H0013, (01)10801902172867(17)250731(10)18F23H0055, (01)10801902172867(17)250831(10)18F23J0003, (01)10801902172867(17)250831(10)18F23J0037, (01)10801902172867(17)250831(10)18F23J0072, (01)10801902172867(17)250930(10)18F23J0073, (01)10801902172867(17)251031(10)18F23K0054, (01)10801902172867(17)251130(10)18F23L0055, (01)10801902172867(17)251130(10)18F23M0043, (01)10801902172867(17)251231(10)18F24A0025, (01)10801902172867(17)251231(10)18F24A0031, (01)10801902172867(17)260131(10)18F24B0005, (01)10801902172867(17)260131(10)18F24B0048, (01)10801902172867(17)260131(10)18F24B0052, (01)10801902172867(17)260228(10)18F24B0078, (01)10801902172867(17)260228(10)18F24C0036, (01)10801902172867(17)260228(10)18F24C0064, (01)10801902172867(17)260228(10)18F24C0065, (01)10801902172867(17)260228(10)18F24C0067, (01)10801902172867(17)260228(10)18F24C0068, (01)10801902172867(17)260228(10)18F24C0105; c) REF IAB-05850-LWS, UDI: (01)10801902167726(17)240831(10)18F22J0037(13)220927, (01)10801902167726(17)240831(10)18F22J0040(13)220929, (01)10801902167726(17)240930(10)18F22K0044(13)221021, (01)10801902167726(17)240930(10)18F22K0045(13)221021, (01)10801902167726(17)241031(10)18F22K0056(13)221028, (01)10801902167726(17)241031(10)18F22L0012(13)221104, (01)10801902167726(17)250930(10)18F23K0006(13)231031, (01)10801902167726(17)260131(10)18F24B0049(13)240227, (01)10801902167726(17)260228(10)18F24B0075(13)240301, (01)10801902172881(17)250930(10)18F23K0038(13)231028, (01)10801902172881(17)260228(10)18F24C0031(13)240316, (01)10801902172881(17)260228(10)18F24C0032(13)240331, (01)10801902172881(17)260228(10)18F24C0033(13)240315, (01)10801902172881(17)260228(10)18F24C0034(13)240315

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K200634 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 94526).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page