ARROW IPN920060Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Two-Lumen PICC with 80 cm Wire · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10801902192643), received between Apr 11, 2023 and Sep 29, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K113277); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Obstruction of Flow2
- Complete Blockage1
- Crack1
- Deformation Due to Compressive Stress1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Complete Blockage
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933190 ↗MalfunctionSep 6, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50638652 ↗MalfunctionJun 9, 2026Obstruction of Flow
IT WAS REPORTED THAT THE PATIENT'S PERIPHERALLY INSERTED CENTRAL CATHETER (PICC), ORIGINALLY PLACED ON (B)(6) 2026, BECAME OCCLUDED ON (B)(6) 2026. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS FINE. THE ONLY INTERVENTION REQUIRED WAS REMOVAL OF THE AFFECTED DEVICE AND REPLACEMENT WITH ANOTHER PICC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444661 ↗MalfunctionJun 9, 2026Obstruction of Flow
IT WAS REPORTED THAT THE PATIENT'S PERIPHERALLY INSERTED CENTRAL CATHETER (PICC), ORIGINALLY PLACED ON (B)(6) 2026, BECAME OCCLUDED ON (B)(6) 2026. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS FINE. THE ONLY INTERVENTION REQUIRED WAS REMOVAL OF THE AFFECTED DEVICE AND REPLACEMENT WITH ANOTHER PICC. THE REPORTER ADDITIONALLY NOTED A TREND AT THE SITE, INDICATING AN INCREASED NUMBER OF PICC OCCLUSIONS DESPITE ADHERENCE TO ROUTINE CARE AND MAINTENANCE PRACTICES. NO SPECIFIC PATIENT INJURY OR ADVERSE OUTCOMES WERE REPORTED IN ASSOCIATION WITH THESE ADDITIONAL EVENTS; MANAGEMENT FOR THOSE CASES SIMILARLY INVOLVED DEVICE REMOVAL AND REPLACEMENT. ASSOCIATED MDRS INCLUDE 9680794-2026-00446 (MULTIPLE EVENTS WITHOUT ADDITIONAL PATIENT-SPECIFIC DETAILS).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25444458 ↗MalfunctionApr 16, 2026Crack
IT WAS REPORTED "PICC LINE INSERTED- PATIENT WENT FOR A CONTRACT INJECTION, INFILTRATION OF CONTRAST AT PICC INSERTION SITE- PICC RE MOVED-LOOKED LIKE THERE WAS A CRACK IN THE PICC LINE AT CM 40CM, 5ML/PER SEC INJECTION OMIPICC 350 INJECTION." THE PICC LINE WAS EXCHANGED. THE PATIENT'S CURRENT CONDITION IS UNKNOWN AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 24903322 ↗MalfunctionApr 11, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
PHYSICIAN FOUND THE GUIDEWIRE KINKED AND BENT DURING USE. IT WAS REPORTED THERE WAS RESISTANCE WITH THE NEEDLE AND THE WIRE WAS FRAYED UPON REMOVAL. A SECOND INSERTION WITH A NEW CATHETER KIT WAS SUCCESSFUL. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16715893 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113277. One submission often covers many models, so check the product and lot details in each record.
- Z-2378-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR
- Z-1254-2020 · Dec 20, 2019 · TerminatedPICC Set- PR-35052-HPHNM
- Z-2023-2018 · Apr 11, 2018 · TerminatedARROW PICC Set; Product codes: PR-34052-HPHNM, PR-34063-HPHNM, PR-35041-HPHNM, PR-35052-HPHNM, PR-35541-HPHNL, PR-35541-HPHNM, PR-35552-HPHNL, PR-35552-HPHNM
- Z-0602-2018 · Jul 25, 2017 · TerminatedArrowADVANTAGE 5 Pressure Injectable Peripherally Inserted Central Catheter (PICC). Product Codes: PR-35041-HPHNM and CDA-35041-HPK1A.
- Z-0182-2018 · Jun 1, 2017 · TerminatedArrowADVANTAGE5" Pressure Injectable Peripherally Inserted Central Catheter (PICC) The Pressure Injectable PICC is indicated for short or long term peripheral access to the central
- Z-0207-2016 · Oct 16, 2015 · TerminatedARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page