Skip to content

ARROW IPN920476Recalls & reported problems

Teleflex Incorporated · Arrow-Clark(TM) VectorFlow(TM) Antegrade Hemodialysis Catheter · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902195149), received between Apr 2, 2026 and Apr 2, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2026Open, Classified

    16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

    Z-2157-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026Leak/Splash

    IT WAS REPORTED THAT " DR (NEPHROLOGIST) INSERTED THE SAID UNIT IN LEFT IJV OF PATIENT AS PER ROUTINE PROCEDURE IN OT. AFTER PROCEDURE, DR NOTICED AIR BUBBLES ALONG WITH BLOOD WHILE ASPIRATING TO CHECK THE FLOW. CATHETER WAS REMOVED AND REPLACED WITH ACUTE HD CATH. THERE WAS NO REPORTED PATIENT HARM. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765415 ↗

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.