ARROW recallZ-2157-2026 · Arrow International, LLC · initiated Apr 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 4/10/2026, recall notices were mailed to customers who were informed of the following Arrow International LLC is recalling product due to the Merit Medical Systems, Inc. recall of the Merit 16F Dual-Valved Splitable Sheath Introducer component, which is included in impacted Arrow Hemodialysis Kits & Sets. Customers were asked to do the following: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Complete and return the Acknowledgement Form, see link: https://fca.myteleflex.com/en/recall/000605 3) This notice should be passed on to all persons who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. In addition, distributors were asked to do the following: - Provide a copy of this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. Should you require further information or support concerning this issue, please contact Customer Service via email ([email protected]), phone (1-866-396-2111), or FAX (1-855-419-8507).
Recall details
- Recall number
- Z-2157-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 10, 2026
- Posted by the FDA
- May 13, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 92,827
- Product code
- MSD
- FDA submissions
- K141051
- Distribution
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
Product description
Arrow-Clark VectorFlow Antegrade Chronic Hemodialysis Catheter: REF: ACS-15192-VFI, ACS-15232-VFI, ACS-15272-VFI, CS-15192-VFI, CS-15192-VFIE, CS-15232-VFI, CS-15232-VFIE, CS-15272-VFI, CS-15272-VFIE, CS-15312-VFI, CS-15312-VFIE, CS-15422-VFI, CS-15422-VFIE, CS-15552-VFI, CS-15552-VFIE; Arrow-Clark VectorFlow Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: REF: CS-15192-VFIM, CS-15232-VFIM, CS-15272-VFIM; Arrow-Clark VectorFlow Retrograde Chronic Hemodialysis Catheter, REF: ACS-15192-VF, ACS-15232-VF, ACS-15272-VF, ACS-15552-VF, CS-15192-VF, CS-15192-VFE, CS-15232-VF, CS-15232-VFE, CS-15272-VF, CS-15272-VFE, CS-15312-VF, CS-15312-VFE, CS-15422-VF, CS-15422-VFE, CS-15552-VF, CS-15552-VFE; Arrow-Clark VectorFlow Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer: CS-15192-VFM, CS-15232-VFM, CS-15272-VFM
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
REF/UDI-DI/Lots: ACS-15192-VFI/ 10801902126709/33F23G0040, 33F25C0689; ACS-15232-VFI/10801902126723/33F23G0045, 33F25C0678; ACS-15272-VFI/ 10801902126747/33F24G0638; CS-15192-VFI/10801902127188, 10801902193572/33F23E0579, 33F23G0042, 33F23K0170, 33F23L0933, 33F24E0271, 33F24G0166, 33F24J0086, 33F24L0064, 33F24M0224, 33F25C0652, 33F25D0052, 33F25D0700, 33F25E1081, 33F25F0015, 33F25F0388, 33F25G0702, 33F25H0785; CS-15192-VFIE/10801902199659/33F23D0853, 33F23E0773, 33F23G0044, 33F23G0532, 33F23H0611, 33F23K0169,33F23K1105,33F23L0934, 33F24A0912, 33F24A0913, 33F24C0058, 33F24D1001, 33F24G0167,33F24G0187, 33F24J0030, 33F24J0032, 33F24J0339, 33F24J0381, 33F24K0659, 33F24L0058, 33F24L0511, 33F24M0155, 33F24M0215, 33F24M0265, 33F25A0143, 33F25A0407, 33F25B0259, 33F25C0037, 33F25D0285, 33F25D0738, 33F25F0060, 33F25G0635; CS-15232-VFI/10801902194463/33F23C0864, 33F23E0578, 33F23E0832, 33F23G0535, 33F23G0840, 33F23H0612, 33F23K0144, 33F24A0914, 33F24C0079, 33F24E0269, 33F24J0042, 33F25B0256, 33F25C0014, 33F25C0709, 33F25C0650, 33F25E1085, 33F25E1130, 33F25F0009, 33F25G0023, 33F25G0450, 33F25H0341, 33F25J1014; CS-15232-VFIE/10801902195149/33F23D0855, 33F23E0774, 33F23G0047, 33F23G0048, 33F23G0049, 33F23K0145, 33F23K0175, 33F23L0938, 33F24A0915, 33F24C0069, 33F24G0169, 33F24J0026, 33F24J0027, 33F24J0028, 33F24J0084, 33F24K0700, 33F24L0178, 33F24M0005, 33F24M0214, 33F24M0219, 33F25C0646, 33F25C0665, 33F25D0097, 33F25E0358, 33F25E0434, 33F25F0012, 33F25F0275, 33F25F0328, 33F25G0069, 33F25H0293, 33F25H0822, 33F25K0235; CS-15272-VFI/10801902193640, 10801902127362/33F23E0871, 33F23F0742, 33F23H0981, 33F23K0181, 33F24A0306, 33F24C0084, 33F24J0062, 33F25D0098, 33F25E0307, 33F25E0332, 33F25E0354, 33F25F0058, 33F25F0245, 33F25J0659; CS-15272-VFIE/10801902195255/33F23E0886, 33F23G0537, 33F23H0982, 33F23K0180, 33F23L0946, 33F24A0922, 33F24C0065, 33F24D1011, 33F24E0686, 33F24E0805, 33F24K0508, 33F24M0223, 33F25A0410, 33F25C0009, 33F25G0717; CS-15312-VFI/10801902197778/33F23G0543, 33F23L0953, 33F24D1007, 33F24E0810, 33F24J0048, 33F25C0039, 33F25D0218, 33F25E0470, 33F25J0021; CS-15312-VFIE/10801902195491/33F23K0183, 33F23L0954, 33F24A0923, 33F24D1013, 33F25A0403, 33F25F0534, 33F25G0722; CS-15422-VFI/10801902193664/33F23G0549, 33F24E0815, 33F25E1074; CS-15422-VFIE/10801902197440/33F23G0550, 33F23K0178, 33F23L0958, 33F24A0919, 33F24C0073, 33F24E0816, 33F25F0303; CS-15552-VFI/10801902193671/33F23K0184, 33F24J0076, 33F25E1075; CS-15552-VFIE/10801902195163/33F23E0865, 33F23F0743, 33F24D0972, 33F24E0819, 33F24G0098, 33F24K0656, 33F25A0032; CS-15192-VFIM/10801902193589/33F24E0792, 33F25C0676, 33F25L0035; CS-15232-VFIM/10801902193619/33F23C0879, 33F23L0939, 33F24A0304, 33F24E0798, 33F24J0047; CS-15272-VFIM/10801902197761/33F23H1027; ACS-15192-VF/10801902126693/33F24G0637; ACS-15232-VF/10801902126716/33F23H1016; ACS-15272-VF/10801902126730/33F23H1024; ACS-15552-VF/10801902126792/33F23H1021; CS-15192-VF/20801902193593/33F23K0186, 33F23L0931, 33F24C0082, 33F24E0788, 33F24J0054, 33F24M0162, 33F24M0230, 33F25D0271; CS-15192-VFE/10801902195460/33F23E0776, 33F23F0728, 33F23G0531, 33F24A0917, 33F24C0063, 33F24D1003, 33F24E0789, 33F24J0039, 33F24L0069, 33F24M0156, 33F24M0159, 33F25C0683, 33F25D0107, 33F25D0220, 33F25E0343, 33F25F0273, 33F25F0462, 33F25G0444, 33F25H0820, 33F25K0384; CS-15232-VF/10801902193626/33F23C0873, 33F23J0546, 33F24A0294, 33F24C0071, 33R24G0168, 33F24J0044, 33F25D0089, 33F25E0333, 33F25E0427, 33F25E1095, 33F25F0404, 33F25H0294, 33F25J1009; CS-15232-VFE/10801902195132/33F23E0775, 33F23F0720, 33F23K0176, 33F23K0194, 33F23L0937, 33F24A0927, 33F24C0059, 33F24D1010, 33F24E0270, 33F24E0796, 33F24G0045, 33F24J0035, 33F24J0083, 33F24L0171, 33F24M0216, 33F25A0004, 33F25A0083, 33F25C0023, 33F25C0659, 33F25D0106, 33F25D0219, 33F25D0726, 33F25E0478, 33F25E1105, 33F25F0298, 33F25J1056, 33F25K0335, 33F25L0011; CS-15272-VF/10801902193657/33F23C0880, 33F23F0740, 33F23L0944, 33F24C0081, 33F24E0804, 33F24J0051, 33F24J0686, 33F24M0228, 33F25C0668, 33F25E0337, 33F25F0403, 33F25J0590; CS-15272-VFE/10801902195248/33F23E0777, 33F23F0732, 33F23H0980, 33F23K0172, 33F23L0945, 33F24A0918, 33F24C0064, 33F24H0044, 33F24H0255, 33F24M0237, 33F25C0049, 33F25E0349, 33F25E1132, 33F25G0637; CS-15312-VF/10801902127447/33F24G0060, 33F25D0216; CS-15312-VFE/10801902195484/33F23F0734, 33F23G0542, 33F23L0952, 33F24A0925, 33F24C0076, 33F24J0053, 33F24M0225, 33F25A0033, 33F25C0667, 33F25E0320, 33F25E0469, 33F25G0603; CS-15422-VF/10801902128666/33F23H1020, 33F25G0734; CS-15422-VFE/10801902197341/33F23G0548, 33F23H0986, 33F23L0957, 33F24A0929, 33F24J0483, 33F24M0229, 33F25C0681, 33F25E0440, 33F25F0250, 33F25G0484; CS-15552-VF/10801902128703/33F23H1012, 33F25C0674, 33F25J0588; CS-15552-VFE/10801902195156/33F23D0868, 33F23G0261, 33F23G0551, 33F23K0173, 33F24A0928, 33F24D1014, 33F24E0818, 33F24J0059, 33F24K0511, 33F25C0670, 33F25E0471, 33F25F0455, 33F25G0447; CS-15192-VFM/10801902193602/33F23K0189, 33F24C0078, 33F24J0049, 33F24M0243, 33F25C0672, 33F25E1077, 33F25F0418; CS-15232-VFM/10801902193633/33F23J0547, 33F24A0289, 33F24C0077, 33F24E0799, 33F24J0045, 33F24J0093, 33F25A0026, 33F25C0050, 33F25E0348, 33F25K0348; CS-15272-VFM/10801902127393/33F25C0687
Devices named in this recall
24 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN920475Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall6 reports
- ARROW IPN920509Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall5 reports
- ARROW IPN920486Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall2 reports
- ARROW IPN001349Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall1 report
- ARROW IPN920476Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall1 report
- ARROW IPN920508Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall1 report
- ARROW IPN001300Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001301Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001302Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001303Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001304Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001305Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001310Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001367Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001370Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001375Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001386Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN001390Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920136Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920137Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920139Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920783Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920784Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
- ARROW IPN920138Teleflex Incorporated · Catheter, hemodialysis, implanted1 recall0 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under K141051 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARROW IPN001356Teleflex Incorporated · Catheter, hemodialysis, implanted34 reports
- ARROW IPN001348Teleflex Incorporated · Catheter, hemodialysis, implanted23 reports
- ARROW IPN001350Teleflex Incorporated · Catheter, hemodialysis, implanted17 reports
- ARROW IPN001179Teleflex Incorporated · Catheter, hemodialysis, implanted15 reports
- ARROW IPN001366Teleflex Incorporated · Catheter, hemodialysis, implanted12 reports
- ARROW IPN001172Teleflex Incorporated · Catheter, hemodialysis, implanted6 reports
- ARROW IPN055267Teleflex Incorporated · Catheter, hemodialysis, implanted4 reports
- ARROW IPN001181Teleflex Incorporated · Catheter, hemodialysis, implanted3 reports
- ARROW IPN035671Teleflex Incorporated · Catheter, hemodialysis, implanted2 reports
- ARROW IPN039803Teleflex Incorporated · Catheter, hemodialysis, implanted2 reports
- ARROW CS-15192-VFETeleflex Incorporated · Catheter, hemodialysis, implanted2 reports
- ARROW IPN035653Teleflex Incorporated · Catheter, hemodialysis, implanted1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98745).
- Z-2156-2026 · Open, ClassifiedArrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
- Z-2158-2026 · Open, ClassifiedCannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter w…
- Z-2159-2026 · Open, ClassifiedNextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep A…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page