1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10801902195248), received between Dec 13, 2022 and Dec 15, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Apr 10, 2026Class IOpen, Classified

    16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.

    Z-2157-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device1
  • Deformation Due to Compressive Stress1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 15, 2022Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THE CATHETER EMPTIED DESPITE THE CLAMPING. THERE WAS NO CLINICAL CONSEQUENCE FOR THE PATIENT. NO MEDICAL INTERVENTION WAS REQUIRED. IT WAS REPORTED "THE GREEN JOINT WAS PROPERLY CONNECTED".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15992231 ↗
  • MalfunctionDec 13, 2022Deformation Due to Compressive Stress; Physical Resistance/Sticking; Deformation Due to Compressive Stress; Physical Resistance/Sticking

    IT WAS REPORTED THAT "DURING THE INSERTION OF THE INTRODUCER, THEY ENCOUNTERED RESISTANCE TO INTRODUCTION. AS THEY PULLED OUT THE ENTIRE DEVICE, THEY NOTICED THAT THE GUIDE-WIRE WAS BENT." NO PATIENT HARM WAS REPORTED. A NEW DEVICE WAS USED. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15974440 ↗

FDA clearance and classification

Cleared through Special 510(k) K141051 on Jul 7, 2014 (decision: Substantially Equivalent). Submitted by Arrow International, Inc. as “Vectorflow; Antegrade Chronic Hemodialysis Catheter, Retrograde Chronic Hemodialysis Catheter, Retrograde Replacement Hu”. The FDA review took 74 days. FDA record ↗

The FDA regulates product code MSD (catheter, hemodialysis, implanted) under 21 CFR 876.5540 as a Class II device. 21 CFR 876.5540 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Common questions about the ARROW IPN920486

Has the ARROW IPN920486 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2157-2026 (Class I), began on Apr 10, 2026: 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ARROW IPN920486?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 13, 2022 and Dec 15, 2022. In those reports, the most common specific device problems are “Air/Gas in Device” (1), “Deformation Due to Compressive Stress” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ARROW IPN920486?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ARROW IPN920486 FDA approved?

It was cleared through Special 510(k) K141051 on Jul 7, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the ARROW IPN920486 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the ARROW IPN920486, and is it still sold?

The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN920486). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page