ARROW IPN920509Recalls & reported problems
Teleflex Incorporated · Arrow-Clark(TM) VectorFlow(TM) Antegrade Hemodialysis Catheter · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10801902199659), received between Apr 2, 2026 and Apr 22, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 10, 2026Open, Classified
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
Z-2157-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Material Deformation4
- Device Misassembled During Manufacturing /Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Device Misassembled During Manufacturing /Shipping
IT WAS REPORTED THAT: "THE MATERIAL NUMBER ON THE SECONDARY PACKAGING WAS DIFFERENT FROM PRIMARY PACKAGING AND THE PRODUCT WAS DIFFERENT THAN WHAT WAS REGULARLY BEEN USED BY THE DOCTOR. SO THE CLINICIAN HAD TO USE ANOTHER TO COMPLETE PERMCATH INSERTION PROCEDURE FOR PATIENT UNDERGOING DIALYSIS.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954540 ↗MalfunctionApr 20, 2026Material Deformation
IT WAS REPORTED THAT: "A LONG-TERM DIALYSIS CATHETER WAS USED INTRAOPERATIVELY TO ESTABLISH ACCESS FOR BLOOD PURIFICATION THERAPY. DURING THE PROCEDURE THE OUTER LAYER OF THE PEEL-AWAY SHEATH TORE, AND THE SHEATH COULD NOT BE INSERTED. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24923599 ↗MalfunctionApr 2, 2026Material Deformation
IT WAS REPORTED THAT" ON (B)(6) 2026, THE DOCTOR FEEDBACK AFTER A PERIOD OF USE THE EXTENSION LINE WAS FOUND DEFORMED DURING USED ON THE PATIENT. " ASSOCIATED COMPLAINTS 9680794-2026-00237, 9680794-2026-00236.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24763263 ↗MalfunctionApr 2, 2026Material Deformation
IT WAS REPORTED THAT" ON (B)(6) 2026, THE DOCTOR FEEDBACK AFTER A PERIOD OF USE THE EXTENSION LINE WAS FOUND DEFORMED DURING USED ON THE PATIENT. " ASSOCIATED COMPLAINTS 9680794-2026-00237 AND 9680794-2026-00238.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24763260 ↗MalfunctionApr 2, 2026Material Deformation
IT WAS REPORTED THAT"06 MAR 2026, THE DOCTOR FEEDBACK AFTER A PERIOD OF USE THE EXTENSION LINE WAS FOUND DEFORMED DURING USED ON THE PATIENT. " ASSOCIATED COMPLAINTS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24763242 ↗
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.