Skip to content

ARROW IPN922859Recalls & reported problems

Teleflex Incorporated · QuickFlash(R) Radial Artery Catheterization Set · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10801902202984), received between Jul 15, 2026 and Sep 17, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

32 other recalls cover devices cleared under the same FDA submission (K810675); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2024Open, Classified

    Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.

    Z-1267-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Difficult to Advance1
  • Material Deformation1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Difficult to Advance; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812477 ↗
  • MalfunctionJul 15, 2026Unraveled Material

    IT WAS REPORTED THAT DURING THE INITIAL INSERTION OF AN ARTERIAL CATHETER UNDER ULTRASOUND GUIDANCE IN THE RADIAL ARTERY, THE SPRING WIRE GUIDE (SWG) WAS OBSERVED TO BE UNCOILED UPON REMOVAL FROM THE CATHETER. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THE EVENT. THE PATIENT'S CONDITION WAS REPORTED AS UNCHANGED ("STATUS QUO"). THE AFFECTED DEVICE WAS REMOVED AND REPLACED WITH A NEW DEVICE, AND NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. IT WAS FURTHER REPORTED THAT THE ARTERY DID NOT CONTAIN SCAR TISSUE, PLAQUE, OR OTHER RIGID ANATOMICAL STRUCTURES, SUCH AS BONE OR MUSCLE MASS, THAT COULD HAVE INTERFERED WITH ARTERIAL PLACEMENT OR CONTRIBUTED TO THE REPORTED DEVICE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25829022 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K810675. One submission often covers many models, so check the product and lot details in each record.

  • Z-2397-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18
  • Z-3299-2024 · Jul 26, 2024 · Open, Classified
    Arrow Arterial Kits and Trays as follows: (1) Arrow Arterial Access Tray with 0.025" Spring-Wire Guide (Catheter Not Included), REF ASK-04001-BW1; (2) Arrow Arterial Access Tray
  • Z-0844-2020 · Dec 12, 2019 · Open, Classified
    Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04
  • Z-1968-2018 · Apr 11, 2018 · Terminated
    Radial Arterial Access Kit; Product Code: ASK-04001-BID2
  • Z-1970-2018 · Apr 11, 2018 · Terminated
    Radial Artery Catheterization Kit; Product Codes: ASK-04020-AH, ASK-04020-MIHS, ASK-04120-HF1
  • Z-1972-2018 · Apr 11, 2018 · Terminated
    Radial Artery Catheterization Kit with Sharps Safety Features; Product Codes: ASK-04020-RIHM
  • Z-1969-2018 · Apr 11, 2018 · Terminated
    Arterial Line Kit with Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-04018-OH1
  • Z-1971-2018 · Apr 11, 2018 · Terminated
    Arterial Catheterization Kit; Product Codes: ASK-04020-DH
  • Z-2019-2018 · Apr 11, 2018 · Terminated
    Arterial Catheterization Kit; Product Code:NA-04550-1A
  • Z-1330-2018 · Dec 22, 2017 · Terminated
    (1) Arterial Catheterization Kit, Part Number (PN): ASK-04018-CC; (2) Arterial Line Kit with Sharps Safety Features, PN: ASK-04510-HUM; (3) Arterial Line Kit with Sharps Safety Fea
  • Z-1281-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04018
  • Z-1283-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04020-C
  • Z-1285-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04122
  • Z-1286-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04220
  • Z-1282-2018 · Oct 26, 2017 · Terminated
    Radial Artery Catheterization Kit; Model: AK-04020

…and 17 more.

Other Wire, guide, catheter devices

All Wire, guide, catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page