ARROW IPN923012Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Advance(R) Two-Lumen Midline · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10801902205428), received between Jul 13, 2023 and Jul 1, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K963257); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2394-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation2
- Material Separation2
- Difficult or Delayed Separation1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2024Material Separation; Material Separation
IT WAS REPORTED THAT: "THE PEEL AWAY WINGS BROKE OFF DURING INSERTION MAKING IT VERY DIFFICULT TO GRIP THE REMAINING PART OF PEEL AWAY SHEATH AND REMOVE. COMPLETED THE PROCEDURE WITH THIS DIFFICULTY. NO ADVERSE EFFECTS TO PATIENT OR MIDLINE PERFORMANCE." ASSOCIATED COMPLAINTS 9680794-2024-00639 AND 9680794-2024-00638.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19642799 ↗MalfunctionMay 24, 2024Material Deformation; Difficult or Delayed Separation; Material Deformation; Difficult or Delayed Separation
IT WAS REPORTED THAT THE "SHEATH BROKE DURING INSERTION AND THEY HAD TO OPEN A STANDALONE". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19391004 ↗MalfunctionDec 15, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT: THE MIDLINE WAS BEING USED IN THE HOSPITAL FOR ONE DAY WHEN IT STARTED LEAKING. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18333753 ↗MalfunctionDec 6, 2023Material Separation; Material Separation
IT WAS REPORTED THAT: THE HANDLE OF THE PEEL AWAY SHEATH BROKE OFF. THE REPORTED DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED. ASSOCIATED TO 9680794-2023-00929.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18273125 ↗MalfunctionJul 13, 2023Material Deformation; Material Deformation
CUSTOMER REPORTED KINKING OF INTRODUCER/GLIDE THRU SHEATH. NO PATIENT HARM WAS REPORTED. A NEW SHEATH WAS USED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17321676 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K963257. One submission often covers many models, so check the product and lot details in each record.
- Z-2385-2018 · May 17, 2018 · TerminatedPeripherally Inserted Midline Catheter Kit with Placement Wire, REF MK-02031-TW Product Usage: To provide real time tip location information of a central venous catheter by utiliz
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page