ARROW IPN924163Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Two-Lumen PICC pre-loaded with ARROW(R) VPS Precision(TM) Stylet · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10801902208672), received between Sep 7, 2023 and Jul 9, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K113277); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2394-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Material Twisted/Bent3
- Material Separation2
- Break1
- Fluid/Blood Leak1
- Material Too Soft/Flexible1
- Obstruction of Flow1
- Product Quality Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED "PICC WAS INSERTED IN MARCH FOR CHEMO. THE PATIENT WAS HAVING CHEMO ON (B)(6) 2025 AND BLOOD STARTED DRIPPING OUT OF THE EXTENSION LINE OF THE WHITE PORT." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22457224 ↗MalfunctionJul 3, 2025Obstruction of Flow; Material Twisted/Bent; Obstruction of Flow; Material Twisted/Bent
IT WAS REPORTED "PICC WAS PLACED ON (B)(6) 2025. THE NURSES WEREN'T ABLE TO ASPIRATE OR FLUSH ON (B)(6) 2025 SO THEY HAD TO REMOVE THE LINE AND SAW HOW THE CATHETER WAS KINKED IN 2 PLACES, WHICH IS THE LOCATION OF WHERE THE KINK WOULD HAVE BEEN IN THE ARMPIT ACCORDING TO INSERTER." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396509 ↗MalfunctionJun 20, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "PICC WAS INSERTED ON (B)(6) 2025 AS AN OUTPATIENT. HOME HEALTH SAID FROM THE BEGINNING THE PICC WAS SLUGGISH. THEY GAVE THE PATIENT TPA MULTIPLE TIMES AND IT WOULD HELP THEN THIS PAST WEEK IT WOULDN'T HELP AT ALL SO PATIENT CAME TO ED TO GET IT REPLACED. WHEN THEY PULLED THE PICC OUT THEY SAW THE PICC WAS KINKED IN MULTIPLE PLACES." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22298416 ↗MalfunctionAug 30, 2024Material Too Soft/Flexible; Material Too Soft/Flexible
IT WAS REPORTED THAT: "PICC'S ARE TOO SOFT AND WE ARE SEEING MORE KINKING. THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE." ASSOCIATED COMPLAINTS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20106607 ↗MalfunctionAug 28, 2024Product Quality Problem; Material Twisted/Bent; Product Quality Problem; Material Twisted/Bent
IT WAS REPORTED THAT: "PICC WAS PLACED AND 2 DAYS LATER IT WAS KINKED IN MULTIPLE PLACES. A NEW PICC WAS NEEDED TO BE PLACED. THE THERAPY WAS DELAYED, INTERRUPTED. THE ACCOUNT IS SAYING THE PICC IS TOO SOFT." THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE. ASSOCIATED COMPLAINTS 9680794-2024-00911 AND 9680794-2024-00811.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20083746 ↗MalfunctionJul 15, 2024Material Separation; Material Separation
IT WAS REPORTED "PATIENT FELL AND PICC HUB BROKE OFF". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19742986 ↗MalfunctionSep 13, 2023Material Separation; Material Separation
IT WAS REPORTED THAT: RIGHT HUB BROKE OFF DURING REMOVAL OF SHEATH OUTSIDE PATIENT. THE DEFECT WAS DETECTED DURING USE ON PATIENT. THE PATIENT CONDITION WAS REPORTED AS "FINE". NO PATIENT INJURY/CONSEQUENCE OR MEDICAL INTERVENTION REPORTED. ADDITIONAL INFORMATION: THE ENTIRE SHEATH WAS REMOVED. NONE LEFT IN PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17735759 ↗MalfunctionSep 7, 2023Break; Break
IT WAS REPORTED THE SHEATH "BROKE AT THE RIGHT SIDE OF THE HUB OF SHEATH INTRODUCER, APART FROM THE CATHETER". THE INSERTION SITE WAS THE UPPER ARM. THE SHEATH WAS REMOVED IN ITS ENTIRETY, NOTHING WAS LEFT BEHIND. NO PATIENT HARM OR INJURY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17703101 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113277. One submission often covers many models, so check the product and lot details in each record.
- Z-2378-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR
- Z-1254-2020 · Dec 20, 2019 · TerminatedPICC Set- PR-35052-HPHNM
- Z-2023-2018 · Apr 11, 2018 · TerminatedARROW PICC Set; Product codes: PR-34052-HPHNM, PR-34063-HPHNM, PR-35041-HPHNM, PR-35052-HPHNM, PR-35541-HPHNL, PR-35541-HPHNM, PR-35552-HPHNL, PR-35552-HPHNM
- Z-0602-2018 · Jul 25, 2017 · TerminatedArrowADVANTAGE 5 Pressure Injectable Peripherally Inserted Central Catheter (PICC). Product Codes: PR-35041-HPHNM and CDA-35041-HPK1A.
- Z-0182-2018 · Jun 1, 2017 · TerminatedArrowADVANTAGE5" Pressure Injectable Peripherally Inserted Central Catheter (PICC) The Pressure Injectable PICC is indicated for short or long term peripheral access to the central
- Z-0207-2016 · Oct 16, 2015 · TerminatedARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.