ARROW IPN928085Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Advance(R) Two-Lumen PICC Kit pre-loaded with Arrow(R) VPS Precision(TM) Stylet · Class 2 · Implant
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902215151), received between Apr 29, 2025 and Apr 29, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).
11 other recalls cover devices cleared under the same FDA submission (K112896); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole1
Patient problems
- Extravasation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED "WE RAN ACROSS A SITUATION TODAY WHERE A PICC HAD A PRETTY BAD EXTRAVASATION DUE TO A HOLE IN THE PICC." ADDITIONAL INFORMATION RECEIVED STATES "IT SOUNDS LIKE THE 'HOLE' WAS ACTUALLY FOUND/DEVELOPED THIS MORNING WHILE GIVING CONTRAST DOWN IN CT. THEY COULD SEE THE CONTRAST BUILDING UP IN THE ARM SO TERMINATED PROCEDURE. UP UNTIL THAT EVENT THIS MORNING, PICC WAS FUNCTIONING NORMALLY." THE LINE WAS REMOVED. THE PATIENT HAD SWELLING IN THERE ARM; HOWEVER, THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Extravasation
MAUDE report 21925153 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112896. One submission often covers many models, so check the product and lot details in each record.
- Z-2378-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR
- Z-1811-2019 · Apr 23, 2019 · TerminatedArrowg+ard Blue Advance PICC Product Code: CDC45541HPK1A
- Z-1819-2019 · Apr 23, 2019 · TerminatedArrow PICC with Chlorag+ard Technology Product Code: PR 45541 HPHNL
- Z-1820-2019 · Apr 23, 2019 · TerminatedArrow PICC with Chlorag+ard Technology Product Code:PR 45541 HPHNM
- Z-1822-2019 · Apr 23, 2019 · TerminatedArrow PICC with Chlorag+ard Technology Product Code:PR 45552 HPHNM
- Z-1327-2018 · Dec 22, 2017 · TerminatedPressure Injectable ARROW(R) PICC Kit with Chlorag+ard Technology, Part Numbers: (1) ASK-45541-RWJ4; (2) ASK-45552-RWJ4
- Z-0996-2018 · Nov 27, 2017 · TerminatedPressure Injectable Multi- Lumen Central Venous Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology Product Usage: The Arrow CVC is indicated to provide short-ter
- Z-2698-2017 · Jun 12, 2017 · TerminatedArrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology
- Z-2331-2017 · May 9, 2017 · Terminated1) Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology The Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology is indicated for
- Z-1705-2017 · Mar 3, 2017 · TerminatedThe CG+ Arrow PIC Catheters permit venous access to the central circulation through a peripheral vein.
- Z-0207-2016 · Oct 16, 2015 · TerminatedARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page