Skip to content

ARROW IPN928176Recalls & reported problems

Teleflex Incorporated · Pressure Injectable Two-Lumen PICC with 130 cm Wire · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10801902215755), received between Feb 24, 2026 and Feb 24, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).

11 other recalls cover devices cleared under the same FDA submission (K112896); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT "DURING PICC PROCEDURE, HCP FELT RESISTANCE DURING PASSING THROUGH LESION AS WELL AS FLUSHING. DRUG INJECTION WAS NOT SMOOTH. THE PRODUCT WAS REMOVED AND CHANGED SAME MODEL NEW PRODUCT. THE PROCEDURE WAS SUCCEEDED AND NO MORE ADDITIONAL PROBLEMS OCCURRED." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24428566 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112896. One submission often covers many models, so check the product and lot details in each record.

  • Z-2378-2026 · Apr 21, 2026 · Open, Classified
    Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR
  • Z-1822-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code:PR 45552 HPHNM
  • Z-1819-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code: PR 45541 HPHNL
  • Z-1811-2019 · Apr 23, 2019 · Terminated
    Arrowg+ard Blue Advance PICC Product Code: CDC45541HPK1A
  • Z-1820-2019 · Apr 23, 2019 · Terminated
    Arrow PICC with Chlorag+ard Technology Product Code:PR 45541 HPHNM
  • Z-1327-2018 · Dec 22, 2017 · Terminated
    Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard Technology, Part Numbers: (1) ASK-45541-RWJ4; (2) ASK-45552-RWJ4
  • Z-0996-2018 · Nov 27, 2017 · Terminated
    Pressure Injectable Multi- Lumen Central Venous Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology Product Usage: The Arrow CVC is indicated to provide short-ter
  • Z-2698-2017 · Jun 12, 2017 · Terminated
    Arrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology
  • Z-2331-2017 · May 9, 2017 · Terminated
    1) Pressure Injectable ARROW(R) PICC Kit with Chlorag+ard(R) Technology The Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology is indicated for
  • Z-1705-2017 · Mar 3, 2017 · Terminated
    The CG+ Arrow PIC Catheters permit venous access to the central circulation through a peripheral vein.
  • Z-0207-2016 · Oct 16, 2015 · Terminated
    ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow Pressure Injectable Arrow PICC powered by Arrow VPS

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page