Infinity Orange Enteral Feeding Pump INFORNGRFRecalls & reported problems
Zevex International, Inc. · Class 2
All models of Infinity Orange Enteral Feeding Pump (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10814844000341), received between Sep 25, 2024 and Dec 2, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Protective Measures Problem” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Protective Measures Problem11
- Insufficient Flow or Under Infusion1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP WAS NOT DELIVERING FORMULA AND DID NOT ALARM NO FLOW OUT AS EXPECTED. THEY STATED IT FAILED TO ALARM. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, AND THEY STATED THAT THE PATIENT HAD NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THE COMPLAINT. NO ADDITIONAL INFORMATION WAS PROVIDED. [(B)(4)].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23697828 ↗MalfunctionOct 16, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP HINGES WERE CRACKED. WHEN THE PUMP WAS RETURNED TO MMD FOR EVALUATION, THE NO FLOW OUT ALARM DID NOT OPERATE AS EXPECTED. IT FAILED TO ALARM. THE INITIAL REPORTER STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NOT HAD ANY ADVERSE EFFECT ON A PATIENT. THE ALARM FAILURE WAS DISCOVERED AT MOOG. [(B)(4).]
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23316043 ↗MalfunctionJun 18, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP WAS NOT ALARMING NO FLOW OUT AS EXPECTED. THEY STATED THAT IT FAILED TO ALARM. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NO EFFECT ON A PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22286545 ↗MalfunctionJun 4, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP WAS NOT ALARMING LOAD SET AS EXPECTED. THEY STATED THAT IT FAILED TO ALARM. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NO EFFECT ON A PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. [(B)(4)].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22145724 ↗MalfunctionJan 15, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP DID NOT ALARM NO FLOW OUT AS EXPECTED. THEY STATED THAT THE PUMP FAILED TO ALARM. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NO EFFECT ON A PATIENT. NO OTHER INFORMATION WAS PROVIDED. [COMPLAINT: (B)(4)].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21160916 ↗MalfunctionJan 15, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP DID NOT ALARM NO FLOW OUT AS EXPECTED. THEY STATED THAT THE PUMP FAILED TO ALARM. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NO EFFECT ON A PATIENT. NO OTHER INFORMATION WAS PROVIDED. (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21160743 ↗MalfunctionJan 9, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP FRONT HOUSING WAS BROKEN NEAR THE SENSORS. WHEN THE PUMP WAS RETURNED TO MMD FOR EVALUATION, THE NO FLOW IN ALARM DID NOT ALARM AS EXPECTED. IT FAILED TO ALARM. THE INITIAL REPORTER STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NOT HAD ANY ADVERSE EFFECT ON A PATIENT. THE ALARM FAILURE WAS DISCOVERED AT MOOG. [(B)(4)]
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21119718 ↗MalfunctionJan 8, 2025Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP DID NOT ALARM NO FLOW OUT AS EXPECTED. THEY STATED THAT IT FAILED TO ALARM. THE INITIAL REPORTER STATED THAT THE COMPLAINT HAD OCCURRED DURING TESTING AND HAD NOT HAD ANY ADVERSE EFFECT ON A PATIENT. [COMPLAINT: (B)(4)].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107244 ↗MalfunctionDec 18, 2024Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion
THE INITIAL REPORTER STATED THAT THE PUMP UNDER INFUSED AT A RATE THE MMD WOULD CONSIDER REPORTABLE. MMD DID FOLLOW UP WITH THE INITIAL REPORTER, WHO STATED THAT THE COMPLAINT OCCURRED DURING TESTING AND HAD NOT HAD ANY EFFECT ON A PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. [COMPLAINT-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20970839 ↗MalfunctionOct 30, 2024Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP HAD A DAMAGED FRONT HOUSING. WHEN THE PUMP WAS RETURNED TO MMD FOR EVALUATION, THE NO FLOW OUT ALARM DID NOT ALARM AS EXPECTED. IT FAILED TO ALARM. THE INITIAL REPORTER STATED THAT THE COMPLAINT HAD NOT HAD ANY ADVERSE EFFECT ON A PATIENT. THE ALARM FAILURE WAS DISCOVERED AT MOOG. [ (B)(4) ].
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20571559 ↗MalfunctionOct 30, 2024Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP HAD A DAMAGED FRONT HOUSING. WHEN THE PUMP WAS RETURNED TO MMD FOR EVALUATION, THE NO FLOW OUT ALARM DID NOT ALARM AS EXPECTED. IT FAILED TO ALARM. THE INITIAL REPORTER STATED THAT THE COMPLAINT HAD NOT HAD ANY ADVERSE EFFECT ON A PATIENT. THE ALARM FAILURE WAS DISCOVERED AT MOOG. (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20571541 ↗MalfunctionSep 25, 2024Protective Measures Problem; Protective Measures Problem
THE INITIAL REPORTER STATED THAT THE PUMP WAS PUMPING AFTER FOOD WAS GONE AND DID NOT ALARM. THEY STATED THAT THEY DID NOT HAVE ANY INFORMATION ABOUT THE PUMP SETTINGS. MMD DID FOLLOW UP WITH THE INITIAL REPORTER WHO STATED THAT THEY DID NOT HAVE ANY ADDITIONAL INFORMATION ABOUT THE COMPLAINT BECAUSE IT WAS NOT AVAILABLE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT THAT WERE REPORTED. COMPLAINT:(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20304735 ↗
Other Pump, Infusion, Enteral devices
- Kangaroo 775100Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL988 reports
- Kangaroo 382400Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL677 reports
- Kangaroo 383400Cardinal Health, Inc. · Pump, Infusion, Enteral1 death659 reports
- Infinity Enteral Feeding Pump INFKIT2Zevex International, Inc. · Pump, Infusion, Enteral2 deaths407 reports
- Infinity Enteral Feeding Pump INFP01Zevex International, Inc. · Pump, Infusion, Enteral1 death207 reports
- Perfusor® 8713030UB. Braun Medical Inc. · Pump, infusion, enteral2 recalls3 deaths76 reports
- Kangaroo 775659Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL1 death139 reports
- Perfusor® 8713031UB. Braun Medical Inc. · PUMP, INFUSION, ENTERAL1 recall21 deaths87 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page