Nexus TKO® N3561Recalls & reported problems
Nexus Medical, LLC · 10" 3-Up Manifold Set with (3) Nexus TKO®-6P Valves & Pinch Clamp · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10815879020731), received between Feb 6, 2018 and Nov 26, 2019 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Connection Problem1
- Crack1
- Device Dislodged or Dislocated1
- Fluid/Blood Leak1
- Improper Flow or Infusion1
- Leak/Splash1
- Material Separation1
Patient problems
- No Known Impact Or Consequence To Patient2
- Low Pulmonary Arterial Wedge Pressure1
- Missed Dose1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2019Crack; Leak/Splash
IV TUBING INFUSING DRIPS FOR PATIENT. WENT TO ASSESS PATIENT AND FOUND THE DEVICE TO BE CRACKED AND LEAKING FLUIDS WHERE THE MANIFOLD ATTACHES TO THE CAP OF PATIENT'S CENTRAL LINE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9375378 ↗MalfunctionNov 22, 2019Break; Material Separation
IV FLUIDS, VERSED, FENTANYL AND VECURONIUM DRIPS INFUSING VIA MANIFOLD INTO PATIENTS CENTRAL LINE. PLASTIC END PIECE OF MANIFOLD FOUND LYING IN BED BROKEN INTO SEVERAL PIECES. IV FLUIDS AND DRIPS INFUSING INTO THE PATIENT'S BED. SUBSEQUENTLY, INFORMED OF ANOTHER MANIFOLD THAT FELL APART DURING USE BUT THIS ONE WAS NOT SAVED.
Patient problems: Missed Dose
MAUDE report 9362318 ↗MalfunctionFeb 6, 2018Fluid/Blood Leak
MANIFOLD LEAKED AND SOAKED THE BED AFTER REPLACING. PREVIOUS MANIFOLD DID NOT LEAK AND WHEN REPLACED THIS NEXT MANIFOLD DID NOT LEAK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7243797 ↗MalfunctionFeb 6, 2018Connection Problem; Device Dislodged or Dislocated; Improper Flow or Infusion
IV MANIFOLD BROKE WHILE DISCONNECTING MANIFOLD FROM PORT TUBING TO DRAW BLOOD CULTURES. THE ROUND PORTION OF THE MANIFOLD THAT SCREWS INTO THE PORTS TUBING POPPED OFF OF THE REST OF THE MANIFOLD, MAKING IT IMPOSSIBLE TO HOOK THE MANIFOLD BACK UP TO THE PORT TUBING. IN ORDER TO REPLACE THE MANIFOLD, THE NEW MANIFOLD HAD TO BE FLUSHED WITH SALINE. THUS IT TOOK A WHILE FOR THE DRIPS THAT HAD BEEN INFUSING IN THE MANIFOLD (FENTANYL, VERSED, AND EPINEPHRINE) TO REACH THE PATIENT. RESIDENT AND CHARGE RN WERE NOTIFIED OF RESULTING DECREASE IN MEAN ARTERIAL PRESSURES (MAPS) DUE TO DELAY IN EPINEPHRINE REACHING THE PATIENT. PT REMAINED STABLE, MAPS RETURNED TO WITHIN DEFINED LIMITS AFTER ~15 MINUTES OF INFUSION THROUGH THE NEW MANIFOLD. DAMAGED MANIFOLD PUT IN PLASTIC BAGGIE AND LEFT IN SUPERVISOR'S MAILBOX FOR VIEWING.
Patient problems: Low Pulmonary Arterial Wedge Pressure
MAUDE report 7243767 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.