ARROW IPN033806Recalls & reported problems
Teleflex Incorporated · OnControl® Bone Lesion Biopsy System Tray 11ga 10cm / 13ga 14cm · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10816000012380), received between Jan 10, 2022 and Apr 30, 2024 — classified as 3 injuries.
The most frequently reported specific device problem is “Material Separation” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K113872); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Break1
- Positioning Failure1
Patient problems
- Foreign Body In Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 30, 2024Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT: "A PIECE OF THE NEEDLE BROKE OFF IN A PATIENT." ADDITIONAL INFORMATION: "THE PROCEDURE WAS A BONE LESION BIOPSY OF THE RIGHT ILIAC CREST. WE RECEIVED NEW BIOPSY GUNS WITHIN THE LAST 6-8 MONTHS. A PIECE OF NEEDLE BROKE OFF IN PATIENT'S RIGHT ILIAC CREST. REMOVED REMAINING PRODUCT FROM PATIENT, REPORTED INCIDENT TO HOSPITAL AND CONTACTED VENDOR." THE PATIENT WAS REPORTED AS FINE/STABLE.
Patient problems: Foreign Body In Patient
MAUDE report 19215149 ↗InjurySep 27, 2023Material Separation; Material Separation
IT WAS REPORTED THAT: THE ACCESS NEEDLE BROKE IN THE PATIENT'S PELVIS DURING THE BIOPSY AND HAD TO BE REMOVED SURGICALLY. THE PROCEDURE WAS SUCCESSFUL, AND THE ENTIRETY OF THE NEEDLE WAS RECOVERED. THE PATIENT IS FINE.
Patient problems: Foreign Body In Patient
MAUDE report 17823857 ↗InjuryJan 10, 2022Positioning Failure; Positioning Failure
AS REPORTED: THE USER APPROACHED THE BONE TUMOR TISSUE WITH THE ON CONTROL ACCESS NEEDLE. HOWEVER, THE NEEDLE ROTATION WAS SLOW ON CT, AND THE PUNCTURE DID NOT PROCEED. THEN, THE NEEDLE BEGAN TO HEAT UP SO THAT THE USER STOPPED USING IT. THE USER CHANGED TO A MANUAL BIOPSY AND CONTINUED THE APPROACH, BUT AS A RESULT, THE PROCEDURE ITSELF WAS COMPLETED. LATER, WHEN THE USER VERIFIED THE CT IMAGE, IT WAS FOUND THAT THE TIP OF THE NEEDLE WAS DULL. THEREFORE, THE USER WAS WORRIED THAT THE TIP OF THE NEEDLE MIGHT HAVE COME OFF DURING THE OPERATION AND THAT THERE WERE NO REMNANTS IN THE PATIENT. ADDITIONAL INFORMATION RECEIVED 20DEC2021: BY RECONFIRMING CT FLUOROSCOPY IMAGE BY THE USER, IT WAS FOUND THAT NO FRAGMENT WAS LEFT IN THE PATIENT, AND THE NEEDLE TIP WAS NOT WORN. THE OSTEOSCLEROTIC LESION WAS TOO HARD FOR THE ACCESS NEEDLE TO REACH NEAR THE LESION. FROM THE IMAGE FINDINGS, IT WAS FOUND THAT THE ROTATION OF THE DRIVER WAS NOT SUFFICIENTLY TRANSMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13204460 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113872. One submission often covers many models, so check the product and lot details in each record.
- Z-1374-2021 · Mar 8, 2021 · Open, ClassifiedArrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-VC-006 (US) Product Code: 9464-EU-001 (sold in European Regions only) - Pr
- Z-0677-2017 · Nov 8, 2016 · TerminatedARROW OnControl Bone Lesion Biopsy System Tray, Sterile, Rx only, The Arrow OnControl Bone Lesion Biopsy System is intended for bone biopsy of the vertebral body and bone lesions.
Other INSTRUMENT, BIOPSY devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page