Teleflex Incorporated · OnControl® Bone Lesion Biopsy System Tray 11ga 10cm / 13ga 14cm · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10816000012380), received between Jan 10, 2022 and Apr 30, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Material Separation” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K113872); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Break1
  • Positioning Failure1

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 30, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT: "A PIECE OF THE NEEDLE BROKE OFF IN A PATIENT." ADDITIONAL INFORMATION: "THE PROCEDURE WAS A BONE LESION BIOPSY OF THE RIGHT ILIAC CREST. WE RECEIVED NEW BIOPSY GUNS WITHIN THE LAST 6-8 MONTHS. A PIECE OF NEEDLE BROKE OFF IN PATIENT'S RIGHT ILIAC CREST. REMOVED REMAINING PRODUCT FROM PATIENT, REPORTED INCIDENT TO HOSPITAL AND CONTACTED VENDOR." THE PATIENT WAS REPORTED AS FINE/STABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 19215149 ↗
  • InjurySep 27, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT: THE ACCESS NEEDLE BROKE IN THE PATIENT'S PELVIS DURING THE BIOPSY AND HAD TO BE REMOVED SURGICALLY. THE PROCEDURE WAS SUCCESSFUL, AND THE ENTIRETY OF THE NEEDLE WAS RECOVERED. THE PATIENT IS FINE.

    Patient problems: Foreign Body In Patient

    MAUDE report 17823857 ↗
  • InjuryJan 10, 2022Positioning Failure; Positioning Failure

    AS REPORTED: THE USER APPROACHED THE BONE TUMOR TISSUE WITH THE ON CONTROL ACCESS NEEDLE. HOWEVER, THE NEEDLE ROTATION WAS SLOW ON CT, AND THE PUNCTURE DID NOT PROCEED. THEN, THE NEEDLE BEGAN TO HEAT UP SO THAT THE USER STOPPED USING IT. THE USER CHANGED TO A MANUAL BIOPSY AND CONTINUED THE APPROACH, BUT AS A RESULT, THE PROCEDURE ITSELF WAS COMPLETED. LATER, WHEN THE USER VERIFIED THE CT IMAGE, IT WAS FOUND THAT THE TIP OF THE NEEDLE WAS DULL. THEREFORE, THE USER WAS WORRIED THAT THE TIP OF THE NEEDLE MIGHT HAVE COME OFF DURING THE OPERATION AND THAT THERE WERE NO REMNANTS IN THE PATIENT. ADDITIONAL INFORMATION RECEIVED 20DEC2021: BY RECONFIRMING CT FLUOROSCOPY IMAGE BY THE USER, IT WAS FOUND THAT NO FRAGMENT WAS LEFT IN THE PATIENT, AND THE NEEDLE TIP WAS NOT WORN. THE OSTEOSCLEROTIC LESION WAS TOO HARD FOR THE ACCESS NEEDLE TO REACH NEAR THE LESION. FROM THE IMAGE FINDINGS, IT WAS FOUND THAT THE ROTATION OF THE DRIVER WAS NOT SUFFICIENTLY TRANSMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13204460 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113872. One submission often covers many models, so check the product and lot details in each record.

  • Z-1374-2021 · Mar 8, 2021 · Open, Classified
    Arrow OnControl Bone Lesion Biopsy System Tray-11 ga 10 cm Access/13 ga 14 cm Biopsy, Product Code: 9464-VC-006 (US) Product Code: 9464-EU-001 (sold in European Regions only) - Pr
  • Z-0677-2017 · Nov 8, 2016 · Terminated
    ARROW OnControl Bone Lesion Biopsy System Tray, Sterile, Rx only, The Arrow OnControl Bone Lesion Biopsy System is intended for bone biopsy of the vertebral body and bone lesions.

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page