Skip to content

Chlor-XTRA 503850Recalls & reported problems

Vista Dental · Vista’s Chlor-XTRA™ is an enhanced 6% sodium hypochlorite solution designed for irrigation, debridement and cleansing of root canals during and after instrumentation. 12ml pre-filled syringes. · Class U

All models of Chlor-XTRA (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 27, 2019Class IITerminated

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

    Z-0017-2020 · Root cause: Process control · Recall details · FDA record ↗

Other Cleanser, Root Canal devices

All Cleanser, Root Canal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page