Skip to content

Chlor-XTRA recallZ-0017-2020 · Inter-Med Llc · initiated Jun 27, 2019

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Inter-Med Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

FDA root cause category: Process control

What the company is doing

The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.

Recall details

Recall number
Z-0017-2020
Status
Terminated
Initiated by the company
Jun 27, 2019
Posted by the FDA
Oct 1, 2019
Recalling firm
Inter-Med Llc
Quantity
384 units
Product code
KJJ
FDA submissions
K082470
Distribution
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Product description

Vista Dental Products Chlor-XTRA - 12mL Prefilled Syringes, QTY: 8, REF 503850

Code information (models, lots, serial numbers)

Lot/Work Order Numbers: 2018-1471 2018-1992 20182211, UDI (01)10818207020502 20182779, UDI (01)10818207020502 20190041, UDI (01)10818207020502 20190309, UDI (01)10818207020502

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 83435).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page