Zenith Flex G32184Recalls & reported problems
Cook Incorporated · Zenith Flex, AAA Main Body Extension · Class 3 · Implant
All models of Zenith Flex (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002321844), received between Jul 7, 2017 and Oct 5, 2017 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Defective Device1
Patient problems
- No Known Impact Or Consequence To Patient1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 5, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT ENDOVASCULAR ANEURYSM REPAIR (EVAR) AGAINST ABDOMINAL AORTIC ANEURYSM (AAA) AND ANEURYSM IN THE LEFT COMMON ILIAC ARTERY. THE PATIENT ANATOMY WAS SUITABLE FOR THE PROCEDURE. SEVERE CALCIFICATION WAS OBSERVED IN THE ACCESS ROUTE, BUT THE DELIVERY SYSTEM OF THE MAIN BODY PASSED THROUGH THE CALCIFIED AREA WITHOUT PROBLEMS. THE DEVICES BELOW WERE PLACED IN THE TARGET SITE FOLLOWING THE INSTRUCTIONS FOR USE. TFB-32-74-ZT, LOT#7987338 - RIGHT FEMORAL APPROACH, MAIN BODY, ZSLE-13-122-ZT, LOT#5691101 - LEFT FEMORAL APPROACH, CONTRALATERAL LEG, ZSLE-20-56-ZT, LOT#7791648 - RIGHT FEMORAL APPROACH, IPSILATERAL LEG. ANGIOGRAPHY CONFIRMED A PROXIMAL TYPE 1 ENDOLEAK. THE PHYSICIAN DETERMINED TO PLACE A PROXIMAL EXTENSION (ESBE-32-39 LOT#6800050) BY USING THE SAME ACCESS ROUTE, RIGHT FEMORAL APPROACH, USED FOR PLACEMENT OF THE MAIN BODY AND IPSILATERAL LEG. HOWEVER, THE DELIVERY SYSTEM GOT STUCK IN THE CALCIFICATION NEAR THE BIFURCATION OF INTERNAL AND EXTERNAL ILIAC ARTERY. AS THE DELIVERY SYSTEM COULD NOT BE ADVANCED FURTHER, PLACEMENT OF THE EXTENSION WAS GIVEN UP AND TOUCH-UP BALLOONING WAS CONDUCTED. THE AMOUNT OF THE ENDOLEAK FLOW DECREASED BY THE BALLOONING, YET IT
Patient problems: Vascular Dissection
MAUDE report 6915527 ↗MalfunctionJul 7, 2017Defective Device; Defective Device
IT WAS REPORTED THAT A PROBLEM WAS NOTICED WITH A ZENITH AAA ENDOVASCULAR GRAFT AAA ANCILLARY COMPONENT MAIN BODY EXTENSION. PRIOR TO PATIENT CONTACT FOR A PLANNED ENDOVASCULAR ANEURYSM REPAIR PROCEDURE (EVAR) THE SURGEON NOTICED THAT THERE WAS A BARB STICKING THROUGH. CLARIFICATION HAS BEEN REQUESTED BUT NOT YET RECEIVED. THE SHEATH WASN'T SMOOTH AND THEY DIDN'T WISH TO USE THIS ON A PATIENT, SO THEY USED THE SPARE DEVICE. THE PROBLEM WAS DISCOVERED PRIOR TO PATIENT CONTACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6693445 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page