Zenith Flex G48403Recalls & reported problems
Cook Incorporated · Zenith Flex, AAA Endovascular Graft Main Body · Class 3 · Implant
All models of Zenith Flex (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002484037), received between Apr 10, 2015 and Apr 12, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Dissection1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Vascular Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED A TYPE 1A ENDOLEAK OCCURRED ON A ZENITH FLEX AAA ENDOVASCULAR GRAFT BIFURCATED MAIN BODY. DURING THE PROCEDURE, THE BALLOONING WAS PERFORMED AT THE FIRST SEALING STENT. THE ENDOLEAK WAS OBSERVED ON THE FINAL ANGIOGRAM UPON COMPLETION OF THE PROCEDURE. NO ADDITIONAL TREATMENT WAS PERFORMED. THE PHYSICIAN INTENDS TO PERFORM A COMPUTED TOPOGRAPHY ANGIOGRAM IN 30 DAYS TO DETERMINE IF ADDITIONAL PROCEDURES WILL BE REQUIRED. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14091434 ↗InjuryMar 30, 2017Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) USING A ZENITH FLEX AAA ENDOVASCULAR GRAFT BIFURCATED MAIN BODY, A TYPE 3 ENDOLEAK WAS FOUND ON THE IPSILATERAL LIMB OF THE ZENITH FLEX-ZT MAIN BODY AFTER BOTH TFLES WERE PLACED. THE LEAK WAS DISCOVERED TO BE RIGHT ABOVE THE OVERLAPPING OF THE TFLE AND THE IPSILATERAL LIMB. ANOTHER TFLE WAS PLACED TO COVER THE LEAK AND THE TYPE 3 ENDOLEAK WAS CONSIDERED RESOLVED. IT WAS REPORTED THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 6446555 ↗InjuryApr 10, 2015Dissection; Leak/Splash
AN (B)(6) FEMALE PATIENT UNDERWENT AAA REPAIR BY RIGHT FEMORAL APPROACH. THE PATIENT WAS SUITABLE FOR ENDOVASCULAR REPAIR; SACCULAR ANEURYSM OF THE ABDOMINAL AORTA, NO TORTUOSITY, ENOUGH LENGTH OF PROXIMAL AND DISTAL NECK. BOTH CIA IS SHORT, HOWEVER IT WAS SECURED 30 MM FOR BOTH SIDE. AFTER PLACING THE MAINBODY GRAFT AND ILIAC LEGS, A PROXIMAL TYPE I ENDOLEAK WAS CONFIRMED BY ANGIOGRAPHY. AFTER PLACING THE MAINBODY GRAFT AND ILIAC LEGS, A PROXIMAL TYPE I ENDOLEAK WAS CONFIRMED BY ANGIOGRAPHY. THE PHYSICIAN PLANNED TO ADD THE EXTENSION IN THE PROXIMAL NECK OF THE MAIN BODY, AND WHEN HE ATTEMPTED TO ADVANCE THE EXTENSION FROM RIGHT, HE FELT RESISTANCE. THEREFORE, HE REMOVED THE DEVICE FROM THE BODY OF THE PATIENT AND OBSERVED THE ENDOSPORIUM BETWEEN THE DILATOR TIP AND SHEATH. THE PHYSICIAN DETERMINED NOT TO ATTEMPT THE DEVICE FROM RIGHT SIDE BECAUSE OF THE RISK, AND ADVANCE THE DEVICE FROM LEFT THIS TIME, HOWEVER HE FELT RESISTANCE AGAIN. HE CONSIDERED TO REMOVE THE DEVICE BUT COULD NOT, SO HE ADVANCED THE DEVICE AS MUCH AS POSSIBLE, AND PLACED THE EXTENSION INSIDE OF THE MAIN BODY WHERE IT WAS NOT THE DESIRED POSITION. AFTERWARDS, HE ATTEMPTED TO USE STRONG FORCE TO ADVANCE THE EXT
Patient problems: Vascular Dissection
MAUDE report 4690253 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.