Skip to content

Zenith Flex G48404Recalls & reported problems

Cook Incorporated · Zenith Flex, AAA Endovascular Graft Main Body · Class 3 · Implant

All models of Zenith Flex (54)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002484044), received between May 9, 2018 and May 9, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Hole In Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Hole In Material1
  • Leak/Splash1

Patient problems

  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 9, 2018Hole In Material; Leak/Splash; Hole In Material; Leak/Splash

    ON (B)(6) 2015, A FEMALE PATIENT HAD UNDERGONE ENDOVASCULAR ANEURYSM REPAIR (EVAR) BY RIGHT APPROACH FROM THE FEMORAL ARTERY (FA). THE PATIENT WAS SUITABLE FOR THE PROCEDURE. ON (B)(6) 2018, A FOLLOW-UP COMPUTERIZED TOMOGRAPHY (CT) SCAN CONFIRMED ANEURYSM ENLARGEMENT FROM 65 MM TO 70 MM. A CAUSE OF THE ANEURYSM ENLARGEMENT COULD NOT BE DETERMINED, ABDOMINAL SURGERY WAS PERFORMED ON (B)(6) 2018. THE SURGERY CONFIRMED BLOOD LEAKING FROM MULTIPLE SUTURE HOLES OF THE MAIN BODY. THE LEAKING POINTS WERE MENDED BY APPLYING MEDICAL ADHESIVE AND BANDING. THE PATIENT RECOVERED BY (B)(6) 2018.

    Patient problems: No Code Available

    MAUDE report 7498323 ↗

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →