Zenith Flex G48451Recalls & reported problems
Cook Incorporated · Zenith Flex, AAA Endovascular Graft Iliac Leg · Class 3 · Implant
All models of Zenith Flex (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002484518), received between Mar 12, 2018 and Mar 30, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Failure to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Failure to Advance1
- Fluid/Blood Leak1
- Positioning Failure1
Patient problems
- Hematoma1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2020Failure to Advance; Failure to Advance
IT WAS REPORTED THAT THERE WAS DIFFICULTY ADVANCING A ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG INTO THE PATIENT. UPON REMOVAL FROM THE PACKAGING, THE DEVICE WAS RINSED PRIOR TO THE PROCEDURE AND IT WAS NOTED THAT THE STAINLESS STEEL SLEEVE AT THE END OF THE GRAFT "WAS NOT SMOOTH". RESISTANCE WAS FELT WHEN REMOVING THE TRIGGER WIRE AND THE OPERATOR ALSO FOUND THAT THE PROXIMAL END OF THE OUTER SHEATH OF THE GRAFT WAS "TWISTED UNEVENLY". THE GRAFT WAS INSERTED INTO THE PATIENT "UNTIL THE TWISTED PART WAS TOUCHED", BUT RESISTANCE WAS FELT DURING ADVANCEMENT. THE PATIENT HAD CALCIFIED AND TORTUOUS VESSELS. THE OPERATOR REMOVED THE GRAFT FROM THE PATIENT AND SUCCESSFULLY COMPLETED THE PROCEDURE WITH ANOTHER COOK ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG, WHICH SUCCESSFULLY REACHED THE TARGET POSITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9898697 ↗InjuryOct 7, 2019Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AFTER AN ENDOVASCULAR REPAIR SURGERY ON A (B)(6) MALE PATIENT ON (B)(6) 2016, A LEAKAGE AT THE JUNCTION OF AN AORTIC STENT (ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT DISTAL BIFURCATED BODY) AND A ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG (WHICH IS THE SUBJECT OF THIS REPORT) WAS DISCOVERED ON (B)(6) 2016 VIA AN EMERGENCY ANGIOGRAPHY PERFORMED THROUGH FEMORAL ARTERY PUNCTURE. THE CONTRAST SPECIFICALLY SHOWED LEAKAGE BETWEEN THE IMPLANTED AORTIC STENT AND THE RIGHT ILIAC ARTERY. PRIOR TO THIS, AN ABDOMINAL ULTRASOUND EXAMINATION PERFORMED ON (B)(6) 2016 HAD SUGGESTED A RIGHT PERIRENAL HEMATOMA. TWO STENTS FROM ANOTHER MANUFACTURER WERE IMPLANTED IN THE RIGHT ILIAC ARTERY UPON DISCOVERING THE LEAKAGE. INTRAOPERATIVE ULTRASOUND EXPLORATION SHOWED NO PROGRESSION OF THE RETROPERITONEAL HEMATOMA. THE FEMORAL ARTERY INCISION WAS THEN CLOSED AND THE PATIENT SHOWED STABLE VITAL SIGNS. HOWEVER, THE PATIENT CONTINUED TO HAVE GROSS HEMATURIA AND DECREASED HEMOGLOBIN POST-SURGERY WHICH RESULTED IN ADDITIONAL RENAL AND ILIAC ARTERIOGRAPHIES ON (B)(6) 2016. CONTRAST LEAKAGE WAS DETECTED AT THE RIGHT RENAL ARTERY; CONSEQUENTLY, RENAL ARTERY EMBOLIZATION WAS PERFORMED. FOLLOW
Patient problems: Hematoma; No Code Available
MAUDE report 9163603 ↗MalfunctionMar 12, 2018Positioning Failure; Positioning Failure
IT WAS REPORTED DURING THE GRAFT IMPLANTATION IN A (B)(6)-YEAR-OLD, FEMALE PATIENT WITH ABDOMINAL AORTIC ANEURYSM, ONE TFFB-26-96-ZT AND ONE TFLE-12-73-ZT HAD BEEN IMPLANTED SUCCESSFULLY. THE OPERATOR COULD NOT RELEASE THE GRAFT OF THE ZENITH FLEX AAA ENDOVASCULAR GRAFT ILIAC LEG (TFLE-12-90-ZT) AFTER THE OUTER SHEATH WAS WITHDRAWN IN THE SAME SIDE OF THE ILIAC ARTERY. AS REPORTED, THE OPERATION WAS COMPLETED WITH ANOTHER TFLE-12-90-ZT. ACCORDING TO THE REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7330604 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.