Zenith Flex G55209Recalls & reported problems
Cook Incorporated · Zenith Flex, AAA Endovascular Graft Main Body Extension · Class 3 · Implant
All models of Zenith Flex (54)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10827002552095), received between Dec 10, 2014 and Mar 26, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak1
- Leak/Splash1
Patient problems
- Blood Loss1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A TYPE 1 ENDOLEAK PERSISTED AFTER PLACEMENT OF A ZENITH FLEX WITH Z-TRAK AAA ENDOVASCULAR GRAFT MAIN BODY EXTENSION. THE PATIENT PRESENTED SYMPTOMATIC FOLLOWING THE IMPLANTATION OF A COOK MAIN BODY GRAFT AND TWO COOK LEG GRAFTS. A LARGE TYPE 1 ENDOLEAK WAS DISCOVERED, RESULTING IN THE PLACEMENT OF A MAIN BODY EXTENSION GRAFT IN A SECONDARY PROCEDURE (WHICH IS THE SUBJECT OF THIS COMPLAINT). FOLLOWING PLACEMENT, THE GRAFT WAS BALLOONED IN THE LEFT RENAL USING A COOK CODA BALLOON INFLATED TO 25 CC. HOWEVER, THE ENDOLEAK STILL REMAINED, SO TEN COMPETITOR SCREWS WERE PLACED AND THE AREA WAS RE-BALLOONED. ULTIMATELY, A COMPETITOR BARE STENT GRAFT WAS PLACED. AFTERWARDS, A "SLIGHT BLUSHING" LEAK WAS STILL PRESENT, BUT THE PHYSICIAN DECIDED TO LEAVE IT AS IS AS THE LEAK HAD "DRASTICALLY IMPROVED". THE PATIENT WAS ON A COUMADIN REGIMEN AND WAS HEPARINIZED DURING THE PROCEDURE. THE PATIENT REMAINED IN THE HOSPITAL BUT WAS SAID TO BE CURRENTLY STABLE AND "DOING WELL". ADDITIONAL INFORMATION REGARDING THE PATIENT HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE. THE INITIAL TYPE 1 ENDOLEAK ON THE COOK MAIN BODY GRAFT IS REPORTED UNDER MDR: 1820334-2020-00696.
Patient problems: No Code Available
MAUDE report 9886248 ↗MalfunctionDec 10, 2014Leak/Splash
PER COMPLAINT FORM: THIS VALVE WAS LEAKING BADLY BETWEEN THE CLEAR PLASTIC AND THE VALVE. IT WAS SPRAYING EVERYWHERE. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: Blood Loss; No Consequences Or Impact To Patient
MAUDE report 4360769 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗