Skip to content

Salvo™ Spine System 11-6530Recalls & reported problems

Spine Wave, Inc. · Yoke, 4.75 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10840642132984), received between Jul 20, 2020 and Jan 5, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation3
  • Detachment of Device or Device Component1

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2026Detachment of Device or Device Component

    APPROXIMATELY FIVE WEEKS POST-OPERATIVELY DURING ROUTINE RADIOGRAPHIC FOLLOW-UP, ONE PEDICLE SCREW YOKE WAS OBSERVED TO BE DISENGAGED FROM ONE BONE SCREW IN THE 2-LEVEL PEDICLE SCREW CONSTRUCT. THE SURGEON REPORTS THAT NO FURTHER SURGERY IS REQUIRED AS THE PATIENT IS ASYMPTOMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23968071 ↗
  • InjuryDec 2, 2020Material Separation

    POST-OPERATIVELY, THE SURGEON IDENTIFIED A SEPARATED SCREW IN THE PEDICLE SCREW CONSTRUCT VIA RADIOGRAPH. A SURGICAL PROCEDURE WAS PERFORMED TO REPLACE THE YOKE WHILE LEAVING THE SCREW AND REMAINING CONSTRUCT IN PLACE.

    Patient problems: Insufficient Information

    MAUDE report 10931926 ↗
  • MalfunctionNov 25, 2020Material Separation

    DURING ROUTINE POST-OPERATIVE FOLLOW UP, FOUR WEEKS POST-OPERATIVELY, THE SURGEON IDENTIFIED A SEPARATED SCREW IN THE PEDICLE SCREW CONSTRUCT VIA RADIOGRAPH. THE PEDICLE SCREW AND CONSTRUCT WERE LEFT IN PLACE AS THE SURGEON REPORTED THE PATIENT IS ASYMPTOMATIC.

    Patient problems: Insufficient Information

    MAUDE report 10906766 ↗
  • InjuryJul 20, 2020Material Separation

    APPROXIMATELY FOUR MONTHS POST-OPERATIVELY, A PATIENT REPORTEDLY EXPERIENCED PAIN. A SURGICAL PROCEDURE WAS PERFORMED AT WHICH TIME THE SURGEON IDENTIFIED A SEPARATED SCREW IN THE PEDICLE SCREW CONSTRUCT. THE PEDICLE SCREW AND CONSTRUCT WERE REPLACED.

    Patient problems: Pain

    MAUDE report 10295474 ↗

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page