GUARDIAN IPN914182Recalls & reported problems

Teleflex Incorporated · Guardian II NC Intro Torque · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10841156101367), received between Feb 1, 2024 and Feb 1, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • Arrhythmia1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 1, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT: ON (B)(6) 2024, DR. WAS FIXING A CX ARTERY WITH A GUARDIAN HEMOSTASIS VALVE. THERE WAS AN AIR LEAK INTO THE VESSEL AN AIR EMBOLUS WAS NOTED IN THE CIRCUMFLEX ARTEY AND PATIENT WENT INTO ARRYTHMIAS. PATIENT HAD CPR INITIATED AND SHE WAS INTUBATED, AND THE CASE WAS ABORTED TO STABILIZE PATIENT. PATIENT WENT TO THE ICU, STILL INTUBATED, BUT EXPECTED TO RECOVER. ADDITIONAL INFORMATION RECEIVED ON 08JAN2024: THE GUARDIAN APPARENTLY WAS LEAKING SOME BLOOD FROM THE PROXIMAL HUB THE ROTOLOCK WAS IN THE OPEN POSITION FOR THE DURATION OF THE CASE THE PATIENT HAD A MINIMAL BLOOD LEAK DURING THE CASE. ADDITIONAL INFORMATION RECEIVED ON 08JAN2024: PATIENT IS EXTUBATED AND DOING FINE, AS OF TODAY.

    Patient problems: Arrhythmia

    MAUDE report 18622604 ↗

Other Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass devices