Skip to content

Synergetics 58.20-6Recalls & reported problems

Synergetics USA, Inc. · ONE-STEP™ Sutureless Cannula · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10841305101699), received between Oct 19, 2023 and Oct 19, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 19, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ONE-STEP SUTURELESS CANNULA FAILED DURING SURGERY. WHEN THE SURGEON ATTEMPTED TO REMOVE THE TROCAR AT THE END OF THE PROCEDURE, THE TOP OF THE TROCAR BROKE OFF LEAVING THE CANNULA IN THE EYE. THE SURGEON HAD TO GO BACK INTO THE EYE AND RETRIEVE THE CANNULA PART OF THE TROCAR THAT FELL DOWN ONTO THE RETINA. A 20GA INCISION WAS MADE TO ENLARGE THE INCISION USING A V-LANCE BLADE. AFTER SEVERAL ATTEMPTS OF TRYING TO RETRIEVE THE CANNULA WITH DIFFERENT TYPES OF FORCEPS, THE SURGEON DECIDED TO USE A CHANDELIER LIGHT FIBER SO THAT TWO FORCEPS COULD BE USED TO GRASP THE CANNULA. THE CANNULA WAS THEN REMOVED FROM THE BACK OF THE EYE. INCISION WAS ENLARGED, BUT NO SUTURES WERE REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17969750 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates