Skip to content

Qube Qube Compact MonitorRecalls & reported problems

Spacelabs Healthcare (Washington), Inc · COMPACT MONITOR,91390 · Class 2

All models of qube (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841522131455), received between Jan 31, 2025 and Dec 11, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Data Problem” (1 report).

3 FDA recall records name this device's identifier, 3 still open.

6 other recalls cover devices cleared under the same FDA submission (K120616); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 5, 2025Class IIOpen, Classified

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

    Z-0167-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Sep 5, 2025Class IIOpen, Classified

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

    Z-0169-2026 · Root cause: Device Design · Recall details · FDA record ↗

  3. Sep 5, 2025Class IIOpen, Classified

    Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

    Z-0168-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Data Problem1
  • Inappropriate or Unexpected Reset1
  • Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Power Problem; Power Problem

    THE CUSTOMER BIOMED REPORTED THAT USERS NOTIFIED HIM OF AN EVENT THAT OCCURRED AROUND 19:30 EST ON (B)(6) 2025 BEDSIDE MONITORS IN THEIR EMERGENCY ROOM DEPARTMENT WENT BLANK AND STAFF HAD TO RE-ADMIT ALL PATIENTS. IT WAS UNKNOWN HOW LONG THE OUTAGE LASTED AND THAT THERE WERE NO INCIDENTS THAT OCCURRED RELATED TO THE ISSUE. THE CUSTOMER STATED THEY WERE UNAWARE OF ANY POWER ISSUES. WHEN THE BIOMED ARRIVED ON SITE, THE DAYSHIFT STAFF STATED THAT THEY HAD NOT OBSERVED ANY ISSUES WITH DEVICES AND NO FAULTS WERE FOUND WHILE TESTING. THE NUMBER OF DEVICES WAS UNKNOWN, BUT THE BIOMED PROVIDED THE DETAILS OF ONE UNIT. THE BIOMED STATED THEY WERE JUST REPORTING THE INCIDENT AND NO FURTHER INFORMATION IS KNOWN. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23785942 ↗
  • MalfunctionSep 3, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER REPORTED THAT WHILE MONITORING PATIENTS, TWO XHIBIT CENTRAL STATIONS (XCS) WOULD RESTART UNEXPECTEDLY. THE CUSTOMER BIOMED PERFORMED TROUBLESHOOTING AND TRACED THE ISSUE TO A QUBE BEDSIDE MONITOR. WHEN THE BEDSIDE MONITOR WAS ADDED TO A ZONE ON THE XCS, THE XHIBIT SOFTWARE WOULD CRASH AND RESTART. ONCE THE MONITOR WAS REMOVED FROM USE, NO FURTHER ISSUES HAVE BEEN REPORTED. A PRODUCT SUPPORT SPECIALIST REQUESTED THE MONITOR BE SHIPPED TO SPACELABS HEALTHCARE FOR FURTHER EVALUATION. AT THIS TIME THE DEVICE HAS NOT BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22959853 ↗
  • MalfunctionJan 31, 2025Data Problem; Data Problem

    THE CUSTOMER REPORTED THAT WHILE IN USE, THE QUBE BEDSIDE MONITOR DISPLAYED AN ERROR MESSAGE "NO SIGNAL" ON THE BEDSIDE MONITOR. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21284194 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120616. One submission often covers many models, so check the product and lot details in each record.

  • Z-1608-2015 · May 8, 2015 · Terminated
    Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter
  • Z-1145-2015 · Jan 28, 2015 · Terminated
    Spacelabs Healthcare qube Compact Monitor, Model 91390. The Spacelabs Healthcare qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displ
  • Z-1852-2014 · Jun 6, 2014 · Terminated
    Spacelabs Healthcare Qube Compact Monitor, Model 91390. Intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other phy
  • Z-0188-2014 · Oct 17, 2013 · Terminated
    Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia detection or alarms.
  • Z-0351-2014 · Oct 17, 2013 · Terminated
    Spacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia detection or alarms. The qube is intended for use under the direct supervi
  • Z-0403-2013 · Oct 19, 2012 · Terminated
    qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →