Skip to content

Brand Name: Branding is not applied to this product recallZ-0169-2026 · Spacelabs Healthcare, Inc. · initiated Sep 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

FDA root cause category: Device Design

What the company is doing

On 09/05/2025, the firm email an "URGENT MEDICAL DEVICE CORRECTION" Letter to customer informing them that the DVI cables which are used for connecting the Qube (91390) or Xprezzon (91393) to the 94267-19, 94267-L19, 94267-L17 and 94267-L15 displays, do not meet the EMI/EMC electrical safety standards specifically for electromagnetic shielding and may pose serious risks to patient safety by interfering with the operation of another device, particularly in environments where electrically sensitive medical devices are in use. Customers are instructed to acknowledge receipt and understanding of this notification and request replacement cables as soon as possible, however no later than 31-Dec-2025. Do not replace cables while a patient is admitted to the monitor to mitigate any potential disruption in active patient monitoring. Any affected cables not currently in use should not be placed into service. For questions and/or assistance with this recall, contact: Spacelabs Technical Support Spacelabs Healthcare, Inc 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support Email: [email protected]

Recall details

Recall number
Z-0169-2026
Status
Open, Classified
Initiated by the company
Sep 5, 2025
Posted by the FDA
Oct 16, 2025
Recalling firm
Spacelabs Healthcare, Inc.
Quantity
16898
Product code
MHX
FDA submissions
K112962, K120616
Distribution
Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GRC, GTM, GUM, HKG, IDN, IND, IRL, IRN, ISR, ITA, JOR, KWT, LBN, MAR, MEX, MYS, NLD, NZL, OMN, PAK, PAN, PER, PHL, PLS, POL, PRI, PRT, PRY, QAT, ROM, SAU, SGP, SLV, SRB, THA, TTO, TUN, TUR, TWN, VNM.

Product description

Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software Version: N/A Product Description: 16 inch DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393

Code information (models, lots, serial numbers)

Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K112962, K120616 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97616).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page