Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10841736100117), received between May 8, 2019 and Oct 4, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
1 other recall cover devices cleared under the same FDA submission (K132439); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Crack1
- Detachment of Device or Device Component1
- Loose or Intermittent Connection1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2021Detachment of Device or Device Component; Loose or Intermittent Connection; Detachment of Device or Device Component
SKYTRON FIELD SERVICE TECHNICIAN SUBMITTED COMPLAINT ALLEGING THAT THE COTTER PINS THAT HOLDS THE WHEEL ASSEMBLY TOGETHER ON THE INTEGRITY 270 STERILIZER CARRIAGE RACK WERE COMING LOOSE. IF A WHEEL COMES OFF THE RACK THERE IS A RISK OF THE RACK TIPPING AND CONTENTS FALLING.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12571611 ↗MalfunctionMay 8, 2019Crack; Crack
THE COMPANY BECAME AWARE OF TWO INCIDENTS OUTSIDE OF THE USA WITH THE STEAM STERILIZER. THE FIRST INCIDENT WAS IN (B)(6) 2017 AND THE SECOND WAS (B)(6) 2017. A WELD ATTACHING THE AIR HEATER/FILTER ASSEMBLY TO THE MAIN STERILIZER UNIT SHOWED SIGNS OF FATIGUE. THE AFFECTED STEAM STERILIZER MODEL IS ONLY AVAILABLE IN (B)(6), HOWEVER THE COMPANY'S US MODEL, SKYTRON INTEGRITY 270, UTILIZES A SIMILAR AIR HEATER/FILTER ASSEMBLY. THE AIR HEATER/FILTER ASSEMBLY SUPPLIES HEATED FRESH AIR INTO THE CHAMBER AFTER THE STERILIZATION AND DRYING PROCESS HAS BEEN COMPLETED. THIS ALLOWS THE CHAMBER TO EQUALIZE TO ATMOSPHERIC PRESSURE PERMITTING THE USER TO OPEN THE CHAMBER AND UNLOAD THE UNITS. IF THERE IS A FAILURE TO THE WELD, THE CHAMBER PRESSURE WOULD STILL EQUALIZE AT THE END OF THE STERILIZATION AND DRYING CYCLE, BUT THE ROOM AIR MAY NOT GO THROUGH THE FILTER. THERE HAS BEEN NO INSTANCES OF FATIGUE OR FAILURE IN THE USA.
Patient problems: No Patient Involvement
MAUDE report 8591145 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132439. One submission often covers many models, so check the product and lot details in each record.
- Z-1779-2019 · Apr 18, 2019 · TerminatedSkytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity
Other Sterilizer, steam devices
- NA P2131-2Fort Defiance Industries, LLC · Sterilizer, Steam1 recall0 reports
- STATCLAVE G4 Chamber Autoclave G4-621103Scican Ltd · Sterilizer, Steam1 recall0 reports
- STATCLAVE G4 Chamber Autoclave G4-622303Scican Ltd · Sterilizer, Steam1 recall0 reports
- NA AX083020003Steris Mexico, S. de R.L. de C.V. · Sterilizer, steam23 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam11 reports
- Atlas AX0660230830000013Steris Mexico, S. de R.L. de C.V. · Sterilizer, steam9 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam9 reports
- AMSCO AMSCO 600 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page