Radius-PPG 4585Recalls & reported problems
Masimo Corporation · Radius-PPG Neo Neonatal/Adult Pulse Oximeter Adhesive Sensor · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10843997014684), received between Nov 7, 2021 and Sep 9, 2022 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation3
- Sensing Intermittently2
- Connection Problem1
- Low Readings1
- Overheating of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2022Overheating of Device; Overheating of Device
PER MEDWATCH RECEIVED: "PATIENT AWOKE TO WARMTH FROM MASIMO WRIST DEVICE (RECENT HISTORY WITHIN FACILITY OF FORCEFUL SEPARATION OF DEVICES)."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15396710 ↗MalfunctionAug 31, 2022Material Separation; Material Separation
PER MEDWATCH RECEIVED: "BATTERY COMPARTMENT ON WRIST DEVICE FORCEFULLY SEPARATED WHILE IN USE ON PATIENT." NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15332428 ↗MalfunctionJul 29, 2022Material Separation; Material Separation
PER MEDWATCH RECEIVED: "BATTERY COMPARTMENT OF WRISTBAND FORCEFULLY SEPARATED/EXPLODED WHILE IN USE ON PATIENT." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15136097 ↗MalfunctionJul 28, 2022Material Separation; Material Separation
PER MEDWATCH RECEIVED: BATTERY COMPONENT OF WRIST BAND FORCEFULLY SEPARATED OR EXPLODED WHILE IN USE ON PATIENT. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15126855 ↗MalfunctionJan 28, 2022Sensing Intermittently; Sensing Intermittently
THE CUSTOMER REPORTED THE SENSOR SUDDENLY STOPS WORKING AFTER 10-30 MINUTES. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13389543 ↗MalfunctionNov 24, 2021Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT THE SENSOR "WAS LOSING CONNECTIVITY." NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12872404 ↗MalfunctionNov 11, 2021Low Readings; Low Readings
THE CUSTOMER REPORTED THE SENSOR WAS CONSISTENTLY READING OXYGEN 3%-5% LOWER THAN THE SITE'S OXIMETER. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12801100 ↗MalfunctionNov 7, 2021Sensing Intermittently; Sensing Intermittently
THE CUSTOMER REPORTED THE SENSOR MALFUNCTIONED AND STOPPED WORKING IN THE MIDDLE OF THE STUDY. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12763520 ↗
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page