CONMED 00509223600Recalls & reported problems
Conmed Corporation · OVAL BUR, LONG, CARBIDE, 4 X 8 MM · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10845854001257), received between Sep 27, 2018 and Jul 24, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2025Material Fragmentation; Material Fragmentation
THE CUSTOMER REPORTED THAT THE THAT THE 00509223600, OVAL BUR, LONG, CARBIDE, 4 X 8 MM DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED ¿DURING PROCEDURE, TIP BROKE OF BUR BROKE OFF. BROKEN TIP RECOVERED, AND NEW PRODUCT OPENED FOR USE.¿. THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ALTERNATE DEVICE. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586260 ↗MalfunctionSep 27, 2018Break; Break
ON BEHALF OF THE CUSTOMER THE DISTRIBUTOR REPORTED AN ISSUE WITH ITEM 00509223600, 4 X 8MM OVAL BUR, LONG, CARBIDE, 5EA. IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2018, WHILE CUTTING OF THE SCREW HEAD, THE BUR WAS BROKEN AT THE JOINT PART OF HEAD AND SHAFT. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE BROKEN PIECE WAS RETRIEVED BY HAND WITH NO IMPACT TO THE PATIENT. ALL PIECES OF THE BROKEN BUR WERE REMOVED FROM THE PATIENT AND NO PATIENT OR USER INJURY WAS REPORTED. THE SURGERY WAS SUCCESSFULLY COMPLETED USING A BACK-UP BUR WITH NO FURTHER ISSUES AND NO DELAY. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7915335 ↗
FDA clearance and classification
The FDA regulates product code GFF (bur, surgical, general & plastic surgery) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CONMED 00509223600
Has the CONMED 00509223600 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10845854001257), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CONMED 00509223600?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Sep 27, 2018 and Jul 24, 2025. In those reports, the most common specific device problems are “Break” (1) and “Material Fragmentation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CONMED 00509223600?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CONMED 00509223600 FDA approved?
Devices of this type (product code GFF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.
Who makes the CONMED 00509223600, and is it still sold?
The FDA's device database lists Conmed Corporation as the company responsible for it (brand CONMED, model 00509223600). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices
- NA 5820066012Stryker Corporation · Bur, Surgical, General & Plastic Surgery1 recall4 reports
- NA 310.25SSynthes GmbH · BUR, SURGICAL, GENERAL & PLASTIC SURGERY1 recall0 reports
- Arthrex® AR-8400OBEArthrex, Inc. · BUR, SURGICAL, GENERAL & PLASTIC SURGERY46 reports
- NA 310.21Synthes GmbH · BUR, SURGICAL, GENERAL & PLASTIC SURGERY29 reports
- NA 351.27Synthes GmbH · BUR, SURGICAL, GENERAL & PLASTIC SURGERY28 reports
- Arthrex® AR-8500OBEArthrex, Inc. · BUR, SURGICAL, GENERAL & PLASTIC SURGERY26 reports
- Arthrex® AR-8500FOTArthrex, Inc. · BUR, SURGICAL, GENERAL & PLASTIC SURGERY17 reports
- Arthrex® AR-8150PX-45Arthrex, Inc. · BUR, SURGICAL, GENERAL & PLASTIC SURGERY15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page