Skip to content

CONMED 00507116100Recalls & reported problems

Conmed Corporation · OSCILLATOR BLADE, 25.4 X 90 X 1.27 MM (.050") · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854006382), received between Nov 20, 2025 and Feb 23, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE 00507116100 OSCILLATOR BLADE, 25.4 X 90 X 1.27 MM (.050"), DEVICE WAS BEING USED ON (B)(6) 2026 DURING AN UNKNOWN PROCEDURE WHEN IT WAS REPORTED ¿TODAY WE HAD A 5071-161 BREAK FROM LOT: 1506918.¿. FURTHER ASSESSMENT FOUND THAT THE DEVICE DID FRAGMENT INTO THE SURGICAL SITE AND WAS REMOVED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24414753 ↗
  • MalfunctionJan 21, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, 00507116100, OSCILLATOR BLADE, 25.4 X 90 X 1.27 MM (.050"), WAS BEING USED DURING A TOTAL KNEE REPLACEMENT PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿TOTAL KNEE REPLACEMENT WAS IN PROGRESS WHEN WE NOTICED A PIECE OF METAL LODGED IN THE TIBIAL BONE.". THE ¿POWER TOOL AS IT WAS THE LAST EQUIPMENT USED IN THE TIBIA¿ WAS CHECKED AND ¿CLOSE INSPECTION SHOWS THAT TWO (BOTH ENDS) OF THE SAW BLADE TEETH HAD COME OFF.¿ ¿THE KNEE JOINT WAS WASHED NUMEROUS TIMES IN AN ATTEMPT TO FLUSH THE METAL PIECE, BUT WITH NEGATIVE RESULTS.¿ ¿THE SURGEON DECIDED TO GIVE UP THE SEARCH AND WROTE IN HIS OPERATION NOTES ABOUT THE INCIDENT.¿ ¿POST-OPERATIVE X-RAY SHOWS NO METAL PIECE IS LEFT ON THE PATIENT'S KNEE.¿ THERE REPORTER STATES THAT THERE IS NO METAL LEFT IN THE PATIENT¿S KNEE CONFIRMED BY X-RAY. THE PROCEDURE WAS COMPLETED AS PLANNED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24129434 ↗
  • MalfunctionNov 20, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE 00507116100 OSCILLATOR BLADE, 25.4 X 90 X 1.27 MM (.050") DEVICE WAS BEING USED DURING A TOTAL KNEE REPLACEMENT PROCEDURE ON AN UNKNOWN DATE WHEN IT WAS REPORTED ¿ONE OF THE END BLADE TEETH CAME OFF DURING TOTAL KNEE REPLACEMENT SURGERY.¿. IT WAS ALSO REPORTED ¿BROKEN SAW BLADE TOOTH REMOVED FROM PATIENT AS CONFIRMED BY X-RAY.¿ THE PROCEDURE WAS COMPLETED AND THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23603760 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page