Skip to content

CONMED C7312Recalls & reported problems

Conmed Corporation · CLEAR FLEXIBLE CANNULA THREADED, 6.5 MM X 73 MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854018873), received between Jan 22, 2020 and Aug 25, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50437592 ↗
  • MalfunctionNov 26, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, C7312, CLEAR FLEXIBLE CANNULA THREADED, 6.5 MM X 73 MM, WAS BEING USED DURING AN ACL NAIL REMOVAL PROCEDURE ON 18OCT25 WHEN IT WAS REPORTED, ¿THE PACKING OF THE FLEXIBLE CANNULA CAME OFF DURING SURGERY AND ENTERED THE JOINT. THE OPERATION WAS COMPLETED AFTER IT WAS SOMEHOW REMOVED.". THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS REPORTED TO HAVE BEEN COMPLETED AS PLANNED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23655288 ↗
  • MalfunctionJan 22, 2020Material Fragmentation; Material Fragmentation

    THIS INCIDENT WAS ORIGINALLY REPORTED TO THE FDA UNDER REGISTRATION NUMBER 1017294 & REPORT NUMBER 1017294-2019-00158. DURING PROCESSING OF THE RECORD, IT WAS DETERMINED THE REGISTRATION NUMBER USED SHOULD BE 3007305485. THIS REPORT INCLUDES THE ORIGINAL INFORMATION SUBMITTED AS WELL AS ADDITIONAL INFORMATION REGARDING INVESTIGATION AND EVALUATION OF THE INCIDENT. CONMED RECEIVED A REPORT FROM THE CUSTOMER DOCUMENTING ISSUES WITH THE 6.5MM X 73MM CLEAR FLEXIBLE CANNULA THREADED, ITEM # C7312, LOT 994150, THAT OCCURRED ON (B)(6) 2019 AT (B)(6) HOSPITAL IN (B)(6). IT WAS NOTED THAT "DURING ACROMIOPLASTY SURGERY, SEAL WAS DETACHED FROM A CANNULA AND LEFT INSIDE OF A PATIENT BODY IN THE MIDDLE OF INSERTING A SHAVER. AS SOON AS THE SURGEON FOUND IT, HE REMOVED THE LEFT SEAL USING A GRASPER FORCEPS. SURGERY WAS COMPLETED SUCCESSFULLY USING ANOTHER C7312 WITH A 5-MINUTE DELAY ". THE INFORMATION RECEIVED NOTED THE PATIENT WAS A (B)(6) YEAR OLD FEMALE AND THERE WAS NO IMPACT OR INJURY TO HER. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS MADE AVAILABLE. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE (FRAGMENTATION FALLING IN THE PATIE

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9616772 ↗

Other Accessories, arthroscopic devices

All Accessories, arthroscopic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page