Skip to content

POWERPRO PRO2033Recalls & reported problems

Conmed Corporation · POWERPRO HIGH SPEED DRILL ATTACHMENT · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854028384), received between Jul 29, 2024 and Jul 29, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PRO2033, POWERPRO HIGH SPEED DRILL ATTACHMENT WAS BEING USED ON (B)(6) 2024 DURING A PODIATRY CASE AND ¿A HIGH SPEED DRILL ATTACHMENT BROKE APART DURING A CASE TODAY. MULTIPLE PIECES LANDED IN THE PATIENT¿. THERE WAS NO IMPACT OR INJURY FOR THE PATIENT OR USER. PER FURTHER ASSESSMENT IT WAS FOUND THAT THE FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE HIGH SPEED DRILL ATTACHMENT AND A DELAY OF A HALF HOUR TO AN HOUR. THE PATIENT IS "RECOVERING AND THERE WAS NO EXTENDED HOSPITALIZATION". THE PRO8500SB, HALL MICROFREE MINI-DRIVER WILL BE LISTED AS A CONCOMITANT DEVICE AND THERE WAS NO ALLEGATION AGAINST IT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19849494 ↗

Other Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment devices

All Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page