Skip to content

Michigan Critical Care Consultants, Inc. 4949Recalls & reported problems

Michigan Critical Care Consultants, Inc. · Soft Flow Extended Aortic Cannula 21FR 7MM VENTED · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10854916006048), received between Jul 28, 2017 and Sep 9, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation1
  • Use of Device Problem1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Material Deformation; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A SOFT-FLOW EXTENDED ARTERIAL CANNULAE DEVICE, IT WAS REPORTED A KINK AT THE CANNULA'S TIP. THE KINK WAS NOT IN THE LOCATION OF THE SUTURES. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23003479 ↗
  • MalfunctionJul 28, 2017Use of Device Problem; Use of Device Problem

    INFORMATION RECEIVED BY MANUFACTURER (MC3) FROM DISTRIBUTOR ((B)(4)) ON MARCH 29, 2017. UPON ARTERIAL CANNULATION FOR A MITRAL VALVE REPAIR, SURGEON INSERTED SOFT FLOW CANNULA TIP INTO AORTA AND INADVERTENTLY PLACED THE PORTION OF THE CANNULA WITH THE SUTURE RING INTO THE AORTA. IT WAS REPORTED THAT THE SURGICAL PROCEDURE WAS MODIFIED TO SAFETY REMOVE THE CANNULA WITH THE SUTURE RING ATTACHED FROM THE AORTA. WHEN IT WAS TIME TO REMOVE THE CANNULA FROM THE PATIENT AT THE END OF THE CASE, THE SURGEON WAS CONCERNED THAT THE SUTURE RING COULD FALL OFF THE CANNULA BODY. AS A RESULT, HE DECIDED TO PUT THE PATIENT ON CIRCULATORY ARREST IN ORDER TO ENSURE SAFE REMOVAL OF THE DEVICE WITHOUT DETACHMENT OF THE SUTURE RING. IT WAS CONFIRMED THAT THE DEVICE WAS SAFELY REMOVED AND THE SUTURE RING REMAINED ATTACHED TO THE DEVICE. THERE WAS NO REPORTED ADVERSE EFFECT TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 6749819 ↗

Other Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass devices

All Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass devices →

Sources and updates