MC3 Nautilus Smart ECMO Module 48135ERecalls & reported problems

Michigan Critical Care Consultants, Inc. · Nautilus ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed. Smart ECMO Module device contains integrated sensors with an electronic touch screen display. The following measured parameters are visible on the electronic display: inlet pressure, inlet oxygen saturation, outlet pressure, outlet oxygen saturation, and outlet blood temperature. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10854916006895), received between Oct 20, 2025 and Oct 20, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

    Z-1310-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device1
  • Increase in Pressure1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Thrombocytopenia1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2025Increase in Pressure; Obstruction of Flow; Air/Gas in Device; Increase in Pressure; Obstruction of Flow; Air/Gas in Device

    IT WAS REPORTED THAT AN ALLEGED PRODUCT ISSUE OCCURRED WITH THE MC3 NAUTILUS EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) OXYGENATOR DURING USE. THE CIRCUIT WITH THE DEVICE WAS INITIALLY PRIMED AND LEFT READY FOR USE. AFTER SEVERAL DAYS, A PATIENT WAS CANNULATED AND CONNECTED TO THE CIRCUIT. JUST BEFORE CONNECTING THE PATIENT, AIR AND BUBBLES WERE OBSERVED IN THE SUPERIOR PART OF THE DEVICE, PROMPTING THE TEAM TO ATTEMPT RE-PRIMING USING A STERILE SYRINGE AND SERUM THROUGH THE CONNECTION INSIDE, COLLECTING IT THROUGH THE OUTSIDE CONNECTION. ONCE RE-PRIMED, THE PATIENT WAS CONNECTED TO THE CIRCUIT. DURING ECMO, A BLANK TOP WAS OBSERVED ON THE IN SIDE OF THE NAUTILUS SMART, APPEARING AS IF THERE WAS AIR AND NO BLOOD, WHILE THE OUT SIDE DID NOT SHOW THIS FEATURE. IT WAS NOTED THAT THE PATIENT WAS NOT COAGULATED. THE USE OF THE DEVICE WAS UNSPECIFIED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Thrombocytopenia

    MAUDE report 23334261 ↗