MC3 Nautilus ECMO Oxygenator 48145ERecalls & reported problems
Michigan Critical Care Consultants, Inc. · Nautilus ECMO oxygenators are diffusion membrane oxygenators used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10854916006901), received between Dec 9, 2021 and Dec 10, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 11, 2023Class IIOpen, Classified
May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
Z-1309-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT PRIOR TO USE OF THE MC3 NAUTILUS EXTRA CORPOREAL MEMBRANE OXYGENATION (ECMO) OXYGENATOR, AFTER PRIMING AND CONNECTION TO THE PATIENT, BLOOD CAME OUT FROM THE GAS OUTLET. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 20890337 ↗MalfunctionJan 18, 2024Infusion or Flow Problem; Infusion or Flow Problem
ON (B)(6) 2023, WITHIN 10 MINUTES OF ECMO BYPASS, THE DELTA P WAS GREATER THAN 380 MMHG. DONOR BLOOD WAS GIVEN TO THE PATIENT AT THE INITIATION OF ECMO AND THE PATIENT WAS NOT SEPTIC. THERE WAS NO LOW GAS EXCHANGE OBSERVED. HEPARIN WAS GIVEN, ACT WAS REPORTED IN THE 200'S. IT WAS UNKNOWN IF THE PATIENT HAD PREVIOUSLY BEEN ON ANTI-COAGULATION THERAPY. THE NAUTILUS OXYGENATOR WAS REPLACED WITH NO IMPACT TO THE PATIENT. IT WAS REPORTED THAT A CLOT WAS OBSERVED IN THE OXYGENATOR AFTER THE DEVICE HAD BEEN REPLACED. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18539860 ↗MalfunctionJan 12, 2024Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2023, AFTER LESS THAN 24 HOURS USE OF A NAUTILUS OXYGENATOR, THE OUTLET HAD A LEAK AND WAS DRIPPING. THE OXYGENATOR WAS REPLACED WITH NO IMPACT TO THE PATIENT. THE PATIENT WAS REPORTED AS ALIVE WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18503752 ↗MalfunctionDec 9, 2021Fluid/Blood Leak; Fluid/Blood Leak
FLUID LEAK OUT OF GAS OUTLET PORT. DEVICE CHANGED OUT AND REPLACED WITH ANOTHER NAUTILUS OXYGENATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12969663 ↗
Other Oxygenator, Long Term Support Greater Than 6 Hours devices
- MC3 Nautilus(tm) ECMO Oxygenator 48145Michigan Critical Care Consultants, Inc. · Oxygenator, Long Term Support Greater Than 6 Hours1 recall3 deaths47 reports
- MC3 Nautilus (tm) Smart ECMO Module 48135Michigan Critical Care Consultants, Inc. · Oxygenator, Long Term Support Greater Than 6 Hours1 recall27 reports
- MC3 Nautilus Smart ECMO Module 48135EMichigan Critical Care Consultants, Inc. · Oxygenator, Long Term Support Greater Than 6 Hours1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page