Inspire 900-007-008Recalls & reported problems

Inspire Medical Systems, Inc. · The Model 3028 IPG contains electronics and a battery that are sealed inside a titanium case. The IPG is implanted subcutaneously, below the clavicle in the upper chest, and connect to the stimulation lead and sensing lead. The Model 3028 does not contain any software or firmware. All functions including the telemetry and algorithm have been designed into the hardware of the IPG. The algorithm synchronizes stimulation of the hypoglossal nerve with respiration signals. The IPG processes the same dynamic range of pressure signals (2-48 cmH2O) in order to account for pressure reading variability. Based on typical settings from the STAR pivotal trial, the longevity of the Model 3028’s battery will average 10 years although the device is smaller than the currently approved version (Model 3024). In addition the Model 3028 IPG will allow patients to safely undergo magnetic resonance imaging (MRI) under specified conditions. · Class 3 · Implant

All models of Inspire Sleep Apnea Implant (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Inspire hypoglossal nerve stimulator for sleep apnea has 248 reports on file, with 1 death and 247 injuries. Many describe pain or infection around the implant site, and some describe the device moving within its pocket, wounds opening, or the implant eroding through the skin. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
248
Deaths reported
1
Injuries reported
247
Recalls naming it
0

Summary

The FDA's MAUDE database contains 248 adverse event reports that list this exact device (DI 10855728005656), received between Sep 16, 2019 and Jul 27, 2026 — classified as 1 death, 247 injuries.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (15 reports).

3 other recalls cover devices cleared under the same FDA submission (P130008); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem219
  • Migration or Expulsion of Device15
  • Migration4
  • Device Sensing Problem2
  • Insufficient Device Problem Information2
  • Malposition of Device2
  • Expulsion1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Unspecified Infection72
  • Pain50
  • Bacterial Infection41
  • Hematoma25
  • Wound Dehiscence17
  • Erosion15
  • Swelling/ Edema14
  • Perforation of Vessels9

Recent adverse event reports

The 25 most recent of 248 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    Patient presented to the physician with pain at the IPG site, including pain radiating from the IPG site to the armpit region when lifting heavy objects. Physician prescribed steroids.

    Patient problems: Implant Pain

    MAUDE report 50011424 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE. PHYSICIAN PRESCRIBED AN ANTI-INFLAMMATORY MEDICATION.

    Patient problems: Implant Pain

    MAUDE report 25462811 ↗
  • InjuryMay 5, 2026Expulsion

    PATIENT PRESENTED TO THE PHYSICIAN WITH WOUND DEHISCENCE AT THE CHEST INCISION SITE, AND THE IPG HAD ERODED THROUGH THE SKIN. ON EXAMINATION, THE PHYSICIAN NOTED REDNESS, SWELLING, AND YELLOW FLUID AT THE SITE. PHYSICIAN CLEANED THE AREA WITH BETADINE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR SEVEN DAYS LATER, EXPLANTED THE IPG AND SENSING LEAD, AND A PORTION OF THE STIMULATION, AND CLOSED.

    Patient problems: Inflammation; Pocket Erosion; Wound Dehiscence

    MAUDE report 25086599 ↗
  • InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE AND INABILITY TO SLEEP WITH THERAPY. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND EXPLANTED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Pain

    MAUDE report 24654176 ↗
  • InjuryFeb 6, 2026Migration

    PATIENT PRESENTED TO THE PHYSICIAN WITH THE IPG MOVING WITHIN THE POCKET, CAUSING PAIN. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, REMOVED THE IPG AND SENSING LEAD, IMPLANTED A NEWER IPG MODEL, WHICH DOES NOT REQUIRE A SENSING LEAD, SUTURED, AND CLOSED.

    Patient problems: Pain

    MAUDE report 24283469 ↗
  • InjuryAug 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, REMOVED THE IPG, SENSING LEAD, AND A PORTION OF THE STIMULATION LEAD BODY, AND CLOSED.

    Patient problems: Pain

    MAUDE report 22849781 ↗
  • InjuryNov 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE NECK AND THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REPLACED THE ENTIRE SYSTEM.

    Patient problems: Pain

    MAUDE report 20750774 ↗
  • InjuryNov 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE AFTER REACHING FOR SOMETHING IN THEIR VEHICLE. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND REPLACED THE IPG AND SENSING LEAD.

    Patient problems: Pain

    MAUDE report 20652215 ↗
  • InjuryOct 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH AN INFECTION AT THE CHEST INCISION SITE. PHYSICIAN PRESCRIBED ANTIBIOTICS. PHYSICIAN BROUGHT THE PATIENT INTO THE OR 6 DAYS LATER AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Unspecified Infection

    MAUDE report 20577063 ↗
  • InjuryApr 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH SCAR PAIN AT THE CHEST INCISION SITE. PHYSICIAN GAVE PATIENT STEROID INJECTIONS. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 19111198 ↗
  • InjuryMar 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH EXCORIATIONS AT THE IPG CHEST INCISION. PHYSICIAN PRESCRIBED ANTIBIOTICS.

    Patient problems: Erythema; Skin Inflammation/ Irritation

    MAUDE report 18923891 ↗
  • InjuryFeb 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH CHEST PAIN AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE ENTIRE SYSTEM.

    Patient problems: Pain

    MAUDE report 18652595 ↗
  • InjuryJan 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN RADIATING IN THE CHEST AROUND THE IPG SITE WHEN USING INSPIRE THERAPY. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND REPLACED THE SENSING LEAD AND IPG. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 18546182 ↗
  • InjuryOct 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR, REPOSITIONED THE IPG AND LEADS, SUTURED THE DEVICE, AND CLOSED THE CHEST INCISION.

    Patient problems: Pain

    MAUDE report 18031055 ↗
  • InjurySep 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG SITE. PHYSICIAN IS UNABLE TO DETERMINE WHAT IS CAUSING THE PAIN. PHYSICIAN BROUGHT THE PATIENT INTO THE OPERATING ROOM AND REMOVED THE INSPIRE SYSTEM.

    Patient problems: Pain

    MAUDE report 17728655 ↗
  • InjuryJul 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH INFECTION AT THE IPG SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Unspecified Infection

    MAUDE report 17385471 ↗
  • InjuryOct 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER THE INSPIRE IMPLANT PROCEDURE, THE PATIENT DEVELOPED A PNEUMOTHORAX AND REQUIRED 2% OXYGEN TO MAINTAIN SAT LEVEL ABOVE 90. PHYSICIAN CONSULTED WITH RADIOLOGIST AND PIGTAIL CATHETER WAS PLACED WITH GUIDANCE OF CT FLUOROSCOPY. PATIENT TO BE KEPT OVERNIGHT. THIS RESOLVED THE ISSUE.

    Patient problems: Pneumothorax

    MAUDE report 15609439 ↗
  • InjuryOct 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH REDNESS, TENDERNESS, AND EXPOSED IPG. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE INSPIRE SYSTEM AND FLUSHED THE AREA. THIS RESOLVED THE ISSUE.

    Patient problems: Erosion

    MAUDE report 15607012 ↗
  • InjuryOct 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH EAR PAIN. PHYSICIAN PRESCRIBED STEROIDS AND ANTIBIOTICS. THIS RESOLVED THE ISSUE.

    Patient problems: Pain

    MAUDE report 15579723 ↗
  • InjuryOct 11, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH MILD REDNESS AND WEEPING FROM THE CHEST WALL INCISION. PHYSICIAN PRESCRIBED A COURSE OF ORAL ANTIBIOTICS AND TOPICAL MUPIROCIN CREAM. PATIENT PRESENTED BACK TO THE PHYSICIAN WITH REDNESS AND IMPROVED WEEPING FROM THE CHEST WALL SITE AND REDNESS ON HIS CHIN LINE INCISION. PHYSICIAN BELIEVES IT IS THE SUTURE TRYING TO EXTRUDE AND INJECTED BOTH SITES WITH A STEROID. PATIENT PRESENTED BACK AGAIN TO THE PHYSICIAN WITH SWELLING AND INCREASED PAIN UNDER THE CHIN. PHYSICIAN PRESCRIBED ANTIBIOTICS, ORAL STEROIDS, AND ORAL BENADRYL.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 15574454 ↗
  • InjuryOct 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH A HEMATOMA 8 DAYS POST IMPLANT PROCEDURE. PATIENT WAS ON A BLOOD THINNER DURING THE PROCEDURE BUT DISCONTINUED ONCE THE HEMATOMA DEVELOPED. PATIENT KEPT OVERNIGHT. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND OBSERVED SUPERFICIAL BLEEDING AND BLOOD CLOTS. PHYSICIAN CLEANED THE POCKET, REMOVED THE IPG, AND DID ADDITIONAL CLEANING OF THE POCKET. PHYSICIAN USED BIPOLAR ENERGY AND ANTICOAGULANT POWDER TO STOP THE BLEEDING. HEMATOMA WAS EVACUATED. IPG WAS PLACED BACK INTO THE POCKET AND RE-SUTURED. THIS RESOLVED THE ISSUE.

    Patient problems: Hematoma

    MAUDE report 15534329 ↗
  • InjuryOct 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH AN INFECTION WITH BACTERIA. PHYSICIAN PRESCRIBED ANTIBIOTICS AND BROUGHT THE PATIENT INTO THE OR 8 DAYS LATER AND REMOVED THE ENTIRE INSPIRE SYSTEM.

    Patient problems: Bacterial Infection

    MAUDE report 15532210 ↗
  • InjurySep 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED WITH A RED IPG SITE WITH SUSPECTED INFECTION. PHYSICIAN PRESCRIBED ANTIBIOTICS FOR 10 DAYS. THIS RESOLVED THE ISSUE.

    Patient problems: Unspecified Infection

    MAUDE report 15489406 ↗
  • InjurySep 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AT THE IPG IMPLANT SITE. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED THE IPG.

    Patient problems: Pain

    MAUDE report 15477953 ↗
  • InjurySep 24, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT PRESENTED TO THE PHYSICIAN WITH PAIN AND SWELLING ON THE RIGHT SIDE OF THE CHEST. PHYSICIAN BROUGHT THE PATIENT INTO THE OR AND REMOVED A MAJORITY OF THE INSPIRE SYSTEM. THE SENSING LEAD WAS CUT NEAR THE IPG AND THE DISTAL END REMAINS.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 15477117 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P130008. One submission often covers many models, so check the product and lot details in each record.

  • Z-2516-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire stimulation Lead Model/Catalog Number: 4063 Product Description: The Inspire Stimulation Lead (Model 4063) is design
  • Z-2515-2026 · May 11, 2026 · Open, Classified
    Brand Name: Inspire Medical Systems, Inc. Product Name: Inspire Respiratory Sensing Lead Model/Catalog Number: 4340 Software Version: N/A Product Description: The Inspire Respirato
  • Z-2969-2018 · Jun 12, 2018 · Terminated
    Inspire II Model 3024 Implantable Pulse Generator (IPG) for Obstructive Sleep Apnea, Inspire Part #900-003-020, Product Usage: The device is an implanted component of the Inspir

Other Stimulator, Hypoglossal Nerve, Implanted, Apnea devices

All Stimulator, Hypoglossal Nerve, Implanted, Apnea devices →