Inspire Sleep Apnea ImplantRecalls & reported problems across all models
Inspire Medical Systems, Inc. · 24 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Inspire Sleep Apnea Implant model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2,200
- Deaths reported
- 4
- Injuries reported
- 2,093
- Recalls
- 3
We found 24 models of Inspire Sleep Apnea Implant by Inspire Medical Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 2,200 adverse event reports in the FDA's MAUDE database, received between Feb 9, 2016 and Sep 30, 2026 — classified as 4 deaths, 2,093 injuries, 103 malfunctions.
The most frequently reported specific device problem is “Migration” (141 reports).
3 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Inspire Sleep Apnea Implant recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 11, 2026Open, Classified
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Z-2516-2026 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
- May 11, 2026Open, Classified
Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By Date may result in unplanned or premature revision surgery.
Z-2515-2026 · Root cause: Process control · Recall details · FDA record ↗
- Jul 31, 2024Open, Classified
Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.
Z-0041-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported Inspire Sleep Apnea Implant problems
Device problems
- Adverse Event Without Identified Device or Use Problem1,713
- Migration141
- Migration or Expulsion of Device118
- Device Sensing Problem104
- Expulsion32
- Malposition of Device32
- Insufficient Device Problem Information13
- Detachment of Device or Device Component11
Patient problems
- Pain600
- Unspecified Infection432
- Erosion202
- Hematoma189
- Wound Dehiscence185
- No Clinical Signs, Symptoms or Conditions120
- Perforation of Vessels120
- Swelling/ Edema104
Inspire Sleep Apnea Implant models
24 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Inspire 900-007-014Stimulator, Hypoglossal Nerve, Implanted, Apnea3 deaths1,012 reports
- Inspire 900-013-001Stimulator, Hypoglossal Nerve, Implanted, Apnea2 recalls359 reports
- Inspire 900-007-008Stimulator, Hypoglossal Nerve, Implanted, Apnea1 death248 reports
- Inspire 900-009-002Stimulator, Hypoglossal Nerve, Implanted, Apnea138 reports
- Inspire 900-001-009Stimulator, Hypoglossal Nerve, Implanted, Apnea125 reports
- Inspire 900-001-011Stimulator, Hypoglossal Nerve, Implanted, Apnea56 reports
- Inspire 900-013-003Stimulator, Hypoglossal Nerve, Implanted, Apnea1 recall0 reports
- Inspire 900-014-005Stimulator, Hypoglossal Nerve, Implanted, Apnea1 recall0 reports
- Inspire 900-007-003Stimulator, Hypoglossal Nerve, Implanted, Apnea46 reports
- Inspire 900-003-015Stimulator, Hypoglossal Nerve, Implanted, Apnea45 reports
- Inspire 900-009-003Stimulator, Hypoglossal Nerve, Implanted, Apnea43 reports
- Inspire 900-001-008Stimulator, Hypoglossal Nerve, Implanted, Apnea26 reports
- Inspire 900-002-005Stimulator, Hypoglossal Nerve, Implanted, Apnea17 reports
- Inspire 900-001-005Stimulator, Hypoglossal Nerve, Implanted, Apnea13 reports
- Inspire 900-002-007Stimulator, Hypoglossal Nerve, Implanted, Apnea12 reports
- Inspire 900-001-010Stimulator, Hypoglossal Nerve, Implanted, Apnea12 reports
- Inspire 900-002-010Stimulator, Hypoglossal Nerve, Implanted, Apnea10 reports
- Inspire 900-007-001Stimulator, Hypoglossal Nerve, Implanted, Apnea9 reports
- Inspire 900-002-008Stimulator, Hypoglossal Nerve, Implanted, Apnea9 reports
- Inspire 900-007-010Stimulator, Hypoglossal Nerve, Implanted, Apnea8 reports
- Inspire 900-002-009Stimulator, Hypoglossal Nerve, Implanted, Apnea5 reports
- Inspire 900-001-007Stimulator, Hypoglossal Nerve, Implanted, Apnea4 reports
- Inspire 900-009-001Stimulator, Hypoglossal Nerve, Implanted, Apnea2 reports
- Inspire 900-004-001Stimulator, Hypoglossal Nerve, Implanted, Apnea1 report
Recent Inspire Sleep Apnea Implant adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
Patient presented to the physician with pain at the chin and neck. Due to these persistent symptoms, the patient requested removal of the device. Physician brought the patient into the OR and explanted the entire Inspire system.
Patient problems: Neck Pain; Pain
MAUDE report 50950137 ↗InjurySep 30, 2026Expulsion
Patient problems: Pocket Erosion
MAUDE report 50949988 ↗InjurySep 30, 2026Migration
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949709 ↗InjurySep 30, 2026Migration
Patient presented to the physician with movement of the IPG within the pocket, causing pain. Imaging was obtained, showing that the IPG appeared to have shifted downward. Physician brought the patient into the OR, re-sutured the IPG securely, and closed.
Patient problems: Pain
MAUDE report 50949538 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient presented to the physician with ongoing right-sided tongue paralysis and left-sided tongue deviation. Upon examination, the physician confirmed nerve damage attributed to Inspire. No intervention has been reported at this time. Physician plans to re-evaluate the patient.
Patient problems: Nerve Damage; Neuropathy
MAUDE report 50936369 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Medical device site infection
MAUDE report 50809535 ↗InjurySep 2, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Nerve Damage; Neuropathy; Speech Disorder
MAUDE report 50568227 ↗InjurySep 1, 2026Expulsion
Patient presented to the physician on (b)(6) 2026 with the stimulation lead eroded through the skin at the neck incision site. No intervention was performed at the time. On (b)(6) 2026, the physician brought the patient into the Operating Room, repositioned the lead beneath the tissue, re-sutured, and closed the incision.
Patient problems: Pocket Erosion
MAUDE report 50549250 ↗InjuryAug 27, 2026Migration
Patient presented to the physician with flipping and movement of the IPG within the pocket, resulting in pain. Physician brought the patient into the OR, re-sutured the IPG, and closed.
Patient problems: Pain
MAUDE report 50484048 ↗InjuryAug 27, 2026Expulsion
Patient presented to the physician with the stimulation lead eroded through the skin at the neck incision site and also reported headaches. Physician noted that the patient had been taking antibiotics and steroids. Physician brought the patient into the OR, repositioned the lead beneath the tissue, irrigated the site with antibiotic solution, re-sutured, and closed.
Patient problems: Headache; Pocket Erosion
MAUDE report 50482845 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
Patient presented to the physician, post-implant procedure, with a hematoma at the neck incision site. Physician brought the patient into the OPERATING ROOM, drained the hematoma, irrigated the neck pocket, and closed.
Patient problems: Hematoma
MAUDE report 50464517 ↗InjuryAug 26, 2026Impedance Problem
During a routine follow-up appointment, the physician identified a device impedance issue. Imaging was obtained, which indicated a possible stimulation lead issue. Physician brought the patient to the operating room, and upon surgical exploration, the stimulation lead was found to be damaged. The physician noted twiddler syndrome, which posed a risk of patient injury. Physician explanted the existing stimulation lead, implanted a new lead, sutured, and closed. The revision surgery restored therapy, addressed the risk, and the issue is now resolved.
Patient problems: Twiddlers Syndrome
MAUDE report 50464499 ↗InjuryAug 26, 2026Migration
Patient presented to the physician with flipping of the IPG within the pocket, posing a potential risk of injury. Physician brought the patient into the OR, and upon surgical exploration, twiddling of the device was noted. Physician explanted the IPG, sensing lead, and a portion of the stimulation lead, and closed. The explant surgery addressed the risk, and the issue is now resolved.
Patient problems: Twiddlers Syndrome
MAUDE report 50458208 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
Patient presented to the physician with a burning and stabbing sensation at the chest incision site. Physician brought the patient into the OR, explanted the IPG and sensing lead, implanted a newer IPG model that does not require a sensing lead, connected the new IPG to the existing stimulation lead, sutured, and closed.
Patient problems: Burning Sensation
MAUDE report 50457460 ↗InjuryAug 26, 2026Adverse Event Without Identified Device or Use Problem
Patient presented to the physician with persistent headaches. Physician prescribed medication to manage the symptoms.
Patient problems: Headache
MAUDE report 50455575 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered, combining the brand names used for this product line.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page