Skip to content

Boppli 54-0002Recalls & reported problems

Pyrames Inc. · Ship Kit, Boppli Bedside Device · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Feb 24, 2025Open, Classified

    Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.

    Z-1508-2025 · Root cause: Software design · FDA record ↗

Other System, Measurement, Blood-Pressure, Non-Invasive devices

All System, Measurement, Blood-Pressure, Non-Invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page