S-CATH™ Esophageal Temperature Probe CS-2001 (65cm Length)Recalls & reported problems
CIRCA Scientific, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10860237000214), received between Feb 24, 2022 and Aug 25, 2023 — classified as 1 death, 6 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method3
- Adverse Event Without Identified Device or Use Problem2
- Difficult to Insert2
Patient problems
- Fistula2
- Laceration(s) of Esophagus2
- Perforation2
- Bronchial Hemorrhage1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 25, 2023Difficult to Insert; Difficult to Insert
A PATIENT HAD A PERFORATION IN THEIR THROAT AND THEY THINK IT MAY BE FROM WHEN THE CRNA INTRODUCED THE PROBE.
Patient problems: Laceration(s) of Esophagus
MAUDE report 17627653 ↗InjuryMay 5, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE PROCEDURE WAS ATRIAL FLUTTER/FIBRILLATION ABLATION. THE ATRIAL FLUTTER WAS ABLATED AND TRANSSEPTAL WAS PERFORMED. THE PATIENT'S BLOOD PRESSURE DROPPED SUDDENLY AND THE PROCEDURE WAS STOPPED. THE PATIENT WAS TRANSFERRED TO THE ICU. PATIENT HAD ECMO DONE LATER IN THE DAY. THE ABLATION IN THE LEFT ATRIUM HAD NOT PERFORMED AND CIRCA HAD NOT CLARIFIED ITS POSITION UNDER X-RAY PRIOR THE INCIDENT OCCURRENCE. ELECTROPHYSIOLOGIST STATED THAT THE CIRCA PROBE WAS PLACED IN THE BRONCHUS WHICH CAUSED HEMORRHAGE. ANESTHESIOLOGIST ASSISTANT STATED THAT THERE WAS NO EVIDENCE OF THE PATIENT'S BRONCHIAL HEMORRHAGE.
Patient problems: Bronchial Hemorrhage
MAUDE report 16881576 ↗DeathFeb 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN ATRIAL FIBRILLATION (PVI) PROCEDURE, AN ESOPHAGEAL PERFORATION CAUSING SEPSIS IN THE PATIENT. THE TEMPERATURE MONITORING DEVICE WAS ALARMING ABOVE THE MANUFACTURER'S RECOMMENDED HIGH USAGE SUGGESTIONS. THE PATIENT WAS ADMITTED 17 DAYS POST PROCEDURE DUE TO SEPSIS CAUSED FROM AN ESOPHAGEAL PERFORATION THE (SIC) LIKELY OCCURRED DURING THE ABLATION PROCEDURE. PER CT: MEDIASTINUM AND LYMPHATIC, THERE WAS PNEUMOMEDIASTINUM, WITH AIR ANTERIOR TO THE ESOPHAGUS AT THE LEVEL OF THE RIGHT MAIN PULMONARY ARTERY, AND EXPENDING INFERIORLY ALONG THE POSTERIOR MARGIN OF THE LEFT ATRIUM. IT IS SUSPECTED TO ARISE FROM ESOPHAGEAL PERFORATION AT THE LEVEL OF THE MID ESOPHAGUS.
Patient problems: Fistula
MAUDE report 16390702 ↗InjuryMay 6, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
DR. REPORTS A PATIENT LUNG PERFORATION CAUSED BY CIRCA PROBE. PATIENT COUGHED UP BLOOD FOR FIVE DAYS, THEN RECOVERED. NO FLUOROSCOPY WAS USED IN THE PROCEDURE.
Patient problems: Perforation
MAUDE report 14311657 ↗InjuryMay 6, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
DR. REPORTS A PATIENT LUNG PERFORATION CAUSED BY CIRCA PROBE. PATIENT COUGHED UP BLOOD FOR FIVE DAYS, THEN RECOVERED. NO FLUOROSCOPY WAS USED IN THE PROCEDURE.
Patient problems: Perforation
MAUDE report 14311468 ↗InjuryFeb 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PHYSICIAN ASKED US TO REPORT THE OCCURRENCE OF A PATIENT WHO HAD AN ESOPHAGEAL FISULA FOLLOWING AN RF ABLATION. CIRCA WAS USED TO MONITOR ESOPHAGEAL TEMPERATURE. PATIENT HAS SINCE RECOVERED. TEE PROBE WAS ALSO USED ON PATIENT FOR WATCHMAN INSERTION.
Patient problems: Fistula
MAUDE report 13602629 ↗InjuryFeb 24, 2022Difficult to Insert; Difficult to Insert
DURING PROBE INSERTION THEY ENCOUNTERED RESISTANCE. DOCTOR ALSO USES ESOSURE WITH AN OG TUBE. OG TUBE WAS INSERTED BEFORE CIRCA PROBE. TOWARD END OF CASE BLEEDING WAS NOTED. ENT WAS CALLED AND DIDN'T NOTE ANY DAMAGE.
Patient problems: Laceration(s) of Esophagus
MAUDE report 13595998 ↗
Other Thermometer, Electronic, Clinical devices
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical1 recall221 reports
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical2 recalls13 reports
- NA B105 VSP3.0 Patient MonitorGE Medical Systems Information Technologies, Inc. · Thermometer, electronic, clinical2 recalls0 reports
- Medline MDSTH1002Medline Industries, Inc. · Thermometer, electronic, clinical65 reports
- Mon-a-Therm 90050Covidien LP · Thermometer, electronic, clinical1 recall3 reports
- Mon-a-Therm 90044Covidien LP · Thermometer, electronic, clinical1 recall2 reports
- CARESCAPE B450 MBA313GE Healthcare Finland Oy · Thermometer, electronic, clinical1 recall1 report
- Temperature Probes Esophageal/Rectal Temperature ProbePhilips Medical Systems Hsg · Thermometer, electronic, clinical1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page