Skip to content

DELTAMAXX CERECYTE CDX181445-30Recalls & reported problems

Medos International Sàrl · DELTAMAXX microcoil DELTAMAXX 18 CERECYTE Coil 14mm x 45cm 190cm Contents: 1 Detachable Coil System Coil Type: DELTAMAXX 18 CERECYTE Coil · Class 2 · Implant

All models of DELTAMAXX CERECYTE (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10878528008658), received between Jul 13, 2017 and Jul 13, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Premature Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Premature Activation1
  • Stretched1

Patient problems

  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 13, 2017Premature Activation; Stretched; Premature Activation; Stretched

    AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING THE PROCEDURE COIL INTRODUCER DAMAGE WAS REPORTED WHEN USING TWO DELTAMAXX COILS (CDX18246030/C38840 AND CDX18144530/C34584) AND ANOTHER DELTAMAXX COIL (CDX18144530/C36948) UNRAVELED AND PREMATURELY DETACHED. THE PROCEDURE WAS COIL EMBOLIZATION OF AN ANEURYSM AT THE SPLENIC ARTERY. THE PATIENT WAS A FEMALE WHOSE VESSEL WAS MODERATELY TORTUROUS AND NOT CALCIFIED. THE DELTAMAXX COIL (CDX18246030/C38840) WAS USED AS THE FRAMING COIL HOWEVER "THE BALLOON COULD NOT BE HANDLED", SO THE PHYSICIAN TRIED TO RE-SHEATH THE COIL BUT INTRODUCER SHEATH BROKE AND THE COIL COULD NOT BE RE-SHEATHED. ANOTHER DELTAMAXX COIL (CDX18144530/C34584) WAS USED AS A FILLING COIL, BUT THE MICRO CATHETER SLID BACK PROXIMALLY AND THE COIL WOULD NOT FIT INTO THE CATHETER, SO THE PHYSICIAN TRIED TO RE-SHEATH THE COIL HOWEVER THE INTRODUCER SHEATH GOT BROKEN AND THE COIL COULD NOT BE RE-SHEATHED. EXCESSIVE FORCE WAS NOT USED WHEN HANDLING THE CATHETER AND THE COIL. AFTER THIS ISSUE, THE MICROCATHETER WAS REINSERTED. A DELTAMAXX COIL (CDX18144530/C36948) WAS INSERTED HOWEVER DUE TO INCORRECT POSITIONING OF THE CATHETER TIP IN THE ANEURYSM THE COIL DID NOT FIT IN THE

    Patient problems: No Code Available