The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10878528009426), received between Jan 6, 2017 and Jan 6, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Physical Resistance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Physical Resistance1
  • Stretched1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2017Stretched; Physical Resistance; Stretched; Physical Resistance

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT DURING AN ELECTIVE COILING PROCEDURE TO TREAT A SECULAR MODERATELY WIDE NECK ACOM ANEURYSM A G2 COMPLEX FILL COIL 2X1.5 (B)(4) GOT STRETCHED WHILE REPOSITIONING. THE COIL WAS REMOVED ALONG WITH THE ECHELON 10 MICROCATHETER AND SLIGHT RESISTANCE WAS FELT. NO RESISTANCE WAS ENCOUNTERED BETWEEN THE GUIDEWIRE AND THE MICROCATHETER WHEN ACCESSING THE TARGET SITE. NOR WAS THERE ANY RESISTANCE WHEN ADVANCING THE COIL THROUGH THE MICROCATHETER. OTHER COILS HAD SUCCESSFULLY BEEN PASSED THROUGH THE SAME MICROCATHETER WITHOUT ISSUE. THE G2 COMPLEX FILL COIL 2X1.5 (COMPLAINT PRODUCT) WAS THE LAST COIL ATTEMPTED TO FILL THE NECK. A ONE TO ONE RELATIONSHIP BETWEEN THE COIL AND DELIVERY TUBE WAS NOT VERIFIED WITH FLUORO PRIOR TO REPOSITIONING. ADDITIONAL INTERVENTION WAS NOT REQUIRED AS THE CASE WAS ABANDONED. THERE WERE NO PATIENT COMPLICATIONS OR PROCEDURAL DELAY AS A RESULT OF THIS EVENT. REPORTEDLY, THE ANEURYSM WAS 4MM IN HEIGHT, 3.5MM IN WIDTH, AND THE NECK WAS 2.5MM WITH NORMAL VESSEL CHARACTERISTICS. NECK REMODELING WAS NOT UTILIZED DURING THIS PROCEDURE. AN ADEQUATE CONTINUOUS FLUSH WAS MAINTAINED THROUGH THE MICROCATHETER AT ALL TIMES. T

    Patient problems: No Consequences Or Impact To Patient