ORBIT GALAXY ENPOWER 641CF0515Recalls & reported problems
Medos International Sàrl · ORBIT GALAXY G2 microcoil Stretch Resistant Complex Fill Coil with Tight Distal Loop Technology 5mm x 15cm 190cm Contents: 1 Detachable Coil System Coil Type: Stretch Resistant Complex Fill Coil with Tight Distal Loop Technology · Class 2 · Implant
All models of ORBIT GALAXY ENPOWER (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10878528009570), received between Oct 19, 2016 and Mar 11, 2019 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Positioning Failure2
- Stretched2
- Material Separation1
- Separation Failure1
Patient problems
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2019Stretched; Separation Failure; Stretched; Separation Failure
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE COIL EMBOLIZATION INTERVENTION PROCEDURE OF A CEREBRAL ANEURYSM AT THE CAROTID SYPHON THAT WAS CARRIED OUT CORRECTLY, IT WAS REPORTED THAT THE 5MM X 15CM G2 STRETCH RESISTANT COMPLEX FILL COIL (641CF0515 / S14134) FAILED TO DETACH. IT WAS NOT CLEAR IF THE DETACHMENT ISSUE IS RELATED TO THE COIL OR TO THE CODMAN® CONNECTING CABLE (CCB00015700 / S13256). THE G2 COIL WAS REMOVED WITHOUT ANY ADVERSE CONSEQUENCE TO THE PATIENT; IT WAS STILL ATTACHED TO THE DELIVERY SYSTEM AND WAS NOT STRETCHED WHEN IT WAS REMOVED. IT WAS CONFIRMED THAT A PRE-DEPLOYMENT ELECTRICAL CHECK WAS PERFORMED. THE CONTACT LIGHT DID NOT LIGHT UP WHEN THE COIL WAS CONNECTED, AND THE GREEN SYSTEM READY LIGHT DID NOT ILLUMINATE. THE AUDIBLE SIGNAL DID NOT SOUND DURING THE DETACHMENT CYCLE. ALL THE CONNECTIONS FIT PROPERLY WITHOUT THE APPLICATION OF EXCESSIVE FORCE. THERE WAS NO REPORT OF ANY PATIENT INJURY AS A RESULT OF THE FAILURE OF THE G2 COIL FAILING TO DETACH. THE REPORTED EVENT DID NOT RESULT IN ANY CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THE G2 COIL IS AVAILABLE TO BE RETURNED FOR EVALUATION AND ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8409940 ↗MalfunctionMar 2, 2017Positioning Failure; Positioning Failure
THE CONTACT FROM THE FACILITY REPORTED THAT THE ORBIT GALAXY COIL (641CF0515/C12282) DID NOT DETACH DURING THE PROCEDURE. NO PATIENT INJURY OCCURRED. THE DEVICE WAS NOT AVAILABLE FOR RETURN. THERE WAS NO SIGNIFICANT CLINICAL DELAY REPORTED. THE PROCEDURE WAS CONTINUED WITH ANOTHER COIL. THERE WERE NO DAMAGES NOTED TO THE COIL AND COIL DELIVERY SYSTEM PRIOR TO USE OR AFTER REMOVAL. THE COIL WAS SUCCESSFULLY REMOVED FROM THE PATIENT AND STILL ATTACHED TO THE DELIVERY SYSTEM WHEN REMOVED. NO PRE-DEPLOYMENT ELECTRICAL CHECK WAS PERFORMED. NO LOW BATTERY LIGHT WAS SEEN DURING THE CASE. NO FAULT LIGHT WAS SEEN. THE SYSTEM READY LIGHT DID NOT ILLUMINATE. THE DETACHMENT LIGHT DID NOT ILLUMINATE DURING THE DETACHMENT CYCLE AND THE AUDIBLE SIGNAL DID NOT BEEP. ALL CONNECTIONS APPEARED TO FIT PROPERLY WITHOUT USE OF EXCESSIVE FORCE. THE SAME CONNECTING CABLE AND DETACHMENT CONTROL BOX WAS USED SUCCESSFULLY WITH SUBSEQUENT COILS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6375132 ↗InjuryFeb 9, 2017Material Separation; Stretched; Material Separation; Stretched
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING A COILING PROCEDURE, A G2 ORBIT COIL (641CF0515/ P10965) STRETCHED AND THEN SEPARATED IN TWO PIECES WHILE REPOSITIONING. RESULTING IN THE COIL PROTRUDING INTO THE INTERNAL CAROTID ARTERY. A STENT WAS DEPLOYED TO JAIL THE HANGING COIL TO THE INTERNAL CAROTID ARTERY. THE ECHELON 10 MICROCATHETER HAD NOT BEEN REPOSITIONED OVER THE COIL WHILE THE COIL WAS DEPLOYED OR PARTIALLY DEPLOYED OUT OF THE DISTAL END OF THE MICROCATHETER. THERE WAS RESISTANCE WHILE RE-POSITIONING THE COIL, AND COIL ENTANGLEMENT WITH PREVIOUSLY PLACED COILS MAY HAVE CONTRIBUTED TO THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6314112 ↗MalfunctionOct 19, 2016Positioning Failure; Positioning Failure
AS REPORTED BY A HEALTHCARE PROFESSIONAL, DURING COIL EMBOLIZATION OF A BASILAR ARTERY ANEURYSM, A G2 TDL COMPLEX FILL COIL (641CF0515/ P11905) COULD NOT BE DETACHED, AND WAS SUCCESSFULLY REMOVED TOGETHER WITH THE SL 10 MICROCATHETER. THE COIL WAS NOT VISIBLY STRETCHED, AND WAS STILL ATTACHED TO THE DELIVERY SYSTEM. THERE WAS NO RESISTANCE BETWEEN THE G2 AND THE MICROCATHETER. A SECOND DETACHMENT CABLE WAS USED; HOWEVER, IT WAS REPORTED THAT THE SAME ENPOWER DETACHMENT CONTROL BOX (DCB) AND CABLE WERE USED SUCCESSFULLY TO DETACH SUBSEQUENT COILS. THE GREEN LIGHT OF THE BOX WAS ON FOR BOTH CABLES, BUT NEVERTHELESS NO DETACHMENT WAS OBTAINED. A PRE-DEPLOYMENT ELECTRICAL CHECK HAD BEEN PERFORMED. LOW BATTERY LIGHT WAS NOT SEEN DURING THE CASE, AND FAULT LIGHT WAS NOT SEEN. UPON PRESSING THE POWER BUTTON, ALL LIGHTS ILLUMINATED. THE GREEN SYSTEM READY LIGHT ILLUMINATED. DURING THE DETACHMENT CYCLE, THE DETACHMENT LIGHT ILLUMINATED AND THE AUDIBLE SIGNAL BEEPED. ALL CONNECTIONS APPEARED TO FIT PROPERLY WITHOUT APPLICATION OF EXCESSIVE FORCE. THE PROCEDURE WAS COMPLETED AND THE DELAY WAS LESS THAN 30 MINUTES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6039723 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page